1. Summary
TI1000 is an AC\DC voltage source with high stability and precision. it is applied for calibrating digital voltmeters\multimeters on-site, e.g, industrial production lines, etc.
2. Features
• ACV\DCV Output: 10 mV ~ 1020 V
• Standard Version(TI1000-B) and Advanced Version(TI1000-C) are available
• Typical Specifications of the Standard Version(TI1000-B):
Frequency: DC, AC 45 Hz ~ 2 kHz.
Optimum Accuracy: DC 100 ppm, AC 200 ppm.
• Typical Specifications of the Advanced Version(TI1000-C):
Frequency: DC, AC 10 Hz ~ 20 kHz
Optimum Accuracy: DC 35ppm, AC 100ppm.
• The Maximum Output Power of AC and DC Voltage is up to 10 VA
• Shortcut Output Keys
• Programmable Calibration Schemes
• RS232, LAN and USB Communication Interfaces
3. Application
☆ Calibrate Voltage Measuring Instruments |
|
• Suitable for Calibrating AC/DC Voltmeters or Voltage Measurement of Multimeter. |
☆ Convenient Operation |
|
• The front panel has 10 shortcut keys that can output the corresponding value on the screen with one key. |
☆ Calibration Scheme Setting |
• Multiple calibration schemes can be edited and stored for the calibration of multimeter with different specifications. |
4. Specifications(TI1000-B)
4.1 DC Voltage
Range | Voltage Output | Resolution | Short-term Stability (%/min) | Accuracy ± (%*RD+mV) [1] | Maximum load [source internal resistance] |
300 mV | ± (10.0000 mV ~ 330.0000 mV) | 100 nV | 0.005 | 0.008 + 0.006 | [60 Ω] |
3 V | ± (0.300000 V ~ 3.300000 V) | 1 μV | 0.005 | 0.008 + 0.06 | 20 mA |
30 V | ± (3.00000 V ~ 33.00000 V) | 10 μV | 0.005 | 0.008 + 0.6 | 15 mA |
300 V | ± (30.0000 V ~ 330.0000 V) | 100 μV | 0.005 | 0.008 + 6 | 10 mA |
1000 V | ± (300.000 V ~ 1020.000 V) | 1 mV | 0.005 | 0.008 + 20 | 10 mA |
Note[1]: RD is the reading value, same below. |
• Output Range:± (10 mV ~ 1020 V)
• 7 Digits Display
• Adjustment Fineness: 0.002%*RG
• Ripple Factor: < 0.1%
• Protection Function: short-circuit protection, overload protection and over-heat protection
4.2 AC Voltage
Range | Voltage Output | Frequency | Resolution | Accuracy ± (%*RD+mV) [1] | Maximum Load [Source Internal Resistance] |
300 mV | 10.0000 mV ~ 330.0000 mV | 45 Hz ~ 400 Hz | 100 nV | 0.02 + 0.01 | [60 Ω] |
400 Hz ~ 2 kHz | 0.03 + 0.01 | ||||
3 V | 0.300000 V ~ 3.300000 V | 45 Hz ~ 400 Hz | 1 μV | 0.02 + 0.1 | 20 mA |
400 Hz ~ 2 kHz | 0.03 + 0.1 | ||||
30 V | 3.00000 V ~ 33.00000 V | 45 Hz ~ 400 Hz | 10 μV | 0.02 + 1 | 15 mA |
400 Hz ~ 2 kHz | 0.03 + 1 | ||||
300 V | 30.0000 V ~ 330.0000 V | 45 Hz ~ 400 Hz | 100 μV | 0.02 + 10 | 10 mA |
400 Hz ~ 2 kHz | 0.03 + 10 | ||||
1000 V | 300.000 V ~ 1020.000 V | 45 Hz ~ 400 Hz | 1 mV | 0.02 + 30 | 10 mA |
400 Hz ~ 1 kHz | 0.03 + 30 |
• Output Range: 10 mV ~ 1020 V
• 7 Digits Display
• Adjustment Fineness: 0.002%*RG
• Distortion: < 0.3%
• Protection Function: short-circuit protection, overload protection and over-heat protection
4.3 Sinusoidal Wave Frequency
Range | Resolution | Accuracy ±(%*RD) |
