Welcome To TUNKIA !      Electrical Measurement    Magnetic Measurement
You are here: Home » Products » Wideband Power Source/Meter » TA3000 Power Measurement Standard » TA3000 Power Measurement Standard

loading

Share to:
facebook sharing button
twitter sharing button
line sharing button
wechat sharing button
linkedin sharing button
pinterest sharing button
whatsapp sharing button
sharethis sharing button

TA3000 Power Measurement Standard

TA3000 is a high-precision wideband AC parametric measurement standard that simultaneously measures multiple amounts of electricity in an AC loop such as: AC Voltage measuremrnt , AC Current measuremrnt, Frequency measuremrnt, Phase measuremrnt, Harmonics measuremrnt, Active Power/ Electrical Energy measuremrnt, Reactive Power/Energy measuremrnt, Apparent Power measuremrnt, Power Factor measuremrnt, etc.
Availability:
Quantity:
  • TA3000

  • TUNKIA

TA3000 Power Measurement Standard

TA3000 Power Measurement Standard

1.Summary

TA3000 is a high-precision wideband AC parametric measurement standard that simultaneously measures multiple amounts of electricity in an AC loop such as: Voltage, Current, Frequency, Phase, Harmonics, Active Power/ Electrical Energy, Reactive Power/Energy, Apparent Power, Power Factor, etc.


2. Features

• Power/Energy Measurement Uncertainty: Class 0.01.

• Voltage Measurement Range: 1 V~600 V.

• Current Measurement Range: 10mA~120 A.

• Fundamental Frequency Range: 45 Hz~1 kHz.

• Phase Measurement Uncertainty is typically 0.003°.

• Voltage and Current support fully automatic range shifting.

• Voltage and Current support 2~127th harmonic measurement.

• Standard Energy Pulse input/output function.

• USB, RS232 and LAN interfaces.

• LCD touch screen.


3. Specifications

3.1 AC Voltage Measurement

Range

Resolution

Measurement Uncertainty at different frequencies (Hz) (k = 2).

( ppm*RD + ppm*RG )

Temperature Coefficient @ (15~30)°C 

(ppm*RD+ppm*RG)/°C

45 ≤ F < 80

80 ≤ F < 200

200 ≤ F < 400

400 ≤ F < 1k

12 V

10 μV

30 + 20

90 + 60

150 + 100

450 + 300

1 + 0.5

60 V

10 μV

30 + 20

90 + 60

150 + 100

450 + 300

1 + 0. 5

120 V

0.1 mV

30 + 20

90 + 60

150 + 100

450 + 300

1 + 0. 5

240 V

0.1 mV

30 + 20

90 + 60

150 + 100

450 + 300

1 + 0. 5

600 V

0.1 mV

30 + 20

90 + 60

150 + 100

450 + 300

1 + 0. 5

Note: [1] RD is the reading value, RG is the range value, the same below

• Measuring range: 1 V~600 V, 7-digit display, manual/automatic range shifting


3.2 AC Current Measurement

Range

Resolution

Measurement Uncertainty at different frequencies (Hz) (k = 2).

( ppm*RD + ppm*RG )

Temperature Coefficient @ (15~30)°C 

(ppm*RD+ppm*RG)/°C

45≤F<80

80≤F<200

200≤F<400

400≤F<1000

100 mA

0.1 μA

30 + 20

90 + 60

150 + 100

450 + 300

1 + 0.5

200 mA

0.1 μA

30 + 20

90 + 60

150 + 100

450 + 300

1 + 0. 5

500 mA

0.1 μA

30 + 20

90 + 60

150 + 100

450 + 300

1 + 0. 5

1 A

1 μA

30 + 20

90 + 60

150 + 100

450 + 300

1 + 0. 5

2 A

1 μA

30 + 20

90 + 60

150 + 100

450 + 300

1 + 0. 5

5 A

1 μA

30 + 20

90 + 60

150 + 100

450 + 300

1 + 0.5

10 A

10 μA

30 + 20

90 + 60

150 + 100

450 + 300

1 + 0. 5

20 A

10 μA

30 + 20

90 + 60

150 + 100

450 + 300

1 + 0. 5

50 A

10 μA

30 + 20

90 + 60

150 + 100

450 + 300

1 + 0. 5

100 A

100 μA

30 + 20

90 + 60

150 + 100

450 + 300

1 + 0. 5

• Measuring range: 10 mA~120 A, 7-digit display, manual/automatic range shifting


3.3 Frequency/Phase Measurement

Frequency

Measuring Range

40.00000 Hz~1000.000 Hz

Minimum Resolution

0.00001 Hz

Measurement Uncertainty (k=2)

0.005%*RD

Phase

Measuring Range

0.0000°~ 359.9999°

Minimum Resolution

0.000 1°

Measurement Uncertainty (k=2)

45 Hz ≤ F < 80 Hz

0.005°

80 Hz ≤ F < 200 Hz

0.015°

200 Hz ≤ F < 400 Hz

0.03°

400 Hz ≤ F < 1 kHz

0.08°


3.4 Power / Energy Measurement

The amount of Power being measured

Measurement Uncertainty at different frequencies (Hz) (k=2).

45≤F<80

80 ≤ F<200

200 ≤ F < 400

400 ≤ F ≤ 1000

Active Power/Energy: |cosφ| ≥0.8 or

Reactive Power/Energy: | SINφ | ≥0.8 or

Apparent Power/Energy

0.01%*RD

0.03%*RD

0.05%*RD

0.15%*RD

Active Power/Energy: |cosφ| <0.8 or

Reactive Power/Energy: | SINφ | <0.8

0.01%*FS[2]

0.03%*FS

0.05%*FS

0.15%*FS

Note [2]: FS = voltage range value × current range value

• Power/energy measurement range: a combination of AC voltage range and AC current range

• Power factor measurement range: -1.000 0 0 0... 0.000 000... 1.000000

• Standard energy pulse output: 60 kHz for high-frequency full-scale values and 6 Hz for low-frequency full-scale values

• Standard energy pulse input: frequency≤ 200 kHz, voltage: 0... 3.3 V... 24 V


4. General specifications

• Power Supply: AC (220 ± 22) V, (50 ± 2) Hz

• Working Environment: 5°C~40°C, (20%~85%) R·H, no condensation

• Storage Environment: -10°C ~ 50°C, < 95% R·H, no condensation

• Communication Interface: RS232, USB, LAN

Previous: 
Next: 

SIGN UP FOR NEWSLETTER

Get latest deals, newsletter and technical information

INFORMATION

CONTACT US

 WhatsApp:+86-18773180215
     Wechat: 18773180215
 Phone: +86-18773180215
 E-mail: global@tunkia.com
E-mail: Eugene@tunkia.com
 Address: NO.16 Panpan Road, Changsha
Copyright  2021 TUNKIA Co., Ltd. All Rights Reserved. Supported by leadong.comSitemap