Availability: | |
---|---|
Quantity: | |
TK6800
TUNKIA
(Pictures reference only)
TK6800 is a calibration device dedicated to transformer testers. The calibration objects include transformer no-load loss testers, transformer capacity testers and transformer short-circuit impedance testers. Reference standard: DL/T 1256-2013 General Specification of Transformer Tester for No-load Loss and Load Loss.
• Built-in three-phase voltage standard source: 6 V ~864 V
• Built-in three-phase current standard source: 50 mA ~ 60 A
• Built-in three-phase current standard meter: 10 mA ~ 10 A
• Voltage / current accuracy: Class 0.02
• Frequency and phase are adjustable; virtual power output
• Harmonic function: 2~63 times, adjustable content and phase
• Ergonomically designed console combined with removable control panel operation.
☆ No-load or Load Test, Short Circuit Impedance Test |
• When calibrating the no-load tester, TK6800 can switch the output transformer rated voltage or 50%~100% rated current, adjust the power output, realize the simulation of no-load and load loss, and calibrate the no-load tester. • During the load test, the short-circuit impedance value can also be obtained at the same time to complete the calibration of the short-circuit impedance tester. Parameters can be used for transformer capacity calculations. |
☆ Capacity Test |
• When calibrating the capacity tester, TK6800 selects the corresponding gear according to the resistance value obtained by the short-circuit impedance test. Set the voltage output according to the calibration coefficient and measure the tester output current value. Calculate and determine the transformer capacity value through the obtained short-circuit impedance voltage and actual short-circuit impedance value, and calibrate the capacity tester. |
(Pictures reference only)
TK6800 is a calibration device dedicated to transformer testers. The calibration objects include transformer no-load loss testers, transformer capacity testers and transformer short-circuit impedance testers. Reference standard: DL/T 1256-2013 General Specification of Transformer Tester for No-load Loss and Load Loss.
• Built-in three-phase voltage standard source: 6 V ~864 V
• Built-in three-phase current standard source: 50 mA ~ 60 A
• Built-in three-phase current standard meter: 10 mA ~ 10 A
• Voltage / current accuracy: Class 0.02
• Frequency and phase are adjustable; virtual power output
• Harmonic function: 2~63 times, adjustable content and phase
• Ergonomically designed console combined with removable control panel operation.
☆ No-load or Load Test, Short Circuit Impedance Test |
• When calibrating the no-load tester, TK6800 can switch the output transformer rated voltage or 50%~100% rated current, adjust the power output, realize the simulation of no-load and load loss, and calibrate the no-load tester. • During the load test, the short-circuit impedance value can also be obtained at the same time to complete the calibration of the short-circuit impedance tester. Parameters can be used for transformer capacity calculations. |
☆ Capacity Test |
• When calibrating the capacity tester, TK6800 selects the corresponding gear according to the resistance value obtained by the short-circuit impedance test. Set the voltage output according to the calibration coefficient and measure the tester output current value. Calculate and determine the transformer capacity value through the obtained short-circuit impedance voltage and actual short-circuit impedance value, and calibrate the capacity tester. |