45.00000 Hz ≤ F ≤ 99.99999 Hz | 10 μHz | 0.01 |
100.0000 Hz ≤ F ≤ 999.9999 Hz | 100 μHz | 0.01 |
1.000000 kHz ≤ F ≤ 2.000000 kHz | 1 mHz | 0.01 |
5. Specifications(TI1000-C)
5.1 DC Voltage
Range | Voltage Output | Resolution | 24 Hour Stability (Tcal±1°C) | Accuracy 1 Year (Tcal±5°C) | Maximum Load [Source Internal Resistance] |
±(ppm*RD+mV) | |||||
100 mV | ±(10.0000 mV ~ 110.0000 mV) | 100 nV | 13 + 0.003 | 25 + 0.004 | [60 Ω] |
1 V | ±(0.100000 V ~ 1.100000 V) | 1 μV | 13 + 0.005 | 25 + 0.01 | 20 mA |
10 V | ±(1.00000 V ~ 11.00000 V) | 10 μV | 13 + 0.05 | 25 + 0.1 | 15 mA |
100 V | (10.0000 V ~ 110.0000 V) | 100 μV | 13 + 0.5 | 25 + 1 | 10 mA |
1000 V | (100.000 V ~ 1020.000 V) | 1 mV | 13 + 6 | 25 + 10 | 10 mA |
• Output Range: ± (10 mV ~ 11 V),11 V ~ 1020 V
• 7 Digits Display
• Adjustment Fineness: 0.002%*RG
• Ripple Factor: < 0.1%
• Protection Function: short-circuit protection, overload protection and over-heat protection
5.2 AC Voltage
Range | Voltage Output | Frequency | Resolution | Accuracy 1 Year (Tcal±5°C) ± (ppm*RD+mV) | Maximum Load [Source Internal Resistance] |
100 mV | 10.0000 mV ~ 110.0000 mV | 10 Hz ~ 45 Hz | 100 nV | 100 + 0.02 | [60 Ω] |
45 Hz ~ 2 kHz | 60 + 0.02 | ||||
2 kHz ~ 20 kHz | 80 + 0.02 | ||||
1 V | 0.100000 V ~ 1.100000 V | 10 Hz ~ 45 Hz | 1μV | 100 + 0.04 | 20 mA |
45 Hz ~ 2 kHz | 60 + 0.04 | ||||
2 kHz ~ 20 kHz | 80 + 0.04 | ||||
10 V | 1.00000 V ~ 11.00000 V | 10 Hz ~ 45 Hz | 10 μV | 100 + 0.4 | 15 mA |
45 Hz ~ 2 kHz | 60 + 0.4 | ||||
2 kHz ~ 20 kHz | 80 + 0.4 | ||||
100 V | 10.0000 V ~ 110.0000 V | 45 Hz ~ 2 kHz | 100 μV | 60 + 4 | 10 mA |
1000 V | 100.000 V ~ 1020.000 V | 45 Hz ~ 2 kHz | 1 mV | 60 + 40 | 10 mA |
• Output Range: 10 mV ~ 1020 V
• 7 Digits Display
• Adjustment Fineness: 0.002%*RG
• Distortion: < 0.3%
• Protection Function: short-circuit protection, overload protection and over-heat protection
5.3 Sinusoidal Wave Frequency
Range [1] | Resolution | Accuracy 1 Year (Tcal±5°C) (%*RD) |
10.00000 Hz ≤ F ≤ 99.99999 Hz | 10 μHz | 0.01 |
100.0000 Hz ≤ F ≤ 999.9999 Hz | 100 μHz | 0.01 |
1.000000 kHz ≤ F ≤ 9.999999 kHz | 1 mHz | 0.01 |
10.00000 kHz ≤ F ≤ 20.00000 kHz | 10 mHz | 0.01 |
6. General Specifications
Power Supply | AC (220±22) V, (50±2) Hz |
Maximum Power Consumption | 100 VA |
Warm Up Time | Twice the time since last warmed up, to a maximum of 30 minutes. |
Operating Environment | 0°C ~ 55°C, 30%R·H ~ 80%R·H, non-condensing |
Storage Environment | -30°C ~ 70°C, 10%R·H ~ 90%R·H, non-condensing |
Communication Interface | USB×1, LAN×1, RS232×1 |
Dimensions | 215 mm (W) × 252 mm (D) ×88 mm(H), feets excluded. |
Display | 3.5-inch color LCD |
Weight | About 5 kg |
Advanced Multifunction Calibrator Advanced Electrical Reference Standards Electric Vehicle (EV)Charging Facility Tester AC/DC Standard Source /Meter Single-phase/Three-phase Standard Meter Magnetic Parameters Measurement and Calibration Magnetic Material Testing Equipment Electrical Steel Testing Equipment