Availability: | |
---|---|
Quantity: | |
TA3000
TUNKIA
1.Summary
TA3000 is a high-precision wideband AC parametric measurement standard that simultaneously measures multiple amounts of electricity in an AC loop such as: Voltage, Current, Frequency, Phase, Harmonics, Active Power/ Electrical Energy, Reactive Power/Energy, Apparent Power, Power Factor, etc.
2. Features
• Power/Energy Measurement Uncertainty: Class 0.01.
• Voltage Measurement Range: 1 V~600 V.
• Current Measurement Range: 10mA~120 A.
• Fundamental Frequency Range: 45 Hz~1 kHz.
• Phase Measurement Uncertainty is typically 0.003°.
• Voltage and Current support fully automatic range shifting.
• Voltage and Current support 2~127th harmonic measurement.
• Standard Energy Pulse input/output function.
• USB, RS232 and LAN interfaces.
• LCD touch screen.
3. Specifications
3.1 AC Voltage Measurement
Range | Resolution | Measurement Uncertainty at different frequencies (Hz) (k = 2). ( ppm*RD + ppm*RG ) | Temperature Coefficient @ (15~30)°C (ppm*RD+ppm*RG)/°C | |||
45 ≤ F < 80 | 80 ≤ F < 200 | 200 ≤ F < 400 | 400 ≤ F < 1k | |||
12 V | 10 μV | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0.5 |
60 V | 10 μV | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
120 V | 0.1 mV | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
240 V | 0.1 mV | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
600 V | 0.1 mV | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
Note: [1] RD is the reading value, RG is the range value, the same below |
• Measuring range: 1 V~600 V, 7-digit display, manual/automatic range shifting
3.2 AC Current Measurement
Range | Resolution | Measurement Uncertainty at different frequencies (Hz) (k = 2). ( ppm*RD + ppm*RG ) | Temperature Coefficient @ (15~30)°C (ppm*RD+ppm*RG)/°C | |||
45≤F<80 | 80≤F<200 | 200≤F<400 | 400≤F<1000 | |||
100 mA | 0.1 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0.5 |
200 mA | 0.1 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
500 mA | 0.1 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
1 A | 1 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
2 A | 1 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
5 A | 1 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0.5 |
10 A | 10 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
20 A | 10 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
50 A | 10 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
100 A | 100 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
• Measuring range: 10 mA~120 A, 7-digit display, manual/automatic range shifting
3.3 Frequency/Phase Measurement
Frequency | Measuring Range | 40.00000 Hz~1000.000 Hz | |
Minimum Resolution | 0.00001 Hz | ||
Measurement Uncertainty (k=2) | 0.005%*RD | ||
Phase | Measuring Range | 0.0000°~ 359.9999° | |
Minimum Resolution | 0.000 1° | ||
Measurement Uncertainty (k=2) | 45 Hz ≤ F < 80 Hz | 0.005° | |
80 Hz ≤ F < 200 Hz | 0.015° | ||
200 Hz ≤ F < 400 Hz | 0.03° | ||
400 Hz ≤ F < 1 kHz | 0.08° |
3.4 Power / Energy Measurement
The amount of Power being measured | Measurement Uncertainty at different frequencies (Hz) (k=2). | |||
45≤F<80 | 80 ≤ F<200 | 200 ≤ F < 400 | 400 ≤ F ≤ 1000 | |
Active Power/Energy: |cosφ| ≥0.8 or Reactive Power/Energy: | SINφ | ≥0.8 or Apparent Power/Energy | 0.01%*RD | 0.03%*RD | 0.05%*RD | 0.15%*RD |
Active Power/Energy: |cosφ| <0.8 or Reactive Power/Energy: | SINφ | <0.8 | 0.01%*FS[2] | 0.03%*FS | 0.05%*FS | 0.15%*FS |
Note [2]: FS = voltage range value × current range value |
• Power/energy measurement range: a combination of AC voltage range and AC current range
• Power factor measurement range: -1.000 0 0 0... 0.000 000... 1.000000
• Standard energy pulse output: 60 kHz for high-frequency full-scale values and 6 Hz for low-frequency full-scale values
• Standard energy pulse input: frequency≤ 200 kHz, voltage: 0... 3.3 V... 24 V
4. General specifications
• Power Supply: AC (220 ± 22) V, (50 ± 2) Hz
• Working Environment: 5°C~40°C, (20%~85%) R·H, no condensation
• Storage Environment: -10°C ~ 50°C, < 95% R·H, no condensation
• Communication Interface: RS232, USB, LAN
1.Summary
TA3000 is a high-precision wideband AC parametric measurement standard that simultaneously measures multiple amounts of electricity in an AC loop such as: Voltage, Current, Frequency, Phase, Harmonics, Active Power/ Electrical Energy, Reactive Power/Energy, Apparent Power, Power Factor, etc.
