*Above picture is only for reference, subject to actual delivery
1. Summary
TS1100 is a special instrument for measuring the AC magnetic properties of electrical steel. It consists of excitation and measurement host, Epstein frame, Single Sheet Tester (optional), computer software, etc. It is suitable for comprehensive analysis of AC magnetic characteristics of electrical steel or core.
2. Features
• Electrical parameters calibration function.
• Measurement of magnetic field by M.C.
• Maximum magnetization field reaches 10 kA/m (700 T Epstein frame).
• Slowly reduce amplitude demagnetization for specimen.
• Software controlled automatic measurement.
• Automatic multi-point measurement.
• The software has been configured with multiple measurement schemes
• Automatic calculates magnetic parameters and curves.
• Modular design, easy to upgrade or maintenance.
3. The Reference Standard
IEC 60404-2 "Magnetic materials-Part 2 Methods of measurement of the magnetic properties of electrical steel strip and sheet by EP".
IEC 60404-3 "Magnetic materials-Part 3 Methods of measurement of the magnetic properties of electrical steel strip and sheet by SST".
IEC 60404-10 "Magnetic materials-Part 10-Methods of measurement of magnetic properties of electrical steel strip and sheet at medium frequencies".
4. Applications
Frequency | Parameter | Range[1] | Uncertainty (k=2) [5] | Repeatability[5] |
50Hz / 60Hz | Hm | 5 A/m...10 kA/m | 0.3% | 0.2% |
Jm / Bm | B5 [2]... B10000 | 0.3% | 0.1% | |
Ps | P0.1 [3]... P1.9 ( P1.7 ) | 0.5% | 0.2% | |
400 Hz [4] | Hm | 5 A/m...2.5 kA/m | 0.5% | 0.2% |
Jm / Bm | B5 [2]... B2500 | 0.5% | 0.2% | |
Ps | P0.1[3]... P1.7 ( P1.5 ) | 1.0% | 0.5% | |
1 kHz [4] | Hm | 5 A/m...800 A/m | 1.0% | 0.3% |
Jm / Bm | B5 [2]... B800 | 1.0% | 0.3% | |
Ps | P0.1[3]... P1.5 ( P1.0 ) | 2.0% | 0.5% |
Conditions :(23 ± 5)℃, demagnetized before measurement.
Accuracy: 1%~2% deviation will occur according to the way and position of loading the sample in the measuring frame.
Repeatability: put the same specimens in the specified position and keep it from moving, and measure it repeatedly for 10 times. Take the maximum or minimum value of 10 measurements minus the average value divided by the average value, and the maximum value of the two represents repeatability.
Note: [1] The table only gives reference measuring points, and the indexes such as Hm, Jm/Bm, Ps that can be reached by the specimens during actual measure are related to the performance of the material, quality of the specimens, frequency, current, voltage, etc.
Note: [2] B5 refers to the magnetic flux density value when the magnetic field H is 5 A/m, and the typical measuring points of non-oriented steel are in parentheses.
Note: [3] P0.1 refers to the specific total loss value when flux B is 0.1T, and the typical measure point for non-oriented steel is in parentheses.
Note: [4] For medium and high frequency measurment, select the corresponding Epstein frame.
Note: [5] The uncertainty and repeatability are at typical measuring points, and are related to the performance of the measured material.
Output Capability | Voltage | 0.1 Vrms ~ 200 Vrms | |
Current | 0.5 mApk ~ 15 Apk | ||
Frequency | 45 Hz .. .65 Hz ... 400 Hz ... 1 kHz (optional) | ||
Power | 0~500 VA | ||
Specification | Frequency (Hz) | Uncertainty of Effective Value of Voltage and Current(k=2) | Uncertainty of Active Power(k=2) |
45 ≤ F ≤ 65 | 0.05%*RG [6] | 0.1%*FS [7] | |
65 < F ≤ 400 | 0.05%*RG [6] | 0.15%*FS [7] | |
400 < F ≤ 1 k | 0.1%*RG [6] | 0.2%*FS [7] |
Note: [6] RG is range;
Note: [7] FS= Voltage range x current range, when the phase Angle is within 75 degrees.
Power Supply | AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz |
Temperature Performance | Working Temperature:0°C~40°C; Storage Temperature:-20°C~70°C |
Humidity Performance | Working Humidity:40%~80% R·H,No condensation Store Humidity:< 95% R·H,No condensation |
8. Epstein Frame (optional)
The built-in coil of air flux compensation conforms to IEC60404-2 or IEC60404-10 standards.
The cutting and placement of specimens shall comply with IEC60404-2.
Model | TS7000-A | TS7010 |
Length | 25 cm | 25 cm |
Turns | 700 T | 200 T |
Lumen Height | 10 mm | 5 mm |
Magnetic Field Strength | 1 A/m~10 kA/m | 1 A/m~800 A/m |
Frequency | ≤ 400Hz | 400 Hz~10 kHz |
Specimens Dimension | 0.2 kg~1 kg | 0.1 kg~0.5 kg |
Specimens Quality | 30 mm (W) × ( 280~320 ) mm (L) | |
Working Temperature | 0℃~45℃ |
Note: The user shall select the corresponding Epstein frame according to the test frequency and the maximum magnetic field.
Model | TS7500 |
Magnetic Field Strength | 1 A/m~10 kA/m |
Frequency | ≤ 100Hz |
Measurement Method | M.C |
Turns | 400 T |
Specimens Size | 500 mm (W) ×500 mm(L) |
Specimens Thickness | 0.1 mm~0.5 mm |
Working Temperature | 0℃~45℃ |
10. System Configuration List
No. | Item Name | Number | Configuration | Note |
1 | TS1100 | 1 | standard | |
2 | TS7000-A Epstein Frame | 1 | standard | 700 T,≤ 400Hz,1 A/m~10 kA/m |
3 | Software | 1 | standard | |
4 | Test Wires and Attachments | 1 | standard | |
5 | TS7010 Epstein Frame | 1 | Optional | 200 T, 400 Hz~10 kHz |
6 | TS7500 Single Sheet Tester | 1 | Optional | 500 mm × 500 mm, MC |
7 | Workbench | 1 | Optional | Third party Products |
8 | Sartorius Electronic Balance | 1 | Optional | 0.1~3.2kg, Precision: 0.1g |
9 | Computer | 1 | Optional | Third party Products |
10 | Printer | 1 | Optional | Third party Products |
Note: The above is only for reference. The specific configuration list depends on the technical protocol.
11. Ordering Information
Advanced Multifunction Calibrator Advanced Electrical Reference Standards Electric Vehicle (EV)Charging Facility Tester AC/DC Standard Source /Meter Single-phase/Three-phase Standard Meter Magnetic Parameters Measurement and Calibration Magnetic Material Testing Equipment Electrical Steel Testing Equipment