Welcome To TUNKIA !      Electrical Measurement    Magnetic Measurement
You are here: Home » Products » Advanced Electrical Reference Standards » TH1000 Ultra-stable Precision Current Source » TH1000 Ultra Stable Precision Current Source

loading

Share to:
facebook sharing button
twitter sharing button
line sharing button
wechat sharing button
linkedin sharing button
pinterest sharing button
whatsapp sharing button
sharethis sharing button

TH1000 Ultra Stable Precision Current Source

TH1000 is a precision DC current standard source, which can accurately output ± (0~110 mA) DC current with an optimal uncertainty of ±15 μA/A.
Availability:
Quantity:
  • TH1000

  • TUNKIA


TH1000 Ultra Stable Precision Current Source

TH1000 Ultra Stable Precision Current Source 4

1.Summary

TH1000 is a precision DC current standard source, which can accurately output ±(0~110 mA)  DC current with an optimal uncertainty of ±15 μA/A.


2. Features

  • Optimal uncertainty: ±15 μA/A.

  • DC current standard source:±(0~110 mA).

  • The maximum load voltage of constant current output reaches 10 V.

  • Bipolar source generates positive and negative current directly through the circuit.

  • Typical value of rise time < 10 ms, typical value of establishment time < 20 ms.



3. Specifications

3.1 DC Current Output

Range

Resolution (nA)

Short-term Stability[1]

Relative Accuracy

Absolute Measurement Accuracy (k=2)

24 hour

(23±1)°C

24 hour

(23±1)°C

24 hour

(23±5)°C

90 day

(23±5)°C

180 day

(23±5)°C

1 year

(23±5)°C

(ppm of reading + nA) [2]

100 μA

0.01

5 + 0.7

5 + 3

7 + 5

8 + 5

9 + 5

10 + 5

1 mA

0.1

5 + 4

5 + 5

7 + 6

8 + 6

9 + 6

10 + 6

10 mA

1

5 + 40

5 + 45

7 + 50

8 + 50

9 + 50

10 + 50

100 mA

10

8 + 250

8 + 300

10 + 500

12 + 500

13 + 500

15 + 500

note [1]:The 24-hour stability is expressed in the form of standard deviation of continuous measured values.

[2]:(ppm = parts per million) (e.g: 10ppm = 0.001%).


3.2 Auxiliary Performance

Range

Temperature Coefficient

Maximum Burden Voltage

Load Influence Error[3]

Noise

(10~40)°C

(0~10)°C (40~50)°C

(0.1~10) Hz

(pk-pk)

(10~10k) Hz

(rms)

(ppm of reading + nA)/°C

(V)

(nA/V)

(ppm of reading + nA)

100 μA

0.5 + 0.2

1.5 + 0.6

10

0.1

5 + 0.9

0 + 8

1 mA

0.5 + 1

1.5 + 5

10

0.2

5 + 5

0 + 8

10 mA

0.5 + 15

1.5 + 80

10

5

5 + 50

0 + 40

100 mA

0.8 + 150

2 + 800

10

50

8 + 300

0 + 400

note [3]:The load influence error is the addition of the absolute accuracy index. This rule applies when the load voltage exceeds 2.5 V; The rest are not applicable.

• Display digits: 7/8, adjustment fineness: 0.1 ppm;

• Establishment time: the time from setting the output to meeting the index requirements is less than 20 ms;

• Test conditions: 1 hour preheating, ambient temperature 23 &deg; C &plusmn; 1 &deg; C, sampling rate 0.2 sps;

• Protection mode: open circuit protection, overload protection


4. General Specifications

Power supply

AC ( 220±22 ) V,( 50 ± 2 ) Hz

Working Environment

0 °C ~ 40°C,20% R·H ~ 80% R·H,non-condensing.

Storage environment

-10°C ~ 50°C,10% R·H ~ 90% R·H,non-condensing.

Interface

RS232, LAN

Previous: 
Next: 

SIGN UP FOR NEWSLETTER

Get latest deals, newsletter and technical information

INFORMATION

CONTACT US

 WhatsApp:+86-18773180215
     Wechat: 18773180215
 Phone: +86-18773180215
 E-mail: global@tunkia.com
E-mail: Eugene@tunkia.com
 Address: NO.16 Panpan Road, Changsha
Copyright  2021 TUNKIA Co., Ltd. All Rights Reserved. Supported by leadong.comSitemap