2. Features
• Power/Energy Measurement Uncertainty: Class 0.01.
• Voltage Measurement Range: 1 V~600 V.
• Current Measurement Range: 10mA~120 A.
• Fundamental Frequency Range: 45 Hz~1 kHz.
• Phase Measurement Uncertainty is typically 0.003°.
• Voltage and Current support fully automatic range shifting.
• Voltage and Current support 2~127th harmonic measurement.
• Standard Energy Pulse input/output function.
• USB, RS232 and LAN interfaces.
• LCD touch screen.
3. Specifications
3.1 AC Voltage Measurement
Range | Resolution | Measurement Uncertainty at different frequencies (Hz) (k = 2). ( ppm*RD + ppm*RG ) | Temperature Coefficient @ (15~30)°C (ppm*RD+ppm*RG)/°C | |||
45 ≤ F < 80 | 80 ≤ F < 200 | 200 ≤ F < 400 | 400 ≤ F < 1k | |||
12 V | 10 μV | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0.5 |
60 V | 10 μV | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
120 V | 0.1 mV | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
240 V | 0.1 mV | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
600 V | 0.1 mV | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
Note: [1] RD is the reading value, RG is the range value, the same below |
• Measuring range: 1 V~600 V, 7-digit display, manual/automatic range shifting
3.2 AC Current Measurement
Range | Resolution | Measurement Uncertainty at different frequencies (Hz) (k = 2). ( ppm*RD + ppm*RG ) | Temperature Coefficient @ (15~30)°C (ppm*RD+ppm*RG)/°C | |||
45≤F<80 | 80≤F<200 | 200≤F<400 | 400≤F<1000 | |||
100 mA | 0.1 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0.5 |
200 mA | 0.1 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
500 mA | 0.1 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
1 A | 1 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
2 A | 1 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
5 A | 1 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0.5 |
10 A | 10 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
20 A | 10 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
50 A | 10 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
100 A | 100 μA | 30 + 20 | 90 + 60 | 150 + 100 | 450 + 300 | 1 + 0. 5 |
• Measuring range: 10 mA~120 A, 7-digit display, manual/automatic range shifting
3.3 Frequency/Phase Measurement
Frequency | Measuring Range | 40.00000 Hz~1000.000 Hz | |
Minimum Resolution | 0.00001 Hz | ||
Measurement Uncertainty (k=2) | 0.005%*RD | ||
Phase | Measuring Range | 0.0000°~ 359.9999° | |
Minimum Resolution | 0.000 1° | ||
Measurement Uncertainty (k=2) | 45 Hz ≤ F < 80 Hz | 0.005° | |
80 Hz ≤ F < 200 Hz | 0.015° | ||
200 Hz ≤ F < 400 Hz | 0.03° | ||
400 Hz ≤ F < 1 kHz | 0.08° |
3.4 Power / Energy Measurement
The amount of Power being measured | Measurement Uncertainty at different frequencies (Hz) (k=2). | |||
45≤F<80 | 80 ≤ F<200 | 200 ≤ F < 400 | 400 ≤ F ≤ 1000 | |
Active Power/Energy: |cosφ| ≥0.8 or Reactive Power/Energy: | SINφ | ≥0.8 or Apparent Power/Energy | 0.01%*RD | 0.03%*RD | 0.05%*RD | 0.15%*RD |
Active Power/Energy: |cosφ| <0.8 or Reactive Power/Energy: | SINφ | <0.8 | 0.01%*FS[2] | 0.03%*FS | 0.05%*FS | 0.15%*FS |
Note [2]: FS = voltage range value × current range value |
• Power/energy measurement range: a combination of AC voltage range and AC current range
• Power factor measurement range: -1.000 0 0 0... 0.000 000... 1.000000
• Standard energy pulse output: 60 kHz for high-frequency full-scale values and 6 Hz for low-frequency full-scale values
• Standard energy pulse input: frequency≤ 200 kHz, voltage: 0... 3.3 V... 24 V
4. General specifications
• Power Supply: AC (220 ± 22) V, (50 ± 2) Hz
• Working Environment: 5°C~40°C, (20%~85%) R·H, no condensation
• Storage Environment: -10°C ~ 50°C, < 95% R·H, no condensation
• Communication Interface: RS232, USB, LAN