TD3250 is a portable standard meter specially used for three-phase energy meter on-site calibration, AC sampling etc. It integrates the functions of electrical parameter measurement, electric energy meter calibration, connection mode identification, harmonic analysis, waveform display, phasor diagram display, data management.
• Accuracy: Class 0.02 or 0.05.
• 3PH voltage measurement: 0~576 V.
• 3PH current measurement (direct) : 50 mA~ 12 A.
• 3PH current measurement (clamp) : 100 mA~ 1kA.
• Support electric energy pulse optical/electrical pulse input.
• LCD touch screen.
• Support AC 100 V~264 V wide range supply.
• Support large capacity lithium battery.
• Internal memory, and quickly record test data.
• USB and RS232 interfaces.
• Compact, lightweight, and easy to carry
☆ Calibration for Energy Meter Online |
|
• TD3250 supports direct measurement of three-phase voltage in the range of 10 V ~ 576 V, direct measurement of three-phase current from 0.05 A ~ 12 A, and clamp-based measurement of three-phase current from 0.1 A ~ 1 kA. For energy metering, it includes an electric pulse input interface, paired with a portable optical sensor. • Application 1 (Direct Measurement of Current): applicable to the calibration of single-phase / three-phase voltage source, current source, power source, etc. in the laboratory or on site. • Application 2 (Use Clamp for Current): When testing the current of an energy meter on-site, it is usually necessary to wait for a power outage to disconnect the circuit, then either insert a standard ammeter into the loop for testing or remove the meter and test it in the laboratory using a standard source method. However, by using a clamp meter for current measurement, there is no need to disconnect the circuit. Simply clamp the current-carrying wire into the jaws, and real-time measurement can be performed online. This method allows for quicker and more convenient on-site testing, meeting the requirements for routine inspection of instrument operating conditions. |
☆ CT Transformation Ratio Calibration |
|
• CT transformation ratio test: use large current clamp for primary signal of CT, and with small current clamp for secondary signal of CT; Calculate the transformation ratio of CT and the phase relationship of primary/secondary current. Help judge if the CT runs well. • Note: Standard configuration for 5 A clamp, which can be used for the secondary side, If user need other clamps should indicate the optional specification in the order. |
☆ Front panel | |
| |
S/N | Function |
1 | LAN/USB communication terminals are used to connect to a computer for data transfer via management software. |
2 | The instrument power supply socket can be used for mains power supply or to charge the lithium battery. |
3 | Reserved functional interface for expanding additional environmental measurement capabilities and more. |
4 | Pulse signal input terminal, compatible with optical sensors for energy metering. |
5 | Clamp current terminal, enabling clamp meter measurement of currents ranging from 0.1 A to 1 kA. |
6 | Three-phase voltage terminal: Allows for a wide voltage measurement range of 10 V to 576 V with a single connection. |
7 | Three-phase current terminal: Directly measures currents from 0.05 A to 12 A. |
8 | Pulse signal output terminal: Generates standard energy pulses for higher-level calibration and verification. |
9 | LCD touch color screen with multi-power display, fully touch-operated, greatly enhancing the instrument's ease of use. |
☆ Convenient Operation |
|
• Large LCD screen: Color LCD with high brightness, supports touch operation, offering comprehensive functionality that is simple and convenient to use.. • Front panel wiring: convenient for the user to replace the test lead. • 220 V mains power supply or the lithium battery mode available. • Built-in large-capacity memory for quickly recording test data of the equipment being tested on-site, capable of storing up to thousands of entries. After calibration is complete, the data can be transferred to the accompanying computer management software. • Solution advantages: This combined operation method effectively enhances ease of use, making it suitable for a variety of application scenarios. |
|
• The volume is equivalent to the conventional tool box, weighs about 8 kg. |
☆ Connection Mode Identify and Graphical Display | |
