TM9200 Alternating Magnetometer Calibration Device

TM9200 is a high-precision alternating magnetometer calibration device. It consists of a high-precision AC current standard source, a magnetic field coil (Helmholtz coil), an intelligent probe position adjustment device, a magnetic field probe position controller, and calibration software. TM9200 is suitable for testing or calibrating equipment such as electromagnetic induction magnetometers, alternating Hall magnetometers, alternating magnetic field measurement coils or probes.
Availability:
Quantity:
  • TM9200

  • TUNKIA

TM9200 Alternating Magnetometer Calibration Device

1. Summary

TM9200 is a high-precision alternating magnetometer calibration device. It consists of a high-precision AC current standard source, a magnetic field coil (Helmholtz coil), an intelligent probe position adjustment device, a magnetic field probe position controller, and calibration software. TM9200 is suitable for testing or calibrating equipment such as electromagnetic induction magnetometers, alternating Hall magnetometers, alternating magnetic field measurement coils or probes. Reference standard: JJG 1049-2009 "Verification Regulations for Weak Field Alternating Magnetometer".


2. Features

The stability of the current source reaches 0.01%/min.

The accuracy of the current source reaches 0.05 %.

Annual error variation is better than 100 ppm.

The coil can generate alternating magnetic field of 1 μT ~ 2 mT.

The current source can drive the coil to work for a long time under full load.

Good linearity between magnetic field and excitation current

The coil is equipped with an intelligent probe position adjustment device.

The system adopts modular integrated design.

Equipped with specialized calibration software.


3. Application

Calibrate The Alternating Magnetometer

DUT: electromagnetic induction magnetometer, alternating Hall magnetometer.

Reference standard: JJG 1049-2009 "Verification Regulations for Weak Field Alternating Magnetometer".

Alternating magnetic field testing or calibration: Use a Helmholtz coil as the magnetic field generator, and generate a standard alternating magnetic field of 1 μT ~ 2 mT at the frequency of 45 Hz ~ 1 kHz through the excitation of a high-precision current standard source to complete the testing or calibration of alternating magnetometer.


4. Specifications

4.1 AC Current Standard Source

Output Range

1 mA3.3 A

Frequency Range

45 Hz1 kHz

Adjust Fineness

0.001%*RG

Short Term Stability

0.01%/min

Optimal Accuracy

± ( 0.03%*RD+ 0.02%*RG )

Full Scale Linearity

< 50 ppm

Annual Error Change

< 100 ppm

Load Capacity

500 VA

Protective Function

Open circuit protection, overload protection functions

Power Supply

AC ( 220 ± 22 ) V( 50 ± 2 ) Hz

Note

RD is the reading value, RG is the range value.


4.2 Helmholtz Coil

Excitation Current

1 mA3.3 A

Magnetic Field Range

1 μT2 mT

Magnetic Field Uniformity

Uniform field within 20 mm is better than 300 ppm

Magnetic Field Linearity

The magnetic field is proportional to the excitation current and has good linearity

Temperature Effect

The coil constant K changes little with temperature, minimizing the influence of temperature.

Coil Average Radius

100mm

Remark

Calibration of alternating magnetometers whose lower limit of the measurement range is less than 3μT should be performed in a shielded cylinder


5. General Specifications

Power Supply

AC ( 220 ± 22 ) V( 50 ± 2 ) Hz

Temperature Performance

Working temperature: 0°C50°C

Storage temperature: -20°C70°C

Humidity Performance

Working humidity: 40%80% R H; non-condensing

Storage humidity: < 80% R H; non-condensing


6. Configuration List

S/N

Name

Quantity

Configuration

Note

1

AC Current Standard Source

1

Standard


2

Helmholtz Coil

1

Standard


3

Probe Position Intelligent Adjustment Device

1

Standard


4

Automatic Calibration Software

1

Standard


5

Complete Set of Test Leads and Power Cords

1

Standard


6

Workbench

1

Optional Accessory

Third party product

7

Computer

1

Optional Accessory

Third party product

8

Printer

1

Optional Accessory

Third party product

Note: The above is for reference only, the specific configuration list is subject to the technical agreement.


Previous: 
Next: 

SIGN UP FOR NEWSLETTER

Get latest deals, newsletter and technical information

INFORMATION

CONTACT US

 WhatsApp:+86-18773180215
     Wechat: 18773180215
 Phone: +86-18773180215
 E-mail: global@tunkia.com
E-mail: Eugene@tunkia.com
 Address: NO.16 Panpan Road, Changsha
Copyright  2021 TUNKIA Co., Ltd. All Rights Reserved. Supported by leadong.comSitemap