1. Summary
TD1400 is a portable instrument dedicated to measuring the loopresistance andcontact resistanceof switches, circuit breakers, copper-aluminum busbars and other equipment, with an accuracy of Class 0.2/ Class 0.5 available.
2. Features
• Resistance measurement range: 0.01 mΩ ~ 180 mΩ;
• Support long-term output of 2A~100A (optional 200A) DC current.
• LCD touch screen
• Built-in micro printer for on-site printing of test results.
• Can work under -20°C ~ 45°C ambient conditions.
• Supports AC (198 V ~ 253 V) wide voltage input range.
• RS232 interface.
☆ Measure loop resistance / contact resistance |
|
• With a built-in constant current source, the instrument can continuously output 100 A (200A selection) large current; • The four-wire wiring method can effectively eliminate the measurement error caused by the resistance of the test wire; • It has a combination of voltage and current test wire. |
3. Characteristics
☆ High stability constant current source |
|
• The instrument adopts the design principle of closed-loop negative feedback system, which can feedback and adjust the output power value in real time to ensure real-time stability. •The typical stability of the DC current output is 0.02%/min. |
4. Specifications
Test the current | Resistance measurement range | Resolution | Measurement uncertainty (k=2) (%*RD + μΩ) [1]. | |
Class 0.5 | Class 0.2 | |||
20 A | 0 ~ 180 mΩ | 10 mΩ | 0.5 + 10 | 0.2 + 10 |
50 A | 0 ~ 60 mΩ | 1 mΩ | 0.5 + 10 | 0.2 + 10 |
100 A | 0 ~ 25 mΩ | 1 mΩ | 0.5 + 10 | 0.2 + 10 |
200 A[2] | 0 ~ 5 mΩ | 0.1 mΩ | 0.5 + 10 | 0.2 + 10 |
Note [1]: RD is the reading value; [2]: 200 A range is optional |
• Current output range: 2 A ~ 120 A (or 240A).
5. General Specifications
Power supply | AC ( 220± 22 )V,( 50±2 ) Hz |
Maximum power consumption | 1000 VA |
Warm-up time | ≥ 30 min |
Temperature performance | Working temperature: -20 °C ~ 45°C Storage temperature: -20 °C ~ 55°C |
Humidity performance | Operating humidity: ( 20% ~ 85% ) R·H, no condensation Storage humidity: < 95% R· H, no condensation |
Communication interface | RS232 |
Quality | Approx 8 kg |
Size | 380 mm(W)X 260 mm(D)X 180 mm(H) |
|
6. Ordering Information
Advanced Multifunction Calibrator Advanced Electrical Reference Standards Electric Vehicle (EV)Charging Facility Tester AC/DC Standard Source /Meter Single-phase/Three-phase Standard Meter Magnetic Parameters Measurement and Calibration Magnetic Material Testing Equipment Electrical Steel Testing Equipment