TK2600 is a pulsed voltage source device that can output a pulse voltage of up to 2 kV, and its pulse amplitude, pulse repetition rate and pulse width can be set and adjusted. The rise and fall time of pulse is less than 10 μs, and the step response performance is faster. The communication interface of the device can realize computer program control operation, which is suitable for measuring the response time of current sensor, Roche coil and other equipment, or calibrating pulse shunt, etc.
Features: | |
---|---|
Availability: | |
Quantity: | |
TK2600
TUNKIA
(Reference only)
1. Summary
TK2600 is a pulsed voltage source device that can output a pulse voltage of up to 2 kV, and its pulse amplitude, pulse repetition rate and pulse width can be set and adjusted. The rise and fall time of pulse is less than 10 μs, and the step response performance is faster. The communication interface of the device can realize computer program control operation, which is suitable for measuring the response time of current sensor, Roche coil and other equipment, or calibrating pulse shunt, etc.
2. Features
• Pulse voltage output: 200 V ~ 2 kV
• Pulse width: 1 ms ~ 10 ms
• Rise time: ≤10 μs
• Rising edge falling edge on-time: adjustable within 10 ms ~ 10 s
3. Applications
☆ Measure the Response Time of the Voltage Sensor |
• The resulting step voltage signal flows into both the oscilloscope and the voltage sensor under inspection. • Observe the waveforms of the sensor's primary and secondary signals through an oscilloscope. • Using the rise time of the pulse signal as a reference, the response time of the inspected sensor is measured. |
4. Characteristics
☆ Rapid Rise |
• The output voltage rises quickly, and the rise time when the output voltage is 2000 V is better than 10 μs, At this time, the dU/dt rate reaches 200 V/μs, which satisfies the electricity Requirements for signal sources in the relevant testing standards for pressure sensors. |
(Reference only)
1. Summary
TK2600 is a pulsed voltage source device that can output a pulse voltage of up to 2 kV, and its pulse amplitude, pulse repetition rate and pulse width can be set and adjusted. The rise and fall time of pulse is less than 10 μs, and the step response performance is faster. The communication interface of the device can realize computer program control operation, which is suitable for measuring the response time of current sensor, Roche coil and other equipment, or calibrating pulse shunt, etc.
2. Features
• Pulse voltage output: 200 V ~ 2 kV
• Pulse width: 1 ms ~ 10 ms
• Rise time: ≤10 μs
• Rising edge falling edge on-time: adjustable within 10 ms ~ 10 s
3. Applications
☆ Measure the Response Time of the Voltage Sensor |
• The resulting step voltage signal flows into both the oscilloscope and the voltage sensor under inspection. • Observe the waveforms of the sensor's primary and secondary signals through an oscilloscope. • Using the rise time of the pulse signal as a reference, the response time of the inspected sensor is measured. |
4. Characteristics
☆ Rapid Rise |
• The output voltage rises quickly, and the rise time when the output voltage is 2000 V is better than 10 μs, At this time, the dU/dt rate reaches 200 V/μs, which satisfies the electricity Requirements for signal sources in the relevant testing standards for pressure sensors. |
5. Specifications
Pulse Type | Unipolar pulses, rising edges, and falling edges |
Output Range | 200 V ~ 2 kV |
Adjust the Fineness | 0.1 V |
Single Pulse Width | Adjustable from 1 ms ~ 10 ms |
Single Pulse Rise Time | ≤10 μs |
Rising Edge, Falling Edge On-Time | Adjustable in the range of 10ms~10s |
Voltage Accuracy | ± 0.5%* range |
6. General Specifications
Power Supply | AC (220±22) V,(50±2) Hz |
Maximum Power Consumption | 150 VA |
Working Environment | Temperature: 15°C ~ 30°C Humidity: 20% R·H ~ 80%R·H, no condensing Others: No electromagnetic field interference. |
Storage Environment | Temperature: -20°C ~ 70°C Humidity: <95% R·H, no condensing |
Communication Interface | RS232×1 |
5. Specifications
Pulse Type | Unipolar pulses, rising edges, and falling edges |
Output Range | 200 V ~ 2 kV |
Adjust the Fineness | 0.1 V |
Single Pulse Width | Adjustable from 1 ms ~ 10 ms |
Single Pulse Rise Time | ≤10 μs |
Rising Edge, Falling Edge On-Time | Adjustable in the range of 10ms~10s |
Voltage Accuracy | ± 0.5%* range |
6. General Specifications
Power Supply | AC (220±22) V,(50±2) Hz |
Maximum Power Consumption | 150 VA |
Working Environment | Temperature: 15°C ~ 30°C Humidity: 20% R·H ~ 80%R·H, no condensing Others: No electromagnetic field interference. |
Storage Environment | Temperature: -20°C ~ 70°C Humidity: <95% R·H, no condensing |
Communication Interface | RS232×1 |