| |
S/N | Function |
1 | Accurately measure/display the amplitude and phase of voltage and current of each phase. |
2 | Accurately measure the phase between voltage and current of each phase, and visually display them in the form of phasor diagram. |
3 | Automatically identify the actual results of voltage and current, including phase sequence display. |
4 | The trend chart display function of the measuring channel can display the trend change of the measured electricity quantity with time. |
5 | Users can customize single or multiple power waveforms for display. |
☆ Harmonic Function | |
| |
S/N | Function |
1 | It can measure 2nd ~63rd harmonics and display the amplitude or content. |
2 | Display the phase value of each harmonic. |
3 | Display the waveform after harmonic superposition. |
4 | Display the phase of each harmonic. The spectrum of each harmonic is displayed visually in the form of histogram (fundamental wave is 100%). |
☆ Data Management |
|
• The equipment can store and manage the test data on site. • The selected data can be saved to the mobile USB flash disk. |
☆ Professional Test Software |
|
• Optional computer-specific testing software can organize and generate test reports from data exported via USB drive, improving testing efficiency. • Software functionality can be customized according to customer needs, with support for future software upgrades to accommodate new testing requirements. |
Type | Measurement Mode | Range | Resolution | Accuracy ± (ppm of reading + ppm of range) [1] | |
Class 0.05 | Class 0.02 | ||||
ACV | Direct Measurement | 120 V | 1 mV | 300 + 200 | 120 + 80 |
240 V | 1 mV | 300 + 200 | 120 + 80 | ||
480 V | 1 mV | 300 + 200 | 120 + 80 | ||
ACI | Direct Measurement | 1 A | 10 μA | 300 + 200 | 120 + 80 |
10 A | 100 μA | 300 + 200 | 120 + 80 | ||
Clamp Measurement | 1 A[2] | 10 μA | 0.2%*RG | 0.1%*RG | |
5 A | 10 μA | 0.2%*RG | 0.1%*RG | ||
10 A[2] | 100 μA | 0.2%*RG | 0.1%*RG | ||
20 A[2] | 100 μA | 0.2%*RG | 0.1%*RG | ||
100 A[2] | 1 mA | < 50 A: 0.2%*RG; ≥ 50 A: 0.5%*RG | < 50 A: 0.1%*RG; ≥ 50 A: 0.2%*RG | ||
Note [1]: (ppm = parts per million) (e.g., 10ppm = 0.001%). Note [2]: The 5A clamp meter is standard, while other models are optional. Optional clamp meters include 200 A and 1 kA models. |
• Voltage input: 10 V~576 V, 6-digits decimal display.
• Current input (direct measurement): 0.05 A~12 A, 6-digits display.
• Current input (clamp measurement): 0.1 A~1 kA, 6-digits display.
Frequency | Range | 45.000 Hz~65.000 Hz |
Resolution | 0.001 Hz | |
Accuracy (k=2) | ± 0.01 Hz | |
Phase | Range | 0.000°~359.999° |
Resolution | 0.001° | |
Accuracy (k=2) | Direct Measurement: ± 0.05° | |
Clamp Measurement: ± 0.2° |
Power/Energy Parameters | Accuracy (k=2) | |||
Direct Measurement | Clamp Measurement | |||
Class 0.05 | Class 0.02 | Class 0.05 | Class 0.02 | |
Active P/E | cos φ| ≥0.5 | ± 0.05%*FS[3] | ± 0.02%*FS[3] | ± 0.2%*FS[3] | ± 0.1%*FS[3] |
Reactive P/E | sin φ| ≥0.5 | ± 0.1%*FS[3] | ± 0.05%*FS[3] | ± 0.5%*FS[3] | ± 0.2%*FS[3] |
Apparent power | ± 0.1%*FS[3] | ± 0.05%*FS[3] | ± 0.5%*FS[3] | ± 0.2%*FS[3] |
Power factor | ± 0.0005 | ± 0.0005 | ± 0.002 | ± 0.002 |
Note [3]: FS=voltage range value x current range value. |
• Power factor measurement range: -1.000 0... 0.000 0... 1.000 0.
• Standard electric energy pulse output: high frequency full range value corresponds to 60 kHz, low frequency full range value corresponds to 6 Hz.
• Standard electric energy pulse input: frequency ≤ 200 kHz, voltage: 0... 3.3 V... 24 V.
Power Supply | Mains power: AC (100 ~ 264) V, (50 ± 2) Hz; Lithium battery power: Supports continuous operation for 8 hours. |
Maximum power | 30 VA |
Temperature Performance | Working temperature: -20°C~45°C Storage temperature: -20°C~70°C |
Humidity Performance | Working humidity: < 80% @ 30°C, < 70% @ 40°C, < 40% @ 50°C Storage humidity: (20%~80%) R·H, non-condensing |
Altitude | <3000 m |
Weight | About 8kg |
Interface | USB, LAN |
Dimensions | 410 mm(W) x 320 mm(D) x 170 mm(H) |
|
| |
| |
| |
| |
Advanced Multifunction Calibrator Advanced Electrical Reference Standards Electric Vehicle (EV)Charging Facility Tester AC/DC Standard Source /Meter Single-phase/Three-phase Standard Meter Magnetic Parameters Measurement and Calibration Magnetic Material Testing Equipment Electrical Steel Testing Equipment