☆ Front Panel | |
Number | Function |
Wiring panel, including no-load loss, capacity and short circuit impedance test voltage / current wiring port; | |
Emergency stop button, emergency stop equipment when the equipment output is abnormal; | |
Power supply switch, equipment power supply switch | |
Connecting terminal: LAN interface and RS232 interface for communication with computer; | |
The tested platform for placing the inspected equipment | |
Control the flat plate, display and operate the equipment output test; | |
Host power supply | |
☆ Wiring Panel | |
Number | Function |
No-load loss tester wiring terminal, including three-phase voltage / current output maximum of 864 V, 60 A. | |
Capacity tester terminal with three-phase voltage output 10 V and three-phase current measurement 10 A. | |
Terminal terminal of short-circuit impedance tester with maximum three-phase voltage output 864 V and AC current output maximum 60 A. |
☆ High-stability output |
• TK6800 standard source output has high stability. It adopts the design principle of closed-loop negative feedback system, which can measure and adjust the output power value in real time, ensure the real-time stability of the output, and can perform high repeatability measurement for a long time. The typical stability of AC voltage/current output is 0.005%/min. |
☆ Easy to operate |
• TK6800 can directly set the calibration point of no-load loss or directly set the capacity value and short-circuit impedance value for testing, without the need for the user to set the voltage value and current value one by one after calculation. |
☆ Harmonic Function | |
Number | Function |
Set the amplitude of the selected harmonic, either by setting the absolute value of the voltage or current harmonic, or by the base wave percentage. | |
Set the phase of the selected harmonic (relative to the base wave). | |
The instrument has the output function of the 2nd~21st harmonics. The embedded software is configured with the user custom "Harmonic parameter Settings" interface, and the user can set it according to the harmonics required by the test demand switch. | |
With the function of oscilloscope, it can observe the AC power waveform of each phase after loading the harmonic. | |
Channel selection, select the waveform of each phase voltage or current channel displayed. | |
Display the spectrum of each harmonic (Fundamental wave is 100%). |
☆ Front Panel | |
Number | Function |
Wiring panel, including no-load loss, capacity and short circuit impedance test voltage / current wiring port; | |
Emergency stop button, emergency stop equipment when the equipment output is abnormal; | |
Power supply switch, equipment power supply switch | |
Connecting terminal: LAN interface and RS232 interface for communication with computer; | |
The tested platform for placing the inspected equipment | |
Control the flat plate, display and operate the equipment output test; | |
Host power supply | |
☆ Wiring Panel | |
Number | Function |
No-load loss tester wiring terminal, including three-phase voltage / current output maximum of 864 V, 60 A. | |
Capacity tester terminal with three-phase voltage output 10 V and three-phase current measurement 10 A. | |
Terminal terminal of short-circuit impedance tester with maximum three-phase voltage output 864 V and AC current output maximum 60 A. |
☆ High-stability output |
