1. Summary
TI1100 is an AC/DC current source with high stability and precision, it is applied for calibrating ammeters on-site, e.g, industrial production lines, etc.
2. Features
• AC/DC current output: 10 μA ~ 10.5 A
• Standard version (TI1100-B) and advanced version (TI1100-C) are available
• Typical specifications of the standard version (TI1100-B):
Frequency: DC, AC 10 Hz ~ 2 kHz.
Typical short-term stability: 0.01%/min
Optimum Accuracy: DC 0.02, AC 0.05.
• Typical specifications of the advanced version (TI1100-C):
Frequency: DC, AC 10 Hz ~ 10 kHz
Typical short-term stability: 0.005%/min
Optimum Accuracy: DC 0.01, AC 0.02.
• Shortcut output keys.
• Programmable calibration schemes.
• RS232, LAN and USB communication interfaces.
3. Application
☆ Calibration for Current Measuring Instruments |
|
• Suitable for calibrating AC/DC ammeters or current measurement of multimeter. |
☆ Convenient Operation |
|
• The front panel has 10 shortcut keys that can output the corresponding value on the screen with one key. |
☆ Calibration Scheme Setting |
• Multiple calibration schemes can be edited and stored for the calibration of multimeter with different specifications. |
4. Specifications (TI1100-B)
4.1 DC Current Output
Range | Current Output | Resolution | Short term stability (%/min) | Accuracy ±(%*output +μA) | Compliance Voltage (V) |
300 μA | ±(10.0000 μA ~ 330.0000 μA) | 100 pA | 0.02 | 0.012 + 0.03 | 11 V |
3 mA | ±(0.300000 mA ~ 3.300000 mA) | 1nA | 0.01 | 0.012 + 0.24 | 11 V |
30 mA | ±(3.00000 mA ~ 33.00000 mA) | 10 nA | 0.01 | 0.012 + 2.4 | 9 V |
300 mA | ±(30.0000 mA ~ 330.0000 mA) | 100 nA | 0.01 | 0.012 + 24 | 7 V |
1A | ±(0.100000 A ~ 1.100000 A) | 1μA | 0.01 | 0.012 + 80 | 6 V |
10 A | ±(1.00000 A ~ 10.50000 A) | 10 μA | 0.01 | 0.012 + 800 | 4 V |
• Output range: ±(10 μA ~ 10.5 A).
• 7 digits display.
• Adjustment fineness: 0.002% * RG.
• Ripple coefficient:<0.1%.
• Protection function: open circuit protection, overload protection, overheating protection.
4.2 AC Current Output
Range | Current Output | Frequency | Resolution | Accuracy ± (%*output +μA) | Compliance Voltage (rms) |
300 μA | 10.0000 μA ~ 330.0000 μA | 10 Hz ~ 45 Hz | 100 pA | 0.05 + 0.2 | 7 V |
45 Hz ~ 400 Hz | 0.03 + 0.1 | ||||
400 Hz ~ 2 kHz | 0.04 + 0.2 | ||||
3 mA | 0.300000 mA ~ 3.300000 mA | 10 Hz ~ 45 Hz | 1nA | 0.05 + 0.5 | 7 V |
45 Hz ~ 400 Hz | 0.03 + 0.3 | ||||
400 Hz ~ 2 kHz | 0.04 + 0.5 | ||||
30 mA | 3.00000 mA ~ 33.00000 mA | 10 Hz ~ 45 Hz | 10 nA | 0.05 + 5 | 6 V |
45 Hz ~ 400 Hz | 0.03 + 3 | ||||
400 Hz ~ 2 kHz | 0.04 + 5 | ||||
300 mA | 30.0000 mA ~ 330.0000 mA | 10 Hz ~ 45 Hz | 100 nA | 0.05 + 50 | 5 V |
45 Hz ~ 400 Hz | 0.03 + 30 | ||||
400 Hz ~ 2 kHz | 0.04 + 50 | ||||
1A | 0.100000 A ~ 1.100000 A | 10 Hz ~ 45 Hz | 1μA | 0.05 + 150 | 4 V |
45 Hz ~ 400 Hz | 0.03 + 100 | ||||
400 Hz ~ 2 kHz | 0.04 + 150 | ||||
10 A | 1.00000 A ~ 10.50000 A | 45 Hz ~ 400 Hz | 10 μA | 0.03 + 1000 | 2.5 V |
400 Hz ~ 2 kHz | 0.04 + 1500 |
• Output range: ± (10 μ A ~ 10.5 A).
• 7 digits display.
• Adjustment fineness: 0.002% * RG.
• Distortion: <0.3%.
• Protection function: open circuit protection, overload protection, overheating protection.
