Introduction
TD1600, a single-phase AC calibrator with wide range, high precision and multifunction.
The apparatus can output high accuracy AC voltage, current and power, and the optional accuracy are 0.05% and 0.02%. Applied to calibrate various AC electrical measuring instruments.
Features
AC Electrical Parameter Output: Voltage is up to 1100 V, current is up to 110 A.
Frequency and Phase Adjustment: Frequency: 45 Hz ~ 70 Hz.
AC Power Output: Measure the active power, reactive power, apparent power and power factor.
Value Output Mode: Standard source and output source modes, easy to calibrate both digital meters and analogue meters.
Value Adjustment Method: Fixed-point, rotary encoder, step and potentiometer (option).
Output Switch: The output can be turned on/off arbitrarily by one-touch operation.
Automatical Error Calculation: Set the measured values of the checked meter, and adjust the value output to automatically calculate the errors of the checked meter.
Excellent Human-computer Interaction: Large LCD touch screen display, touch and button operation modes, easy for users to measure manually.
Harmonic (Option): Load 2nd to 31st harmonics with variable amplitude and phase for impact test.
Proprietary Software (Option): Support the calibration of the checked meters by semi-automatic or full-automatic, data management and report export.
Introduction
TD1600, a single-phase AC calibrator with wide range, high precision and multifunction.
The apparatus can output high accuracy AC voltage, current and power, and the optional accuracy are 0.05% and 0.02%. Applied to calibrate various AC electrical measuring instruments.
Features
AC Electrical Parameter Output: Voltage is up to 1100 V, current is up to 110 A.
Frequency and Phase Adjustment: Frequency: 45 Hz ~ 70 Hz.
AC Power Output: Measure the active power, reactive power, apparent power and power factor.
Value Output Mode: Standard source and output source modes, easy to calibrate both digital meters and analogue meters.
Value Adjustment Method: Fixed-point, rotary encoder, step and potentiometer (option).
Output Switch: The output can be turned on/off arbitrarily by one-touch operation.
Automatical Error Calculation: Set the measured values of the checked meter, and adjust the value output to automatically calculate the errors of the checked meter.
Excellent Human-computer Interaction: Large LCD touch screen display, touch and button operation modes, easy for users to measure manually.
Harmonic (Option): Load 2nd to 31st harmonics with variable amplitude and phase for impact test.
Proprietary Software (Option): Support the calibration of the checked meters by semi-automatic or full-automatic, data management and report export.
Specification
AC Voltage Output
Range | Stability (%/min) | Accuracy (k = 2) (ppm*RD + ppm*RG) | Max. Load Current (mA) | ||
0.05% | 0.02% | 0.05% | 0.02% | ||
1 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 500 |
3 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 500 |
10 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 500 |
30 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 500 |
60 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 800 |
100 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 500 |
150 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 400 |
200 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 250 |
300 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 150 |
400 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 100 |
600 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 80 |
750 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 60 |
1000 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 50 |
AC Current Output
Range | Stability (%/min) | Accuracy (k = 2) (ppm*RD + ppm*RG) | Max. Inductive Load (V) | ||
0.05% | 0.02% | 0.05% | 0.02% | ||
1 mA | 1 μA | 0.5 μA | 300 + 2 μA | 120 + 1 μA | 120 |
2 mA | 1 μA | 0.5 μA | 300 + 2 μA | 120 + 1 μA | 120 |
5 mA | 1 μA | 0.5 μA | 300 + 2 μA | 120 + 1 μA | 120 |
10 mA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 120 |
20 mA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 120 |
50 mA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 120 |
100 mA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 120 |
200 mA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 120 |
500 mA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 100 |
1 A | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 50 |
2 A | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 25 |
5 A | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 10 |
10 A | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 5 |
20 A | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 2.5 |
30 A | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 2.5 |
50 A | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 1.5 |
100 A | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 0.8 |
Output Characteristics of Voltage and Current
Output Range of Voltage: 100 mV ~ 1100 V
Output Range of Current: 0.1 mA/ 20 mA ~ 110 A
Range Switching: Automatic or Manual.
Resolution: 0.001 %*RG; Display: 6 digits; Distortion: <0.2 %
1 mA、2 mA、5 mA are small range option.