• TK6800 standard source output has high stability. It adopts the design principle of closed-loop negative feedback system, which can measure and adjust the output power value in real time, ensure the real-time stability of the output, and can perform high repeatability measurement for a long time. The typical stability of AC voltage/current output is 0.005%/min. |
☆ Easy to operate |
• TK6800 can directly set the calibration point of no-load loss or directly set the capacity value and short-circuit impedance value for testing, without the need for the user to set the voltage value and current value one by one after calculation. |
☆ Harmonic Function | |
Number | Function |
Set the amplitude of the selected harmonic, either by setting the absolute value of the voltage or current harmonic, or by the base wave percentage. | |
Set the phase of the selected harmonic (relative to the base wave). | |
The instrument has the output function of the 2nd~21st harmonics. The embedded software is configured with the user custom "Harmonic parameter Settings" interface, and the user can set it according to the harmonics required by the test demand switch. | |
With the function of oscilloscope, it can observe the AC power waveform of each phase after loading the harmonic. | |
Channel selection, select the waveform of each phase voltage or current channel displayed. | |
Display the spectrum of each harmonic (Fundamental wave is 100%). |
Three-phase Voltage Output | ||||
Range | Resolution | Short-term Stability ( %/min ) | Uncertainty (k=2) ( ppm*RD+ppm *RG)[1] | Maximum Load |
10 V | 0.1 mV | 0.005 | 120 + 80 | 10 mA |
Note [1]: RD is the read value and RG is the range value. |
Voltage output range: 0.1 V ~10 V, distortion degree: <0.2%
Symmetry: voltage above 0.2% and phase better than 0.5°
Frequency adjustment range: 45 Hz ~70 Hz, adjustment fineness: 0.001 Hz, measurement uncertainty: 0.005 Hz
Phase adjustment range: 0.000°~359.999°, adjustment fineness: 0.001°, measurement uncertainty: 0.05°
Three-phase Current Measurement | ||
Range | Resolution | Uncertainty (k=2) ( ppm*RD+ppm *RG) |
100 mA | 1 μA | 120 + 80 |
1 A | 10 μA | 120 + 80 |
5 A | 10 μA | 120 + 80 |
10A | 0.1 mA | 120 + 80 |
Range of current measurement: 10 mA ~ 10 A
Frequency range: 45 Hz ~70 Hz; measurement uncertainty: 0.003 Hz
Phase range: 0.000°~359.999°; measurement uncertainty: 0.01°
Three-phase voltage output | ||||
Range | Resolution | Short-term Stability ( %/min ) | Uncertainty (k=2) ( ppm*RD+ppm *RG)[1] | Maximum Load |
60 V | 0.1 mV | 0.005 | 120 + 80 | 25 VA |
120 V | 1 mV | 0.005 | 120 + 80 | 25 VA |
240 V | 1 mV | 0.005 | 120 + 80 | 25 VA |
480 V | 1 mV | 0.005 | 120 + 80 | 25 VA |
720 V | 1 mV | 0.005 | 120 + 80 | 25 VA |
Voltage output range: 6V ~864V, distortion degree: <0.2%
Symmetry: voltage above 0.2% and phase better than 0.5°
Three-phase Current Output | |||
Range | Resolution | Short-term stability ( %/min ) | Uncertainty (k=2) ( ppm*RD+ppm *RG)[1] |
500 mA | 1 μA | 0.005 | 120 + 80 |
1 A | 10 μA | 0.005 | 120 + 80 |
5 A | 10 μA | 0.005 | 120 + 80 |
10 A | 0.1 mA | 0.005 | 120 + 80 |
20 A | 0.1 mA | 0.005 | 120 + 80 |
50 A | 0.1 mA | 0.005 | 120 + 80 |
Current output range: 50 mA ~ 60 A, distortion degree: <0.2%
Symmetry degree: Current is better than 0.5%, and phase is better than 0.5°
Frequency | Range | 45 Hz ~400 Hz |
Fineness Adjustment | 0.001 Hz | |
Uncertainty (k=2) | 0.005 Hz | |
Phase | Range | 0.000°~359.999° |
Fineness Adjustment | 0.001° | |
Uncertainty (k=2) | 0.01° | |
Harmonic | Harmonic Times | 2~63 |
Amplitude Range | 0~25% | |
Phase range | 0~359.99° |
