TD3750
TUNKIA
* For reference only. Configuration and details may vary depending on the application scenario.。
1. Summary
TD3750 is a device used to test the electromagnetic compatibility (EMC) of single-phase or three-phase AC energy meters, mainly composed of a three-phase standard power source, a filter module, system software, etc. The device has a measurement uncertainty of class 0.05 and can be used in conjunction with relevant EMC influence quantity generation devices to complete relevant energy meter EMC tests.
2. Features
• High reliability: Maximum output 456 V / 60 A, continuous operation at the load point (1.1Umax, Imax) for 1000h can ensure the stability of the value output.
• Single-phase and three-phase compatibility: Supports single-phase meter test for single meter position, and supports three-phase meter test for three meter positions.
• Filter network: The device has voltage signal, current signal, and pulse signal filter networks, which can effectively filter interference signals in Class A acceptance tests and conduct real-time tests on the metering performance of the meter under test.
• Supports multiple energy metering methods such as forward and reverse active energy, four-quadrant reactive energy, etc.
• The software has comprehensive data management functions, convenient data retrieval, and automatic backup functions.
• The software can output statistical reports such as verification certificates and test reports; supports document saving and printing in multiple formats.
3. Test Items
For Class A acceptance tests, the metering performance of the DUT can be verified in the EMC impact test. The detailed test items are as follows:
• Electrical fast transient pulse group
• Damped oscillating wave
• RF electromagnetic field radiation (with current)
• RF field induced conducted interference
• Differential mode current interference test (only supports single meter position test)
For Class B acceptance test, the metering performance of the DUT can be verified after the EMC influence is removed. The detailed items are as follows:
• Electrostatic discharge
• Radiated RF electromagnetic field
• Surge
• Radiated RF electromagnetic field (no current)
* For reference only. Configuration and details may vary depending on the application scenario.。
1. Summary
TD3750 is a device used to test the electromagnetic compatibility (EMC) of single-phase or three-phase AC energy meters, mainly composed of a three-phase standard power source, a filter module, system software, etc. The device has a measurement uncertainty of class 0.05 and can be used in conjunction with relevant EMC influence quantity generation devices to complete relevant energy meter EMC tests.
2. Features
• High reliability: Maximum output 456 V / 60 A, continuous operation at the load point (1.1Umax, Imax) for 1000h can ensure the stability of the value output.
• Single-phase and three-phase compatibility: Supports single-phase meter test for single meter position, and supports three-phase meter test for three meter positions.
• Filter network: The device has voltage signal, current signal, and pulse signal filter networks, which can effectively filter interference signals in Class A acceptance tests and conduct real-time tests on the metering performance of the meter under test.
• Supports multiple energy metering methods such as forward and reverse active energy, four-quadrant reactive energy, etc.
• The software has comprehensive data management functions, convenient data retrieval, and automatic backup functions.
• The software can output statistical reports such as verification certificates and test reports; supports document saving and printing in multiple formats.
