TK6500 is a precision Electric Welding Machine AC-DC Power Calibrator for on-site calibration, which is suitable for calibrating the internal ammeter and voltmeter without disassembling the welding machine. TK6500 is a precision AC and DC power calibrator. The maximum measurement current can reach 3 kA, and the accuracy can be selected in two specifications: Class 0.05 and Class 0.1.
Availability: | |
---|---|
Quantity: | |
TK6500
TUNKIA
1. Summary
TK6500 is a precision Electric Welding Machine AC-DC Power Calibrator for on-site calibration, which is suitable for calibrating the internal ammeter and voltmeter without disassembling the welding machine. TK6500 is a precision AC and DC power calibrator. The maximum measurement current can reach 3 kA, and the accuracy can be selected in two specifications: Class 0.05 and Class 0.1.
2. Features
• AC and DC Voltage Measurement: 5 V~180 V (optional 500 V).
• AC and DC Current Through-center Measurement: 10 A~3 kA
• F:DC,40 Hz~1 kHz
• Voltage/Current Accuracy: class 0.05 / class 0.1
• Harmonic Measurement Function
• USB Data Export Function
3. Applications
☆ Calibrate the AC and DC Welding Machine Power Supply |
• The AC-DC voltage measurement range is 5 V~180 V (optional 500 V). AC-DC current measurement range: 10 A~3 kA; • The best accuracy of this equipment is class 0.05, which is suitable for calibrating AC-DC arc welding power supply or other high-current output equipment of level 0.2 and below |
1. Summary
TK6500 is a precision Electric Welding Machine AC-DC Power Calibrator for on-site calibration, which is suitable for calibrating the internal ammeter and voltmeter without disassembling the welding machine. TK6500 is a precision AC and DC power calibrator. The maximum measurement current can reach 3 kA, and the accuracy can be selected in two specifications: Class 0.05 and Class 0.1.
2. Features
• AC and DC Voltage Measurement: 5 V~180 V (optional 500 V).
• AC and DC Current Through-center Measurement: 10 A~3 kA
• F:DC,40 Hz~1 kHz
• Voltage/Current Accuracy: class 0.05 / class 0.1
• Harmonic Measurement Function
• USB Data Export Function
3. Applications
☆ Calibrate the AC and DC Welding Machine Power Supply |
• The AC-DC voltage measurement range is 5 V~180 V (optional 500 V). AC-DC current measurement range: 10 A~3 kA; • The best accuracy of this equipment is class 0.05, which is suitable for calibrating AC-DC arc welding power supply or other high-current output equipment of level 0.2 and below |
4. Characteristics
☆ Wide Range of Current Measurement |
• The instrument adopts the method of measuring the current through the core, which can reduce the wiring work of the current loop and improve the work efficiency. • The wide range measurement capability can cover the calibration of the commonly used AC and DC arc welding machine power supply, and the user can choose different specifications of current sensors (500 A/ 1 kA/ 2 kA/ 3kA) according to the needs, such a combination. It is beneficial to reduce the size and volume of the host, and the choice of lightweight sensor is more convenient to carry to the field for calibration work. |
☆ USB Data Export | |
• Built-in large-capacity memory, the test data of the inspected equipment can be quickly saved on site. • After the calibration work is completed, the original data can be exported to the computer through the U disk for sorting and analysis. |
☆ APP Graphical and Intuitive Display | |
S/N | Function |
AC power measurement has an oscilloscope function that displays the waveform of the AC signal under test in real time. | |
Calculate two types of harmonic distortion: THD/T (harmonic relative to full wave) and THD/F ( harmonic relative to fundamental wave). | |
The amplitude (RMS), content (%) and phase of the 2nd~128th harmonic of voltage or current can be analyzed in real time. | |
Visualize the spectrum of each harmonic in the form of a histogram (100% fundamental). |
☆ Data Statistical Analysis | |
S/N | Function |
Visual Display Of The Measured Power: The measured voltage and current measurement results are displayed at the same time | |
Power Trend Test: Curve the power over time in real time during the test cycle. | |
Power Stability Analysis: calculate the maximum (MAX), minimum value (min), and peak-peak value of the measured electricity ( P-P ), mean ( Avg ) , standard deviation ( S .dev ) , divergence ( Div ), etc. |
4. Characteristics
☆ Wide Range of Current Measurement |
• The instrument adopts the method of measuring the current through the core, which can reduce the wiring work of the current loop and improve the work efficiency. • The wide range measurement capability can cover the calibration of the commonly used AC and DC arc welding machine power supply, and the user can choose different specifications of current sensors (500 A/ 1 kA/ 2 kA/ 3kA) according to the needs, such a combination. It is beneficial to reduce the size and volume of the host, and the choice of lightweight sensor is more convenient to carry to the field for calibration work. |
☆ USB Data Export | |
• Built-in large-capacity memory, the test data of the inspected equipment can be quickly saved on site. • After the calibration work is completed, the original data can be exported to the computer through the U disk for sorting and analysis. |
