TI1200 Precision DC Resistance Calibrator

TI1200 is a high-precision DC resistance source with wide range, it is applied for calibrating ohmmeter on-site, e.g, industrial production lines, etc. It can calibrate DC resistance meters, calibrate digital multimeters, calibrate DC bridges, microhmmeters.

Features:
Availability:
Quantity:
  • TI1200

  • TUNKIA

TI1200 Precision DC Resistance Calibrator

TI1200 Precision DC Resistance Calibrator 3

1. Summary

TI1200 is a high-precision DC resistance source with wide range, it is applied for calibrating ohmmeter on-site, e.g, industrial production lines, etc.


2. Features

  • Resistance Output: 0 Ω ~ 1100 MΩ

  • Accuracy: 50 ppm

  • Shortcut output keys

  • Programmable calibration schemes

  • RS232, LAN and USB communication interfaces

 

3. Application

  • Calibrate DC resistance meters

  • Calibrate digital multimeters

  • Calibrate DC bridges, microhmmeters.




4. Features

Convenient Operation

TI1200 Precision DC Resistance Calibrator The front panel

The front panel has 10 shortcut keys that can output the corresponding value on the screen with one key.


Calibration Scheme Setting

TI1200 Precision DC Resistance Calibrator Calibration Scheme Setting

Multiple calibration schemes can be edited and stored for the calibration of multimeter with different specifications.


5. Specifications

Range

Resistance Output

Resolution

Accuracy

         ±(ppm*output+Ω)

Allowable Current

10 Ω

0 Ω ~ 11.00000 Ω

10 μΩ

50 + 0.01

1 mA ~ 150 mA

30 Ω

10.00000 Ω ~ 33.00000 Ω

10 μΩ

40 + 0.015

1 mA ~ 150 mA

100 Ω

30.0000 Ω ~ 110.0000 Ω

100 μΩ

35 + 0.015

1 mA ~ 80 mA

300 Ω

100.0000 Ω ~ 330.0000 Ω

100 μΩ

35 + 0.02

1 mA ~ 40 mA

1kΩ

0.300000 kΩ ~ 1.100000 kΩ

1 mΩ

35 + 0.02

1 mA ~ 20 mA

3 kΩ

1.000000 kΩ ~ 3.300000 kΩ

1 mΩ

35 + 0.15

0.1 mA ~ 6 mA

10 kΩ

3.00000 kΩ ~ 11.00000 kΩ

10 mΩ

35 + 0.15

0.1 mA ~ 2 mA

30 kΩ

10.00000 kΩ ~ 33.00000 kΩ

10 mΩ

35 + 2

10 μA ~ 0.6 mA

100 kΩ

30.0000 kΩ ~ 110.0000 kΩ

100 mΩ

35 + 3

10 μA ~ 0.2 mA

300 kΩ

100.0000 kΩ ~ 330.0000 kΩ

100 mΩ

40 + 30

1 μA ~ 60 μA

1 MΩ

0.300000 MΩ ~ 1.100000 MΩ

1 Ω

40 + 30

1 μA ~ 20 μA

3 MΩ

1.000000 MΩ ~ 3.300000 MΩ

1 Ω

50 + 300

0.25 μA ~ 6 μA

10 MΩ

3.00000 MΩ ~ 11.00000 MΩ

10 Ω

200 + 500

0.25 μA ~ 2 μA

30 MΩ

10.00000 MΩ ~ 33.00000 MΩ

10 Ω

400 + 5000

25 nA ~ 600 nA

100 MΩ

30.0000 MΩ ~ 110.0000 MΩ

100 Ω

1000 + 6000

25 nA ~ 200 nA

300 MΩ

100.0000 MΩ ~ 330.0000 MΩ

100 Ω

5000 + 200000

2.5 nA ~ 60 nA

1000 MΩ

300.000 MΩ ~ 1100.000 MΩ

1 kΩ

20000 + 600000

1 nA ~ 20 nA

Output range: 0 Ω ~ 1100 MΩ, 7 digits display.

Adjustment fineness: 0.002%*RG.

Manual or automatic range switch.

Four-terminals or Two-terminals.

Reverse connect protection, Overload protection

6. General Specifications

Power Supply

AC (220±22) V, (50±2) Hz

Maximum Power Consumption

100 VA

Warm-up Time

Twice the time since last warmed up, to a maximum of 30 minutes.

Temperature Performance

0°C ~ 55°C, 30%R·H ~ 80%R·H, non-condensing

Humidity Performance

-30°C ~ 70°C, 10%R·H ~ 90%R·H, non-condensing

Communication Interface

USB×1LAN×1RS232×1

Dimensions

215 mm (width) x 252 mm (depth) x 88 mm (height), excluding feets.

Display

3.5-inch color LCD

Weight

About 5 kg

 

Previous: 
Next: 

SIGN UP FOR NEWSLETTER

Get latest deals, newsletter and technical information

INFORMATION

CONTACT US

 WhatsApp:+86-18773180215
     Wechat: 18773180215
 Phone: +86-18773180215
 E-mail: global@tunkia.com
E-mail: Eugene@tunkia.com
 Address: NO.16 Panpan Road, Changsha
Copyright  2021 TUNKIA Co., Ltd. All Rights Reserved. Supported by leadong.comSitemap