Frequency, Phase and Harmonic (Option)
Type | Range | Resolution | Accuracy |
Frequency | 45.000 Hz ~ 70.000 Hz | 0.001 Hz | 0.005 Hz |
Phase | 0.0000 ° ~ 359.9999 ° | 0.005 ° | 0.05 ° |
Type | Order | Amplitude | Variable Phase |
Harmonic | 2nd to 31st | 0 ~ 25 % | 0 ~ 359.9 ° |
Power Output
Stability (%/min) | Accuracy (k = 2) %*FS | |||
0.05% | 0.02% | 0.05% | 0.02% | |
0.01 | 0.005 | 0.05 | 0.025 | |
Reactive Power | 0.02 | 0.01 | 0.1 | 0.05 |
Apparent Power | 0.02 | 0.01 | 0.1 | 0.05 |
Power Factor | 0.02 | 0.01 | 0.1 | 0.05 |
Output Range of AC Power: Combination of AC voltage range and AC current range.
Range of Power Factor: -1.0000…0.0000…+1.0000
Active Power (|cosφ| ≥0.5), Reactive Power (|sinφ| ≥0.5)
Specification
AC Voltage Output
Range | Stability (%/min) | Accuracy (k = 2) (ppm*RD + ppm*RG) | Max. Load Current (mA) | ||
0.05% | 0.02% | 0.05% | 0.02% | ||
1 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 500 |
3 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 500 |
10 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 500 |
30 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 500 |
60 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 800 |
100 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 500 |
150 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 400 |
200 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 250 |
300 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 150 |
400 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 100 |
600 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 80 |
750 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 60 |
1000 V | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 50 |
AC Current Output
Range | Stability (%/min) | Accuracy (k = 2) (ppm*RD + ppm*RG) | Max. Inductive Load (V) | ||
0.05% | 0.02% | 0.05% | 0.02% | ||
1 mA | 1 μA | 0.5 μA | 300 + 2 μA | 120 + 1 μA | 120 |
2 mA | 1 μA | 0.5 μA | 300 + 2 μA | 120 + 1 μA | 120 |
5 mA | 1 μA | 0.5 μA | 300 + 2 μA | 120 + 1 μA | 120 |
10 mA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 120 |
20 mA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 120 |
50 mA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 120 |
100 mA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 120 |
200 mA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 120 |
500 mA | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 100 |
1 A | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 50 |
2 A | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 25 |
5 A | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 10 |
10 A | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 5 |
20 A | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 2.5 |
30 A | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 2.5 |
50 A | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 1.5 |
100 A | 0.01 | 0.005 | 300 + 200 | 120 + 80 | 0.8 |
Output Characteristics of Voltage and Current
Output Range of Voltage: 100 mV ~ 1100 V
Output Range of Current: 0.1 mA/ 20 mA ~ 110 A
Range Switching: Automatic or Manual.
Resolution: 0.001 %*RG; Display: 6 digits; Distortion: <0.2 %
1 mA、2 mA、5 mA are small range option.
Frequency, Phase and Harmonic (Option)
Type | Range | Resolution | Accuracy |
Frequency | 45.000 Hz ~ 70.000 Hz | 0.001 Hz | 0.005 Hz |
Phase | 0.0000 ° ~ 359.9999 ° | 0.005 ° | 0.05 ° |
Type | Order | Amplitude | Variable Phase |
Harmonic | 2nd to 31st | 0 ~ 25 % | 0 ~ 359.9 ° |
Power Output
Stability (%/min) | Accuracy (k = 2) %*FS | |||
0.05% | 0.02% | 0.05% | 0.02% | |
0.01 | 0.005 | 0.05 | 0.025 | |
Reactive Power | 0.02 | 0.01 | 0.1 | 0.05 |
Apparent Power | 0.02 | 0.01 | 0.1 | 0.05 |
Power Factor | 0.02 | 0.01 | 0.1 | 0.05 |
Output Range of AC Power: Combination of AC voltage range and AC current range.
Range of Power Factor: -1.0000…0.0000…+1.0000
Active Power (|cosφ| ≥0.5), Reactive Power (|sinφ| ≥0.5)