Three-phase Voltage Output | ||||
Range | Resolution | Short-term Stability ( %/min ) | Uncertainty (k=2) ( ppm*RD+ppm *RG)[1] | Maximum Load |
60 V | 0.1 mV | 0.005 | 120 + 80 | 25 VA |
120 V | 1 mV | 0.005 | 120 + 80 | 25 VA |
240 V | 1 mV | 0.005 | 120 + 80 | 25 VA |
480 V | 1 mV | 0.005 | 120 + 80 | 25 VA |
Voltage output range: 6V ~576V, distortion degree: <0.2%
Symmetry: voltage above 0.2% and phase better than 0.5°
AC Current Output | |||
Range | Resolution | Short-term Stability ( %/min ) | Uncertainty (k=2) ( ppm*RD+ppm *RG)[1] |
500 mA | 1 μA | 0.005 | 120 + 80 |
1 A | 10 μA | 0.005 | 120 + 80 |
5 A | 10 μA | 0.005 | 120 + 80 |
10 A | 0.1 mA | 0.005 | 120 + 80 |
20 A | 0.1 mA | 0.005 | 120 + 80 |
50 A | 0.1 mA | 0.005 | 120 + 80 |
Current output range: 50 mA ~ 60 A, distortion degree: <0.2%
Symmetry: Current is better than 0.5%, and phase is better than 0.5°
Frequency | Range | 45 Hz ~400 Hz |
Fineness Adjustment | 0.001 Hz | |
Uncertainty (k=2) | 0.005 Hz | |
Phase | Range | 0.000°~359.999° |
Fineness Adjustment | 0.001° | |
Uncertainty (k=2) | 0.01° | |
Harmonic | Harmonic Times | 2~63 Times |
Amplitude Range | 0~25% | |
Phase range | 0~359.99° |
Uncertainty of No-load loss measurement: 0.1%; capacity test measurement uncertainty: 0.1%
Uncertainty of impedance voltage measurement: 0.05%; short circuit impedance: 1.5 Ω ~1000 Ω, measurement uncertainty: 0.1%
Power Supply | AC ( 220 ± 22 ) V ,( 50 ± 2 ) Hz |
Warm-up time | 20 Minutes |
Temperature Performance | Operating temperature: 15°C~35°C Storage temperature: -20°C~50°C |
Humidity Performance | Working Humidity: <80% @ 30°C, <70% @ 40°C, and <40% @ 50°C Storage humidity: (20%~80%) R·H, no condensation |
Altitude | < 3000 m |
Communication Interface | LAN 、RS232 |
Dimension | 505 mm (W) × 700 mm (D) × 1200 mm (H) |
Number | Picture | Name | Specifications | Quantity | Note |
1 | Voltage Test Leads | 1.5m / 2.1mm2/ Φ 4- Φ 4-Socket | Red 1 Green 1 Yellow 1 Black 3 | Standard accessories | |
2 | Test Leads | 1.5m / 8mm2/ Φ 4- Φ 4-Socket | Red 1 Green 1 Yellow 1 Black 3 | Standard accessories | |
3 | Tablet | With the wireless WiFi module, equipped with the APP operation software | 1 | Standard accessories | |
4 | Network Cable | 1 Meters | 1 | Standard accessories | |
5 | Network Cable | 5 Meters | 1 | Standard accessories | |
6 | Power Cable | AC 220V 、10A | 1 | Standard accessories | |
7 | Glass fuse | F 8A 、250V | 3 | Standard accessories |
Three-phase Voltage Output | ||||
Range | Resolution | Short-term Stability ( %/min ) | Uncertainty (k=2) ( ppm*RD+ppm *RG)[1] | Maximum Load |
10 V | 0.1 mV | 0.005 | 120 + 80 | 10 mA |
Note [1]: RD is the read value and RG is the range value. |
Voltage output range: 0.1 V ~10 V, distortion degree: <0.2%
Symmetry: voltage above 0.2% and phase better than 0.5°
Frequency adjustment range: 45 Hz ~70 Hz, adjustment fineness: 0.001 Hz, measurement uncertainty: 0.005 Hz
Phase adjustment range: 0.000°~359.999°, adjustment fineness: 0.001°, measurement uncertainty: 0.05°
Three-phase Current Measurement | ||
Range | Resolution | Uncertainty (k=2) ( ppm*RD+ppm *RG) |
100 mA | 1 μA | 120 + 80 |
1 A | 10 μA | 120 + 80 |
5 A | 10 μA | 120 + 80 |
10A | 0.1 mA | 120 + 80 |
Range of current measurement: 10 mA ~ 10 A
Frequency range: 45 Hz ~70 Hz; measurement uncertainty: 0.003 Hz
Phase range: 0.000°~359.999°; measurement uncertainty: 0.01°
Three-phase voltage output | ||||
Range | Resolution | Short-term Stability ( %/min ) | Uncertainty (k=2) ( ppm*RD+ppm *RG)[1] | Maximum Load |
