1. Summary
TH1200 is an instrument that uses the standard source method to calibrate the nanovoltmeter, and is also suitable for calibrating the voltage-measurement function of DC shunt verification device, small signal voltage-measurement function of DC energy reference standard and other equipment.
2. Features
• Standard voltage output:1 μV~110 V
• Uncertainty of measurement 6 ppm@10 V.
• Typical stability at mV output:10 nV/min.
• 8-digital,minimun resolution:1 nV.
• LEMO interface for 2-channel voltage output.
• Constant temperature control technology to reduce the environmental fator
• Specific test wire
• Specific automatic calibration software(optional)
3. Application
☆ Standard source method to calibrate the nanovoltmeter |
|
• Calibrate 2-chanel voltage simultaneously:1 μV~110 V(CH1)& 1 mV~11 V(CH2)。 • Meet Calibration requirements of nanovoltmeter(such as Agilent 34420A、Keithley 2182A) |
☆ Interface and Automatic Calibration Software |
• support RS232、LAN、USB,and accept customer requirements for software function customization。 |
4. Specifications
Range | Resolution | Short Term Stability | Accuracy (k=2) | Temperature Coefficient (/°C) | Maximum Burden Current | ||
24H Tcal±1°C | 24H Tcal±5°C | 90 day Tcal±5°C | 1year Tcal±5°C | 0°C~18°C 28°C~50°C | |||
ppm of reading + ppm of range [1] | mA | ||||||
CH1 | |||||||
1 mV | 1 nV | 10 + 10 | 10 + 10 | 13 + 10 | 13 + 10 | 2 + 0.5 | 50 |
10 mV | 1 nV | 10 + 2 | 10 + 2 | 13 + 2 | 15 + 2 | 1 + 0.5 | 50 |
100 mV | 10 nV | 3 + 2 | 3 + 2 | 6 + 2 | 8 + 2 | 1 + 0.2 | 50 |
1 V | 100 nV | 2 + 1 | 2 + 2 | 4 + 2 | 6 + 2 | 1 + 0.1 | 50 |
10 V | 1 μV | 1 + 0.5 | 1 + 1 | 3 + 2 | 4 + 2 | 1 + 0.1 | 50 |
100 V | 10 μV | 2 + 1 | 3 + 2 | 4 + 2 | 6 + 2 | 1 + 0.5 | 10 |
CH2 | |||||||
100 mV | 10 nV | 3 + 2 | 3 + 2 | 6 + 2 | 8 + 2 | 1 + 0.2 | 50 |
1 V | 100 nV | 2 + 1 | 2 + 2 | 4 + 2 | 6 + 2 | 1 + 0.1 | 50 |
10 V | 1 μV | 1 + 0.5 | 1 + 1 | 3 + 2 | 4 + 2 | 1 + 0.1 | 50 |
Note [1] : (ppm = parts per million) (e.g: 10ppm = 0.001%). |
• Output range:±(10%~110%)*RANGE
• range change:manual/auto
• 8-digital
• Stable time:< 3 s to total uncertainty,+1 s in case of range or polarity changing
5. General Specifications
Power Supply | AC 220 V / 50 Hz |
Working Environment | Temperature:20±5°C humidity:30%R·H~80%R·H,non-condensing。 others:No electromagnetic interference |
Storage Environment | Temperature:0°C~50°C humidity:10%R·H~90%R·H,non-condensing。 |
Interface | RS232*1、LAN*1、USB*1 |
Advanced Multifunction Calibrator Advanced Electrical Reference Standards Electric Vehicle (EV)Charging Facility Tester AC/DC Standard Source /Meter Single-phase/Three-phase Standard Meter Magnetic Parameters Measurement and Calibration Magnetic Material Testing Equipment Electrical Steel Testing Equipment