1. Summary
TH0410 is a set of nine precision AC/DC shunts of different ranges that convert a nominal input current of 20 mA ~ 10 A to 1V Small voltage, easy to access voltage high-speed sampling system or high-precision data meter, to achieve wideband or complex waveform current measurement, or used to calibrate wideband current sources, transformers, sensors, etc.
2. Features
• Operating frequency: DC ~ 1 MHz
• AC and DC difference: ≤ 30 ppm @53Hz
• Optimal phase displacement: 5 μrad
• The nominal output voltage is 1 V (other voltage values can be customized).
• The typical annual variation is typically 5 ppm and the maximum is 18 ppm.
• The 1 mA, 10 mA versions mate buffers to reduce output impedance.
3. Applications
☆ Build Wideband Current Measurement System |
|
TH0410 can be used with wideband current source, wideband standard voltmeter (such as TH2000 vector voltage analyzer), etc. to form a wideband current measurement system, which is used to measure wideband current signals, calibrate wideband current sources, wideband ammeters, etc. |
☆ Set up wideband power precision measurement system |
|
TH0410 can be used with high-frequency V/V conversion standard and TH2000 vector voltage analyzer to form a wideband power/energy precision measurement system for measuring wideband power/ Power signal, calibration of wideband power source, wideband power meter, power analyzer and other equipment. |
☆ Build a wideband instrument transformer / sensor measurement system |
|
For the calibration of the transformer or sensor under test with a nominal current of ≤ 10A, TH0410 can be used directly as the standard device, with a widebandconstant current source and a wideband sensor tester ( vector voltage analyzer), etc. to form a wideband transformer/sensor measurement system. |
☆ Build a wideband transformer/sensor calibration system |
|
For calibration of transformers or sensors with nominal currents greater than 10 A, TH0410 is used as a secondary signal I/V converter for standard and tested equipment, with wideband constant current source, wideband sensor testers (vector voltage analyzers), etc. form wideband transformers/sensor measurement systems. |
☆ Set up sensor bandwidth calibration system |
|
TH0410 can be used with wideband standard voltmeters (such as TH2000 vector voltage analyzer) to form a sensor bandwidth calibration system. |
4. Specifications
4.1 Resistance
Specifications | Nominal resistance ( Z ) | Initial deviation [1] ( ±mH/Z ) @ (23±1)°C | Annual stability ( ±mH/Z ) | Temperature coefficient ( ±ppm/K ) | Power factor [2] ( ±ppm/rated power ) |
TH0410-20mA | 50 | 50 | 18 | 2.5 | 1 |
TH0410-50mA | 20 | 50 | 18 | 2.5 | 1 |
TH0410-100mA | 10 | 50 | 18 | 2.5 | 2 |
TH0410-200mA | 5 | 50 | 18 | 2.5 | 4 |
TH0410-500mA | 2 | 50 | 18 | 2.5 | 12 |
TH0410-1A | 1 | 50 | 18 | 4.0 | 25 |
TH0410-2A | 500 m | 50 | 18 | 4.0 | 25 |
TH0410-5A | 200 m | 50 | 18 | 4.0 | 30 |
TH0410-10A | 100 m | 50 | 18 | 4.0 | 60 |
Note [1]: The initial deviation is the maximum offset of the resistance value under DC conditions, usually the manufacturer's factory warranty; Higher accuracy can be customized according to user requirements, such as 20 μΩ/Ω. Note [2]: When measuring DCI with a coaxial shunt, the resistance value changes due to the heat generation of the resistor itself at different input currents, resulting in an error, and the calculation formula is: |
4.2 Current measurement
Specifications | Primary nominal current (A) | Secondary nominal Voltage (V) | AC/DC difference [3] (±ppm) | Maximum operating Current (A) | Instantaneous current (A, <5 s) | ||
50 Hz | 1 kHz | 1 MHz | |||||
TH0410-20mA | 20 m | 1 | 20 | 25 | 50 | 40m | 250m |
TH0410-50mA | 50 m | 1 | 20 | 25 | 50 | 100m | 500m |
TH0410-100mA | 100 m | 1 | 20 | 25 | 25 | 200m | 1.5 |
TH0410-200mA | 200 m | 1 | 20 | 25 | 25 | 500m | 2 |
TH0410-500mA | 500 m | 1 | 20 | 25 | 25 | 1 | 3 |
TH0410-1A | 1 | 1 | 20 | 25 | 30 | 1.5 | 4 |
TH0410-2A | 2 | 1 | 20 | 30 | 50 | 2.5 | 6 |
TH0410-5A | 5 | 1 | 20 | 30 | 80 | 5.5 | 20 |
TH0410-10A | 10 | 1 | 20 | 30 | 100 | 11 | 30 |
Note [3]: The measured value of AC current should be calculated according to the following formula, and the unit of AC and DC difference is ppm: |
4.3 Phase displacement
Specifications | Phase displacement (±μrad) | ||||||
53 Hz | 400 Hz | 1 kHz | 10 kHz | 50 kHz | 100 kHz | 1 MHz | |
TH0410-20mA | 5 | 5 | 10 | 100 | 400 | 800 | 800 |
TH0410-50mA | 5 | 5 | 10 | 100 | 400 | 800 | 800 |
TH0410-100mA | 5 | 5 | 10 | 60 | 300 | 500 | 500 |
TH0410-200mA | 5 | 5 | 10 | 60 | 300 | 500 | 500 |
TH0410-500mA | 5 | 5 | 10 | 60 | 300 | 500 | 500 |
TH0410-1A | 5 | 5 | 10 | 60 | 300 | 500 | 500 |
TH0410-2A | 10 | 10 | 15 | 110 | 500 | 1000 | 1000 |
TH0410-5A | 10 | 15 | 25 | 200 | 1000 | 2000 | 2000 |
TH0410-10A | 10 | 20 | 40 | 300 | 1500 | 3000 | 3000 |
5. General Specifications
Power supply | —— |
Temperature performance | Working temperature: 13°C~33°C; Calibration temperature: 18°C~28°C; Storage temperature: 5°C ~ 45°C Note: When stored in harsh temperature environments, it may cause a temporary change in the resistance value and return to the original resistance value within the next 30 days. |
Humidity performance | Operating humidity: < 50% R·H Storage humidity: (15%~80%) R·H, no condensing |
Altitude | < 3000 m |
Connection method | Input connection: Type-N type female plug; Output connection: Type-N type female plug. |
Specifications | Size( mm ) ( W×D×H ) ( without terminal blocks ) | Quality ( kg ) | Dimensional drawing |
TH0410-2 0 mA TH0410-5 0 mA TH0410-100mA | 65 × 68 × 68 | 0.1 | |
TH0410-200mA TH0410-500mA TH0410-1A TH0410-2A | 82 × 94 × 94 | 0.2 | |
TH0410-5A TH0410-10A | 154 × 205 × 205 | 1 | |
6. Ordering Information
| Model | Nominal current |
TH0410-20mA | 20 mA | |
TH0410-50mA | 50 mA | |
TH0410-100mA | 1,00 mA | |
TH0410-200mA | 200 mA | |
TH0410-500mA | 500 mA | |
TH0410-1A | 1 A | |
TH0410-2A | 2 A | |
TH0410-5A | 5 A | |
TH0410-10A | 10 A |
Advanced Multifunction Calibrator Advanced Electrical Reference Standards Electric Vehicle (EV)Charging Facility Tester AC/DC Standard Source /Meter Single-phase/Three-phase Standard Meter Magnetic Parameters Measurement and Calibration Magnetic Material Testing Equipment Electrical Steel Testing Equipment