60 V | 0.1 mV | 0.005 | 120 + 80 | 25 VA |
120 V | 1 mV | 0.005 | 120 + 80 | 25 VA |
240 V | 1 mV | 0.005 | 120 + 80 | 25 VA |
480 V | 1 mV | 0.005 | 120 + 80 | 25 VA |
720 V | 1 mV | 0.005 | 120 + 80 | 25 VA |
Voltage output range: 6V ~864V, distortion degree: <0.2%
Symmetry: voltage above 0.2% and phase better than 0.5°
Three-phase Current Output | |||
Range | Resolution | Short-term stability ( %/min ) | Uncertainty (k=2) ( ppm*RD+ppm *RG)[1] |
500 mA | 1 μA | 0.005 | 120 + 80 |
1 A | 10 μA | 0.005 | 120 + 80 |
5 A | 10 μA | 0.005 | 120 + 80 |
10 A | 0.1 mA | 0.005 | 120 + 80 |
20 A | 0.1 mA | 0.005 | 120 + 80 |
50 A | 0.1 mA | 0.005 | 120 + 80 |
Current output range: 50 mA ~ 60 A, distortion degree: <0.2%
Symmetry degree: Current is better than 0.5%, and phase is better than 0.5°
Frequency | Range | 45 Hz ~400 Hz |
Fineness Adjustment | 0.001 Hz | |
Uncertainty (k=2) | 0.005 Hz | |
Phase | Range | 0.000°~359.999° |
Fineness Adjustment | 0.001° | |
Uncertainty (k=2) | 0.01° | |
Harmonic | Harmonic Times | 2~63 |
Amplitude Range | 0~25% | |
Phase range | 0~359.99° |
Three-phase Voltage Output | ||||
Range | Resolution | Short-term Stability ( %/min ) | Uncertainty (k=2) ( ppm*RD+ppm *RG)[1] | Maximum Load |
60 V | 0.1 mV | 0.005 | 120 + 80 | 25 VA |
120 V | 1 mV | 0.005 | 120 + 80 | 25 VA |
240 V | 1 mV | 0.005 | 120 + 80 | 25 VA |
480 V | 1 mV | 0.005 | 120 + 80 | 25 VA |
Voltage output range: 6V ~576V, distortion degree: <0.2%
Symmetry: voltage above 0.2% and phase better than 0.5°
AC Current Output | |||
Range | Resolution | Short-term Stability ( %/min ) | Uncertainty (k=2) ( ppm*RD+ppm *RG)[1] |
500 mA | 1 μA | 0.005 | 120 + 80 |
1 A | 10 μA | 0.005 | 120 + 80 |
5 A | 10 μA | 0.005 | 120 + 80 |
10 A | 0.1 mA | 0.005 | 120 + 80 |
20 A | 0.1 mA | 0.005 | 120 + 80 |
50 A | 0.1 mA | 0.005 | 120 + 80 |
Current output range: 50 mA ~ 60 A, distortion degree: <0.2%
Symmetry: Current is better than 0.5%, and phase is better than 0.5°
Frequency | Range | 45 Hz ~400 Hz |
Fineness Adjustment | 0.001 Hz | |
Uncertainty (k=2) | 0.005 Hz | |
Phase | Range | 0.000°~359.999° |
Fineness Adjustment | 0.001° | |
Uncertainty (k=2) | 0.01° | |
Harmonic | Harmonic Times | 2~63 Times |
Amplitude Range | 0~25% | |
Phase range | 0~359.99° |
Uncertainty of No-load loss measurement: 0.1%; capacity test measurement uncertainty: 0.1%
Uncertainty of impedance voltage measurement: 0.05%; short circuit impedance: 1.5 Ω ~1000 Ω, measurement uncertainty: 0.1%
Power Supply | AC ( 220 ± 22 ) V ,( 50 ± 2 ) Hz |
Warm-up time | 20 Minutes |
Temperature Performance | Operating temperature: 15°C~35°C Storage temperature: -20°C~50°C |
Humidity Performance | Working Humidity: <80% @ 30°C, <70% @ 40°C, and <40% @ 50°C Storage humidity: (20%~80%) R·H, no condensation |
Altitude | < 3000 m |
Communication Interface | LAN 、RS232 |
Dimension | 505 mm (W) × 700 mm (D) × 1200 mm (H) |
Number | Picture | Name | Specifications | Quantity | Note |
1 | Voltage Test Leads | 1.5m / 2.1mm2/ Φ 4- Φ 4-Socket | Red 1 Green 1 Yellow 1 Black 3 | Standard accessories | |
2 | Test Leads | 1.5m / 8mm2/ Φ 4- Φ 4-Socket | Red 1 Green 1 Yellow 1 Black 3 | Standard accessories | |
3 | Tablet | With the wireless WiFi module, equipped with the APP operation software | 1 | Standard accessories | |
4 | Network Cable | 1 Meters | 1 | Standard accessories | |
5 | Network Cable | 5 Meters | 1 | Standard accessories | |
6 | Power Cable | AC 220V 、10A | 1 | Standard accessories | |
7 | Glass fuse | F 8A 、250V | 3 | Standard accessories |