1. Summary
TA1300 is a series of AC current standard source devices, which can output high stability and high accuracy AC high current, suitable for calibrating AC ammeters, current transformers, current sensors and other equipment.
2. Features
• Supports combined outputs from multiple module sources up to 6 kA.
• Maximum fundamental frequency: 65 Hz / 400 Hz / 1kHz optional.
• Accuracy: Class 0.02 or Class 0.05
• Harmonic output function (optional)
• Support external control console to adjust the current output.
3. Applications
☆ Calibrate the AC ammeter |
|
• Class 0.02 specification: Suitable for calibrating high power frequency ammeter of class 0.1 and below. • Class 0.05 specification: Suitable for calibrating high power frequency ammeter of class 0.2 and below |
☆ Calibrate current sensor / transformer |
|
• TA1300 can be used as a primary current input to the sensor or transformer under test. • In combination with a precision ammeter or voltmeter, the sensor or transformer is calibrated using the direct measurement method. |
4. Characteristics
☆ Measurement output and adjustment mode |
|
• With the "fixed-point output" method, you can directly set the required output value by clicking the touch screen. • The unit switches fully automatically to the optimal range output. |
☆ Harmonic output (optional). |
|
• Optional harmonic output function, the current can be superimposed 2~11th harmonic @ 50 Hz. • The amplitude and phase of each harmonic can be set independently, and the output waveform and spectrogram can be displayed. |
5. Specifications
5.1 AC current output
Range | Short-term stability (%/min) | Measurement uncertainty (k=2) (ppm*RD+ppm*RG) [2] | Compliance voltage (rms) | Degree of distortion (%) | ||
Class 0.02 | Class 0.05 | Class 0.02 | Class 0.05 | |||
100 A | 0.005 | 0.01 | 120 + 180 | 300 + 200 | 0.7 V | < 0.5 |
300 A | 0.005 | 0.01 | 120 + 180 | 300 + 200 | 0.7 V | < 0.5 |
1 kA | 0.005 | 0.01 | 120 + 180 | 300 + 200 | 0.7 V | < 0.5 |
... | ... | ... | ... | ... | ... | ... |
N*300 A | 0.005 | 0.01 | 120 + 180 | 300 + 200 | 0.7 V | < 0.5 |
Note [1]: RD is the reading value, R G is the range value, the same below. |
• Output range: 0~110%*RG, maximum expandable output to 6 kA
• Fineness: 0.002%*RG, 6-digit decimal display
• Protection function: open circuit protection, overload protection, overheating protection
5.2 Frequency and Harmonics
Sine wave frequency | Frequency range | 45 Hz ~ 65 Hz(400 Hz、1 k Hz optional) |
Minimum fineness | 0.0001 Hz | |
Measurement uncertainty (k=2). | 0.005% | |
Harmonic output (Optional) | Number of harmonics | 2nd ~ 11th |
Harmonic amplitude | 0.0% ~ 10.0% adjustable (relative to fundamental). | |
Measurement uncertainty (k=2). | 0.5% |
6. General Specifications
Power supply | AC (220 ± 22) V,(50 ± 2) Hz |
Maximum power consumption | 1 kA: 2 kVA 2 kA: 4 kVA |
Warm-up time | 30 minutes |
Temperature performance | Working temperature: 0°C~40°C Calibration temperature: 18°C~28°C Storage temperature: -20°C ~ 70°C |
Humidity performance | Operating humidity: <80% @ 30°C, <70% @ 40°C Storage humidity: (20%~80%) R·H, no condensation |
altitude | < 3000 m |
Communication interface | RS232 interface |
7. Ordering Information
Advanced Multifunction Calibrator Advanced Electrical Reference Standards Electric Vehicle (EV)Charging Facility Tester AC/DC Standard Source /Meter Single-phase/Three-phase Standard Meter Magnetic Parameters Measurement and Calibration Magnetic Material Testing Equipment Electrical Steel Testing Equipment