4.3 Frequency
Range | Resolution | Accuracy (± %*RD) |
10.00000 Hz ≤ F ≤ 99.99999 Hz | 0.01 mHz | 0.01 |
100.0000 Hz ≤ F ≤ 999.9999 Hz | 0.1 mHz | 0.01 |
1.000000 kHz ≤ F ≤ 2.000000 kHz | 1 mHz | 0.01 |
5. Specifications (TI1100-C)
5.1 DC Current Output
Range | Current Output | Resolution | 24-hour Stability (Tcal ± 1°C) | Accuracy 1year(Tcal ± 5°C) | Compliance Voltage |
± (ppm*output + μA) | |||||
100 μA | ±(10.0000 μA ~ 110.0000 μA) | 100 pA | 50 + 0.01 | 100 + 0.02 | 10 V |
1 mA | ±(0.100000 mA ~ 1.100000 mA) | 1nA | 40 + 0.02 | 80 + 0.04 | 10 V |
10 mA | ±(1.00000 mA ~ 11.00000 mA) | 10 nA | 30 + 0.2 | 60 + 0.4 | 9 V |
100 mA | ±(10.0000 mA ~ 110.0000 mA) | 100 nA | 30 + 2 | 60 + 4 | 7 V |
1 A | ±(0.100000 A ~ 1.100000 A) | 1 μA | 30 + 20 | 60 + 40 | 4 V |
10 A | ±(1.00000 A ~ 10.50000 A) | 10 μA | 50 + 300 | 100 + 500 | 4 V |
• Output range: ± (10 μA ~ 10.5 A).
• 7 digits display.
• Adjustment fineness: 0.002% * RG.
• Ripple coefficient: <0.1%.
• Protection function: open circuit protection, overload protection, overheating protection
5.2 AC Current Output
Range | Current Output | Frequency | Resolution | Accuracy ± (%*output +μA) | Compliance Voltage (rms) |
100 μA | 10.0000 μA ~ 110.0000 μA | 10 Hz ~ 45 Hz | 100 pA | 800 + 0.04 | 6.5V |
45 Hz ~ 2 kHz | 600 + 0.04 | ||||
2 kHz ~ 10 kHz | 800 + 0.1 | ||||
1 mA | 0.100000 mA ~ 1.100000 mA | 10 Hz ~ 45 Hz | 1 nA | 300 + 0.1 | 6.5V |
45 Hz ~ 2 kHz | 200 + 0.1 | ||||
2 kHz ~ 10 kHz | 300 + 0.2 | ||||
10 mA | 1.00000 mA ~ 11.00000 mA | 10 Hz ~ 45 Hz | 10 nA | 200 + 0.8 | 6 V |
45 Hz ~ 2 kHz | 120 + 0.8 | ||||
2 kHz ~ 10 kHz | 300 + 2 | ||||
100 mA | 10.0000 mA ~ 110.0000 mA | 10 Hz ~ 45 Hz | 100 nA | 200 + 8 | 5 V |
45 Hz ~ 2 kHz | 120 + 8 | ||||
2 kHz ~ 10 kHz | 300 + 20 | ||||
1 A | 0.100000 A ~ 1.100000 A | 10 Hz ~ 45 Hz | 1μA | 200 + 80 | 3.5 V |
45 Hz ~ 2 kHz | 120 + 80 | ||||
2 kHz ~ 10 kHz | 300 + 200 | ||||
10 A | 1.00000 A ~ 10.50000 A | 10 Hz ~ 45 Hz | 10 μA | 200 + 1000 | 2.5 V |
45 Hz ~ 2 kHz | 120 + 1000 | ||||
2 kHz ~ 5 kHz | 300 + 2000 |
• Output range: ± (10 μA ~ 10.5 A).
• 7 digits display.
• Adjustment fineness: 0.002% * RG.
• Distortion:<0.5%
• Protection function: open circuit protection, overload protection, overheating protection.
5.3 Frequency
Range | Resolution | Accuracy (± %*RD) |
10.00000 Hz ≤ F ≤ 99.99999 Hz | 0.01 mHz | 0.01 |
100.0000 Hz ≤ F ≤ 999.9999 Hz | 0.1 mHz | 0.01 |
1.000000 kHz ≤ F ≤ 9.999999 kHz | 1 mHz | 0.01 |
10.00000 kHz | 10 mHz | 0.01 |
6. General Specifications
Power Supply | AC (220±22) V,(50±2) Hz |
Maximum Power Consumption | 100 VA |
Warm Up Time | Twice the time since last warmed up, to a maximum of 30 minutes. |
Work Environment | 0 °C~55 °C, 30% R·H~80% R·H, non-condensing |
Storage Environment | -30 °C~70 °C, 10% R·H~90% R·H, non-condensing |
Communication Interface | USB×1、LAN×1、RS232×1 |
Dimensions | 215 mm (Width) x 252 mm (Depth) x 88 mm (Height), excluding feet. |
Display | 3.5-inch color LCD |
Weight | About 5 kg |
| |
| |
| |
Advanced Multifunction Calibrator Advanced Electrical Reference Standards Electric Vehicle (EV)Charging Facility Tester AC/DC Standard Source /Meter Single-phase/Three-phase Standard Meter Magnetic Parameters Measurement and Calibration Magnetic Material Testing Equipment Electrical Steel Testing Equipment