1. Summary
TA1500 is a series of AC current standard source devices, which can output high stability and high accuracy AC current large current, suitable for calibrating AC current ammeters, current transformers, Current sensors and other devices.
2. Features
• Supports combined outputs from multiple module sources up to 6 kA.
• Test frequency: DC, 10 Hz ~ 65 Hz (400, 1kHz optional).
• Accuracy class:Class 0.02 or Class 0.05
• AC and DC current outputs use the same bus and terminal block
• DC current is bipolar and does not require relay commutation.
• Support external control console to adjust the current output.
• The software controls the slow rise and fall of the large current, and the overshoot amplitude is small.
3. Applications
☆ Calibrate the AC current ammeter |
|
• Class 0.02 specification: Suitable for calibrating large AC-DC ammeter of class 0.1 and below. • Class 0.05 specification: Suitable for calibrating large AC-DC ammeter of class 0.2 and below. |
☆ Calibrate current sensor / transformer |
|
• TA1500 can be used as a primary current input to the sensor or transformer under test. • In combination with a precision ammeter or voltmeter, the sensor or transformer is calibrated using the direct measurement method. |
4. Functional features
☆ Measurement output and adjustment mode |
|
• With the "fixed-point output" method, you can directly set the required output value by clicking the touch screen. • The unit switches fully automatically to the optimal range output. |
☆ Harmonic output (optional). |
|
• Optional harmonic output function, current can be superimposed 2~21 harmonics @ 50 Hz. • The amplitude and phase of each harmonic can be set independently, and the output waveform and spectrogram can be displayed. |
5. Specifications
5.1 AC direct current output
Range | Short-term stability (%/min) | Measurement uncertainty (k=2) (ppm*RD+ppm*RG) [1] | Maximum load Voltage (Vpk). | ||||
Class 0.01 [2] | Class 0.02 | Class 0.05 | Class 0.01 [2] | Class 0.02 | Class 0.05 | ||
10 A | 0.003 | 0.005 | 0.01 | 60 + 40 | 150 + 50 | 400 + 100 | 8 |
30 A | 0.003 | 0.005 | 0.01 | 60 + 40 | 150 + 50 | 400 + 100 | 8 |
100 A | 0.003 | 0.005 | 0.01 | 60 + 40 | 150 + 50 | 400 + 100 | 8 |
300 A | 0.003 | 0.005 | 0.01 | 60 + 40 | 150 + 50 | 400 + 100 | 8 |
600 A | 0.003 | 0.005 | 0.01 | 60 + 40 | 150 + 50 | 400 + 100 | 8 |
1 kA | 0.003 | 0.005 | 0.01 | 60 + 40 | 150 + 50 | 400 + 100 | 8 |
... | ... | ... | ... | ... | ... | ... | ... |
N*1 kA | 0.003 | 0.005 | 0.01 | 60 + 40 | 150 + 50 | 400 + 100 | 8 |
Note: [1] RD is the reading value, RG is the range value; [2] The 0.01 level only supports DC, 45 ~ 65 Hz frequency range. |
• Output range: range below 1 kA (excluding 1 kA): 10%~112%*RG
Range above 1 kA (including 1 kA): 10%~100%*RG (maximum expandable output to 6 kA )
• Fineness adjustment: 5 ppm*RG, 7-digit decimal display
• DC ripple coefficient: ≤ 0.5% @ 5 kHz or less, AC distortion: ≤ 0.5%.
• Settling time: Using the slow rise output mode, the time to stabilize to the output accuracy is about 3 s
• Protection function: open circuit protection, overload protection, overheating protection
5.2 Frequency and Harmonics
Sine wave frequency | Frequency range | 10 Hz ~ 65 Hz(400 Hz、1 k Hz optional) |
Minimum fineness | 0.0001 Hz | |
Measurement uncertainty (k=2). | 0.005% | |
Harmonic output (Optional) | Number of harmonics | 2nd ~ 21st time |
Harmonic amplitude | 0.0% ~ 40.0% adjustable (relative to fundamental). | |
Measurement uncertainty (k=2). | 0.5% |
6. General Specifications
Power supply | AC (380 ± 38) V,(50 ± 2) Hz |
Maximum power consumption | 16 kVA / per1 kA |
Warm-up time | 30 minutes |
Temperature performance | Working temperature: 10°C~35°C Calibration temperature: 18°C~28°C Storage temperature: -20°C~ 70°C |
Humidity performance | Operating humidity: <80% @ 30°C, <70% @ 40°C Storage humidity: (20%~80%) R·H, no condensation |
Altitude | < 3000 m |
Communication interface | RS232 interface |
7. Ordering Information
Advanced Multifunction Calibrator Advanced Electrical Reference Standards Electric Vehicle (EV)Charging Facility Tester AC/DC Standard Source /Meter Single-phase/Three-phase Standard Meter Magnetic Parameters Measurement and Calibration Magnetic Material Testing Equipment Electrical Steel Testing Equipment