☆ APP Graphical and Intuitive Display | |
S/N | Function |
AC power measurement has an oscilloscope function that displays the waveform of the AC signal under test in real time. | |
Calculate two types of harmonic distortion: THD/T (harmonic relative to full wave) and THD/F ( harmonic relative to fundamental wave). | |
The amplitude (RMS), content (%) and phase of the 2nd~128th harmonic of voltage or current can be analyzed in real time. | |
Visualize the spectrum of each harmonic in the form of a histogram (100% fundamental). |
☆ Data Statistical Analysis | |
S/N | Function |
Visual Display Of The Measured Power: The measured voltage and current measurement results are displayed at the same time | |
Power Trend Test: Curve the power over time in real time during the test cycle. | |
Power Stability Analysis: calculate the maximum (MAX), minimum value (min), and peak-peak value of the measured electricity ( P-P ), mean ( Avg ) , standard deviation ( S .dev ) , divergence ( Div ), etc. |
5. Specifications
5.1 AC/DC Voltage Measurement
Voltage range | Resolution | Measurement uncertainty (k=2). ( ppm*RD+ppm*RG)[1] | Temperature coefficient ppm*RD /℃ @ ( 15~30 ) ℃ | ||
Class 0.1 | Class 0.05 | Class 0.1 | Class 0.05 | ||
50 V | 100 μV | 600 + 400 | 300 + 200 | <20 | <10 |
150 V | 1 mV | 600 + 400 | 300 + 200 | <20 | <10 |
500 V[2] | 1 mV | 600 + 400 | 300 + 200 | <20 | <10 |
Note [1]: RD is the reading value, RG is the range value, the same below. Note [2]: 500 V ranges are optional, other ranges can be customized upon request. |
• Measuring Range: 5 V~180 V or 5 V~500 V (optional), direct measurement, manual or automatic range switching
• Measurement Band: DC, 40 Hz~1 kHz
5.2 AC/DC Current Measurement
Current sensing selection [3]. | Current range | Resolution | Measurement uncertainty (k=2). ( ppm*RD+ppm*RG)[1] | |
Class 0.1 | Class 0.05 | |||
TK6500-T-500A | 100A, 500A | 1 mA | 600 + 400 | 300 + 200 |
TK6500-T-1kA | 200A, 1kA | 10 mA | 600 + 400 | 300 + 200 |
TK6500-T-2kA | 500A, 2kA | 10 mA | 600 + 400 | 300 + 200 |
TK6500-T-3kA | 1kA, 3kA | 10 mA | 600 + 400 | 300 + 200 |
Note [3]: The current conversion is carried out by an external current sensor and measured by the host. Users can select the sensor with the right specification according to the size of the measured current to achieve the best measurement accuracy. Other current ranges can be customized by customers. |
• Measuring Range: 10 A... 3 kA, indirect measurement, manual or automatic range switching
• Measurement Frequency: DC, 40 Hz~1 kHz
6. General Specifications
Power supply | AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz |
Temperature performance | Working Temperature: 0°C~45°C; Storage Temperature: -20 °C ~ 70 °C |
Humidity performance | Operating Humidity: < 80% @ 30°C, < 70% @ 40°C, < 40% @ 50°C Storage Humidity: (20%~80%) R·H, no Condensing |
altitude | < 3000 m |
Quality | Approx. 6.5 kg (without sensor). |
Communication interface | USB, LAN |
Size | 390 mm(W) × 395 mm(D) × 260 mm(H) |
7. Ordering Information
7.1 Main Instrument Ordering Information
7.2 Current Sensor Ordering Information
5. Specifications
5.1 AC/DC Voltage Measurement
Voltage range | Resolution | Measurement uncertainty (k=2). ( ppm*RD+ppm*RG)[1] | Temperature coefficient ppm*RD /℃ @ ( 15~30 ) ℃ | ||
Class 0.1 | Class 0.05 | Class 0.1 | Class 0.05 | ||
50 V | 100 μV | 600 + 400 | 300 + 200 | <20 | <10 |
150 V | 1 mV | 600 + 400 | 300 + 200 | <20 | <10 |
500 V[2] | 1 mV | 600 + 400 | 300 + 200 | <20 | <10 |
Note [1]: RD is the reading value, RG is the range value, the same below. Note [2]: 500 V ranges are optional, other ranges can be customized upon request. |
• Measuring Range: 5 V~180 V or 5 V~500 V (optional), direct measurement, manual or automatic range switching
• Measurement Band: DC, 40 Hz~1 kHz
5.2 AC/DC Current Measurement
Current sensing selection [3]. | Current range | Resolution | Measurement uncertainty (k=2). ( ppm*RD+ppm*RG)[1] | |
Class 0.1 | Class 0.05 | |||
TK6500-T-500A | 100A, 500A | 1 mA | 600 + 400 | 300 + 200 |
TK6500-T-1kA | 200A, 1kA | 10 mA | 600 + 400 | 300 + 200 |
TK6500-T-2kA | 500A, 2kA | 10 mA | 600 + 400 | 300 + 200 |
TK6500-T-3kA | 1kA, 3kA | 10 mA | 600 + 400 | 300 + 200 |
Note [3]: The current conversion is carried out by an external current sensor and measured by the host. Users can select the sensor with the right specification according to the size of the measured current to achieve the best measurement accuracy. Other current ranges can be customized by customers. |
• Measuring Range: 10 A... 3 kA, indirect measurement, manual or automatic range switching
• Measurement Frequency: DC, 40 Hz~1 kHz
6. General Specifications
Power supply | AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz |
Temperature performance | Working Temperature: 0°C~45°C; Storage Temperature: -20 °C ~ 70 °C |
Humidity performance | Operating Humidity: < 80% @ 30°C, < 70% @ 40°C, < 40% @ 50°C Storage Humidity: (20%~80%) R·H, no Condensing |
altitude | < 3000 m |
Quality | Approx. 6.5 kg (without sensor). |
Communication interface | USB, LAN |
Size | 390 mm(W) × 395 mm(D) × 260 mm(H) |