3. Test Items
For Class A acceptance tests, the metering performance of the DUT can be verified in the EMC impact test. The detailed test items are as follows:
• Electrical fast transient pulse group
• Damped oscillating wave
• RF electromagnetic field radiation (with current)
• RF field induced conducted interference
• Differential mode current interference test (only supports single meter position test)
For Class B acceptance test, the metering performance of the DUT can be verified after the EMC influence is removed. The detailed items are as follows:
• Electrostatic discharge
• Radiated RF electromagnetic field
• Surge
• Radiated RF electromagnetic field (no current)
4. Specification
4.1 Three-phase Voltage Output
Range | Resolution | Short-term Stability ( % / min ) | Measurement Uncertainty (k=2) ( ppm*RD+ppm*RG) [1] | Distortion ( % ) |
60 V | 0.1 mV | 0.01 | 200 + 50 | < 0.3 |
120 V | 1 mV | 0.01 | 200 + 50 | < 0.3 |
240 V | 1 mV | 0.01 | 200 + 50 | < 0.3 |
380 V | 1 mV | 0.01 | 200 + 50 | < 0.3 |
Note [1]: RD is the reading value, RG is the range value, the same below |
• Output range: 6 V ~ 456 V, 6-digit display, load capacity: 15 VA / (phase meter position)
• Symmetry: voltage ≤ 0.2%, phase ≤ 0.5°
• Protection function: short circuit protection, overload protection
4.2 Three-phase Current Output
Range | Resolution | Short-term Stability ( % / min ) | Measurement Uncertainty (k=2) ( ppm*RD+ppm*RG) | Distortion ( % ) |
100 mA | 1μA | 0.01 | 200 + 50 | < 0.5 |
200 mA | 1μA | 0.01 | 200 + 50 | < 0.5 |
500 mA | 1μA | 0.01 | 200 + 50 | < 0.5 |
1 A | 10μA | 0.01 | 200 + 50 | < 0.5 |
2 A | 10μA | 0.01 | 200 + 50 | < 0.5 |
5 A | 10μA | 0.01 | 200 + 50 | < 0.5 |
10 A | 100μA | 0.01 | 200 + 50 | < 0.5 |
20 A | 100μA | 0.01 | 200 + 50 | < 0.5 |
50 A | 100μA | 0.01 | 200 + 50 | < 0.5 |
• Output range: 10 mA ~ 60 A, 6-digit display, load capacity: 30 VA / (phase meter position)
• Symmetry: current ≤ 0.5%, phase ≤ 0.5°
• Protection function: open circuit protection, overload protection
4.3 Frequency/Phase
Frequency | Measuring range | 45 Hz ~ 65 Hz |
Minimum resolution | 0.000 01 Hz | |
Measurement uncertainty (k=2) | 0.005%*RD | |
Phase (I ≥ 50mA) | Measuring range | 0 ~ 360° |
Minimum resolution | 0.000 1° | |
Measurement uncertainty (k=2) | 0.012° |
4.4 Power/Energy
Voltage Range | Current Range | Power Factor | Measurement Uncertainty(k=2) |
30 V ≤ U ≤ 480 V | 50 mA ≤ I ≤ 60 A | 0.5L~1~0.5C | 0.05%*RD |
10 mA ≤ I < 50 mA | 1 | 0.05%*RD | |
0.5L~1~0.5C | 0.08%*RD | ||
5 mA ≤ I < 10 mA | 1 | 0.08%*RD | |
0.5L~1~0.5C | 0.15%*RD |
• Output power stability: 0.02% / 2min;
• Power/energy measurement range: combination of AC voltage range and AC current range;
• Power factor measurement range: -1.000 00…0.000 00…1.000 00;
• Standard electric energy pulse output: high frequency full range value corresponds to 60 kHz, low frequency full range value corresponds to 6 Hz;
• Standard electric energy pulse input: frequency ≤ 200 kHz, voltage: 0…3.3 V…24 V;
• Energy error display: automatic, with a resolution of 0.0001%.
