TA1300 is a series of AC current standard source devices, which can output high stability and high accuracy AC high current, suitable for calibrating AC ammeters, current transformers, current sensors and other equipment.
• Supports combined outputs from multiple module sources up to 6 kA.
• Maximum fundamental frequency: 65 Hz / 400 Hz / 1kHz optional.
• Accuracy: Class 0.02 or Class 0.05
• Harmonic output function (optional)
• Support external control console to adjust the current output.
☆ Calibrate the AC ammeter |
|
• Class 0.02 specification: Suitable for calibrating high power frequency ammeter of class 0.1 and below. • Class 0.05 specification: Suitable for calibrating high power frequency ammeter of class 0.2 and below |
☆ Calibrate current sensor / transformer |
|
• TA1300 can be used as a primary current input to the sensor or transformer under test. • In combination with a precision ammeter or voltmeter, the sensor or transformer is calibrated using the direct measurement method. |
☆ Measurement output and adjustment mode |
|
• With the "fixed-point output" method, you can directly set the required output value by clicking the touch screen. • The unit switches fully automatically to the optimal range output. |
☆ Harmonic output (optional). |
|
• Optional harmonic output function, the current can be superimposed 2~11th harmonic @ 50 Hz. • The amplitude and phase of each harmonic can be set independently, and the output waveform and spectrogram can be displayed. |
Range | Short-term stability (%/min) | Measurement uncertainty (k=2) (ppm*RD+ppm*RG) [2] | Compliance voltage (rms) | Degree of distortion (%) | ||
Class 0.02 | Class 0.05 | Class 0.02 | Class 0.05 | |||
100 A | 0.005 | 0.01 | 120 + 180 | 300 + 200 | 0.7 V | < 0.5 |
300 A | 0.005 | 0.01 | 120 + 180 | 300 + 200 | 0.7 V | < 0.5 |
1 kA | 0.005 | 0.01 | 120 + 180 | 300 + 200 | 0.7 V | < 0.5 |
... | ... | ... | ... | ... | ... | ... |
N*300 A | 0.005 | 0.01 | 120 + 180 | 300 + 200 | 0.7 V | < 0.5 |
Note [1]: RD is the reading value, R G is the range value, the same below. |
• Output range: 0~110%*RG, maximum expandable output to 6 kA
• Fineness: 0.002%*RG, 6-digit decimal display
• Protection function: open circuit protection, overload protection, overheating protection
Sine wave frequency | Frequency range | 45 Hz ~ 65 Hz(400 Hz、1 k Hz optional) |
Minimum fineness | 0.0001 Hz | |
Measurement uncertainty (k=2). | 0.005% | |
Harmonic output (Optional) | Number of harmonics | 2nd ~ 11th |
Harmonic amplitude | 0.0% ~ 10.0% adjustable (relative to fundamental). | |
Measurement uncertainty (k=2). | 0.5% |
Power supply | AC (220 ± 22) V,(50 ± 2) Hz |
Maximum power consumption | 1 kA: 2 kVA 2 kA: 4 kVA |
Warm-up time | 30 minutes |
Temperature performance | Working temperature: 0°C~40°C Calibration temperature: 18°C~28°C Storage temperature: -20°C ~ 70°C |
Humidity performance | Operating humidity: <80% @ 30°C, <70% @ 40°C Storage humidity: (20%~80%) R·H, no condensation |
altitude | < 3000 m |
Communication interface | RS232 interface |
Advanced Multifunction Calibrator Advanced Electrical Reference Standards Electric Vehicle (EV)Charging Facility Tester AC/DC Standard Source /Meter Single-phase/Three-phase Standard Meter Magnetic Parameters Measurement and Calibration Magnetic Material Testing Equipment Electrical Steel Testing Equipment