TH0690 is an intelligent instrument dedicated to the comprehensive performance evaluation of current transducers. It can accurately test the electrical characteristics such as ratio error (k) and phase displacement (δ) of I/I and I/V current transformers, and can also measure the AC magnetic characteristics of the transformer core. It solves the problem of complex calculation of the relationship between magnetic properties and electrical properties in the process of transducers design, and provides the most complete means for the analysis and measurement of current transformers.
AC Power Output/Measurement Accuracy: Class 0.005
Working Frequency Range: 45 Hz~65 Hz.
Support multi-frequency point measurement of ratio error (k) and phase displacement (δ)
Draw frequency or current characteristic curve with the software。
Support various magnetic characteristic parameter measurement and curve drawing.
With automatic demagnetization function, it can evaluate the magnetic resistance of the transformer.
Special Test Software (optional)
Current transformer ratio error (k) and phase displacement (δ) calibration.
Magnetic performance testing of finished transformers and semi-finished iron cores.
Current transformer quality inspection, design verification, iron core pairing.
Ratio error (k) and phase displacement (δ) measurement of the current measurement circuit.
Anti-magnetic evaluation of current transformers.
☆ Current transformer ratio error and phase displacement calibration. |
|
Measure and calibrate the ratio error and phase displacement of the current transformer. Draw the frequency characteristic curve k(F)、δ(F),current characteristic curve k(I)、δ(I). |
☆ Measuring the magnetic properties of core semi-finished products. |
|
It can measure the magnetic properties of finished transformers and semi-finished iron cores。 Parameter Measurement: μi、μm、Hc、Br、Bs、μa、Bm、Hm、δ etc. |
Current Range | Short-term Stability (%/min) | Measurement Uncertainty (k=2, ppm*RD+ppm*RG)① | Maximum Load Voltage(V) | Distortion (%) | ||
24 h (23±1)°C | 90 days (23±5)°C | 1 year (23±5)°C | ||||
2 mA | 0.003 | 20 + 20 | 25 + 20 | 30 + 20 | 1.2 | < 0.2 |
5 mA | 0.003 | 20 + 20 | 25 + 20 | 30 + 20 | 1.2 | < 0.2 |
10 mA | 0.003 | 20 + 20 | 25 + 20 | 30 + 20 | 1.2 | < 0.2 |
20 mA | 0.003 | 20 + 20 | 25 + 20 | 30 + 20 | 1.2 | < 0.2 |
50 mA | 0.003 | 20 + 20 | 25 + 20 | 30 + 20 | 1.2 | < 0.2 |
100 mA | 0.003 | 20 + 20 | 25 + 20 | 30 + 20 | 1.2 | < 0.2 |
200 mA | 0.003 | 20 + 20 | 25 + 20 | 30 + 20 | 1.2 | < 0.2 |
500 mA | 0.003 | 20 + 20 | 25 + 20 | 30 + 20 | 1.2 | < 0.2 |
1 A | 0.003 | 20 + 20 | 25 + 20 | 30 + 20 | 1.2 | < 0.2 |
2 A | 0.003 | 20 + 20 | 25 + 20 | 30 + 20 | 1.2 | < 0.2 |
5 A | 0.003 | 20 + 20 | 25 + 20 | 30 + 20 | 1.2 | < 0.2 |
10 A | 0.003 | 20 + 20 | 25 + 20 | 30 + 20 | 1.2 | < 0.2 |
20 A | 0.003 | 20 + 20 | 25 + 20 | 30 + 20 | 1.2 | < 0.2 |
50 A | 0.003 | 20 + 20 | 25 + 20 | 30 + 20 | 1.2 | < 0.2 |
100 A | 0.003 | 20 + 20 | 25 + 20 | 30 + 20 | 1.2 | < 0.2 |
Note: ① RD is the reading value, RG is the range value, the same below. |
ACI Output:200 μA~120 A,7-digit display.
Frequency range: 45 Hz~65 Hz; Adjustment Fineness: 0.0001 Hz; Measurement Uncertainty: 0.001 Hz
Protection Function: short circuit protection, overload protection.
AC | Range | Resolution | Measurement Uncertainty (k=2, ppm*RD +ppm*RG) | ||
24 h (23±1)°C | 90 days (23±5)°C | 1 year (23±5)°C | |||
ACV/U2 | 20 mV | 10 nV | 40 + 40 | 50 + 40 | 60 + 40 |
50 mV | 10 nV | 20 + 20 | 25 + 20 | 30 + 20 | |
100 mV | 100 nV | 20 + 20 | 25 + 20 | 30 + 20 | |
200 mV | 100 nV | 20 + 20 | 25 + 20 | 30 + 20 | |
500 mV | 100 nV | 20 + 20 | 25 + 20 | 30 + 20 | |
1 V | 1 μV | 20 + 20 | 25 + 20 | 30 + 20 | |
2 V | 1 μV | 20 + 20 | 25 + 20 | 30 + 20 | |
5 V | 1 μV | 20 + 20 | 25 + 20 | 30 + 20 | |
ACI/I2 | 200 μA | 0.1 nA | 20 + 20 | 25 + 20 | 30 + 20 |
500 μA | 0.1 nA | 20 + 20 | 25 + 20 | 30 + 20 | |
1 mA | 1 nA | 20 + 20 | 25 + 20 | 30 + 20 | |
2 mA | 1 nA | 20 + 20 | 25 + 20 | 30 + 20 | |
5 mA | 1 nA | 20 + 20 | 25 + 20 | 30 + 20 | |
10 mA | 10 nA | 20 + 20 | 25 + 20 | 30 + 20 | |
20 mA | 10 nA | 20 + 20 | 25 + 20 | 30 + 20 | |
50 mA | 10 nA | 20 + 20 | 25 + 20 | 30 + 20 | |
100 mA | 100 nA | 20 + 20 | 25 + 20 | 30 + 20 |
U2 Measurement Range: 2 mV~5.5 V,7-digit display.
I2 Measurement Range: 20 μA~110 mA,7-digit display.
Harmonic Measurement: 2nd to 63rd.
Ratio Error | Measurement Range | Current transformer input current like I1 Current transformer output current like I2 |
Measurement Uncertainty (k=2) | 0.01% @ 45 Hz ~ 65 Hz | |
Phase Displacement | Measurement Range | 0.000°~359.999° |
Measurement Uncertainty (k=2) | 0.001° @ 45 Hz~65 Hz |
Measured Parameters | Measurement Uncertainty (k=2) | Measurement Repeatability |
μi | 1.5% | 4.0% |
μm | 1.0% | 2.0% |
Hc | 0.5% | 1.0% |
Br | 0.3% | 1.0% |
Bs | 0.3% | 1.0% |
μa | 1.0% | 3.0% |
Bm | 0.5% | 1.0% |
Hm | 1.0% | 2.0% |
δ | 1.5% | 5.0% |
Hc | 1.0% | 3.0% |
Br | 1.0% | 3.0% |
μ’ | 1.0% | 3.0% |
μ′′ | 1.0% | 3.0% |
μL | 1.0% | 3.0% |
Note: The iron core should be demagnetized before the test, and the parameters of the iron core should be input as required. The measurement temperature T: (23±5)°C. |
Advanced Multifunction Calibrator Advanced Electrical Reference Standards Electric Vehicle (EV)Charging Facility Tester AC/DC Standard Source /Meter Single-phase/Three-phase Standard Meter Magnetic Parameters Measurement and Calibration Magnetic Material Testing Equipment Electrical Steel Testing Equipment