5. General Specification
Power Supply | Single-phase AC ( 220 ± 22 ) V, ( 50 ±2 ) Hz |
Power Consumption | About 1500 VA |
Warm-up Time | 30-min |
Temperature Performance | Working temperature: 5°C~45°C; Storage temperature: -10°C~55°C |
Humidity Performance | Working humidity: < 80% @ 30°C, < 70% @ 40°C, < 40% @ 50°C Storage humidity: (20%~80%) R H, no condensation |
Altitude | < 3000 m |
Communication Interface | RS232 |
Dimensions | 550 mm(W) x 700 mm(D) x 1472 mm(H) |
4. Specification
4.1 Three-phase Voltage Output
Range | Resolution | Short-term Stability ( % / min ) | Measurement Uncertainty (k=2) ( ppm*RD+ppm*RG) [1] | Distortion ( % ) |
60 V | 0.1 mV | 0.01 | 200 + 50 | < 0.3 |
120 V | 1 mV | 0.01 | 200 + 50 | < 0.3 |
240 V | 1 mV | 0.01 | 200 + 50 | < 0.3 |
380 V | 1 mV | 0.01 | 200 + 50 | < 0.3 |
Note [1]: RD is the reading value, RG is the range value, the same below |
• Output range: 6 V ~ 456 V, 6-digit display, load capacity: 15 VA / (phase meter position)
• Symmetry: voltage ≤ 0.2%, phase ≤ 0.5°
• Protection function: short circuit protection, overload protection
4.2 Three-phase Current Output
Range | Resolution | Short-term Stability ( % / min ) | Measurement Uncertainty (k=2) ( ppm*RD+ppm*RG) | Distortion ( % ) |
100 mA | 1μA | 0.01 | 200 + 50 | < 0.5 |
200 mA | 1μA | 0.01 | 200 + 50 | < 0.5 |
500 mA | 1μA | 0.01 | 200 + 50 | < 0.5 |
1 A | 10μA | 0.01 | 200 + 50 | < 0.5 |
2 A | 10μA | 0.01 | 200 + 50 | < 0.5 |
5 A | 10μA | 0.01 | 200 + 50 | < 0.5 |
10 A | 100μA | 0.01 | 200 + 50 | < 0.5 |
20 A | 100μA | 0.01 | 200 + 50 | < 0.5 |
50 A | 100μA | 0.01 | 200 + 50 | < 0.5 |
• Output range: 10 mA ~ 60 A, 6-digit display, load capacity: 30 VA / (phase meter position)
• Symmetry: current ≤ 0.5%, phase ≤ 0.5°
• Protection function: open circuit protection, overload protection
4.3 Frequency/Phase
Frequency | Measuring range | 45 Hz ~ 65 Hz |
Minimum resolution | 0.000 01 Hz | |
Measurement uncertainty (k=2) | 0.005%*RD | |
Phase (I ≥ 50mA) | Measuring range | 0 ~ 360° |
Minimum resolution | 0.000 1° | |
Measurement uncertainty (k=2) | 0.012° |
4.4 Power/Energy
Voltage Range | Current Range | Power Factor | Measurement Uncertainty(k=2) |
30 V ≤ U ≤ 480 V | 50 mA ≤ I ≤ 60 A | 0.5L~1~0.5C | 0.05%*RD |
10 mA ≤ I < 50 mA | 1 | 0.05%*RD | |
0.5L~1~0.5C | 0.08%*RD | ||
5 mA ≤ I < 10 mA | 1 | 0.08%*RD | |
0.5L~1~0.5C | 0.15%*RD |
• Output power stability: 0.02% / 2min;
• Power/energy measurement range: combination of AC voltage range and AC current range;
• Power factor measurement range: -1.000 00…0.000 00…1.000 00;
• Standard electric energy pulse output: high frequency full range value corresponds to 60 kHz, low frequency full range value corresponds to 6 Hz;
• Standard electric energy pulse input: frequency ≤ 200 kHz, voltage: 0…3.3 V…24 V;
• Energy error display: automatic, with a resolution of 0.0001%.
5. General Specification
Power Supply | Single-phase AC ( 220 ± 22 ) V, ( 50 ±2 ) Hz |
Power Consumption | About 1500 VA |
Warm-up Time | 30-min |
Temperature Performance | Working temperature: 5°C~45°C; Storage temperature: -10°C~55°C |
Humidity Performance | Working humidity: < 80% @ 30°C, < 70% @ 40°C, < 40% @ 50°C Storage humidity: (20%~80%) R H, no condensation |
Altitude | < 3000 m |
Communication Interface | RS232 |
Dimensions | 550 mm(W) x 700 mm(D) x 1472 mm(H) |