1. Summary
TA4300 is a 3PH power quality calibrator. It integrates voltage/current output function,harmonic function,AC power output function,AC energy test function. It can test voltage fluctuate,flicker,harmonic,interharmonic,voltage swell,voltage dip and 3PH imbalance for power quality analyzer.
2. Features
• 3PH Voltage Output:1 V~528 V
• 3PH Current Output:0.2 mA~22 A
• Frequency:45 Hz~65 Hz (400 Hz as option)
• U-I phase:0°~360°
• Voltage/Current Accuracy Class:0.02
• Harmonic:2nd~99st output
• Interharmonic:0.1~99.9 times output
• Flicker,Voltage Swell,Voltage Dip and Short Interruption Functions
• Electrical Energy Pulse Input/Output Function
• LCD touch screen
• Testing software(Option)
3. Application
4. Characteristics
☆ Wide Output |
|
• It can meet most three-phase /single-phase power quality meters. |
☆ Multiple Output Mode |
|
• “Direct output”mode,User can set output value by physical key or touch screen. • Three-phase unit adjustment or single phase adjustment. |
☆ Multiple Output Mode |
|
• Touch “Calibration point”of screen for “% setting”. |
|
• “Rotary Knob”mode,User can setting in clockwise direction or anticlockwise direction. |
☆ Harmonics Function | |
| |
S/N | Function Description |
| User can select test function for harmonic,interharmonic,flicker,voltage swell/dip,three-phase imbalance. |
| User can set 2nd~99st harmonics output,and set amplitude or percentage. |
| Oscilloscope function, user can observe voltage and current waves. |
| Display frequency spectrum of harmonic by histogram.(fundament wave is 100%) |
☆ Interharmonic / Flicker Function | |
| |
S/N | Function declaration |
| Set up interharmonic number. |
| Set up amplitude of interharmonic and select voltage and current channel. |
| Set up flicker,select sine wave or square wave by click,and set up duty factor of square wave,frequentness,Pst. |
| Set up amplitude of flicker and select voltage channel. |
☆ Voltage Drop/ Interruption and Three-Phase Imbalance | |
| |
S/N | Function declaration |
| Set up duration time and drop circulation of voltage,set up duration time before drop and after drop,singel drop or times by circle. |
| Set up amplitude of drop and select voltage channel. |
| Set up voltage/current of three-phase imbalanc. |
| Set up three-phase imbalance. |
☆ Electric Quantity Analysis | |
| |
S/N | Function declaration |
| Multi-electric quantity display area for real-time display A/B/C phase multi-group electric quantity values. |
| Phasor diagram,display phasor diagram for voltage and current. |
| Trend chart,display trend of selector electric quantity. |
| Analysis Max value,Min value,P-P:Max- Min,Avg value and standard variance. |
☆ Testing Software(Option) |
|
• USB, LAN, RS232 communication interface,software customizable. |
6. Specifications
6.1 Three-Phase Voltage / Current Output
Voltage Range | Resolution | Stability ( %/min ) | Accuracy ±(ppm of reading + ppm of range) [1] | Max Burden (mA) |
60 V | 0.01 mV | 0.005 | 120+80 | 240 |
120 V | 0.1 mV | 0.005 | 120+80 | 120 |
240 V | 0.1 mV | 0.005 | 120+80 | 60 |
480 V | 0.1 mV | 0.005 | 120+80 | 30 |
Note [1] : (ppm = parts per million) (e.g.,10ppm = 0.001%). |
Current Range | Resolution | Stability ( %/min ) | Accuracy ±(ppm of reading + ppm of range) [1] | Max Burden (V) |
200 mA | 0.1 μA | 0.005 | 120+80 | 50 |
1 A | 1 μA | 0.005 | 120+80 | 20 |
5 A | 1 μA | 0.005 | 120+80 | 6 |
20 A | 10 μA | 0.005 | 120+80 | 1.5 |
• Output range:1 V~528 V,0.2 mA~22 A;manual/automatic range switch;
• Degree of distortion:< 0.1%,voltage / current < 0.02%;
• Voltage short circuit, current open-circuit and overload protection
6.2 Frequency / Phase
Frequency | Range:45 Hz~65 Hz (400 Hz as option); Adjustment fineness:0.001 Hz; Accuracy:±0.01 Hz |
Phase | Range:0.000 0°~359.999 9° Adjustment fineness:0.001° Accuracy:±0.02° |
6.3 Three-Phase Power Output
Type | Stability ( %/min ) | Accuracy ( ± %*FS)[2] |
Active power |cosφ|≥0.5 | 0.005 | 0.05 |
Reactive power |sinφ|≥0.5 | 0.01 | 0.05 |
Apparent power | 0.01 | 0.05 |
Power factor | 0.01 | 0.05 |
Note [1] : FS= voltage range x current range |
• Power factor range:-1.000 000…0.000 000…1.000 000
6.4 AC Energy Testing
Type | Accuracy |
Active energy | ± 0.05%*reading |
Reactive energy | ± 0.1%* reading |
• Energy pulse output:full range for 60 kHz
• Energy pulse input:max frequency is 200 kHz,pulse level:3 V~24 V
• Constant setting of energy meter:1…1000000imp./kwh or 1…1000000imp./ws
6.5 Harmonic / Interharmonic / Flicker / Voltage Swell or Dip
Harmonic | Voltage/current : 2nd~99st output;Amplitude: 0~40%;Phase: 0~359.99°;Accuracy:± 0.1% |
Interharmonic | Voltage/current : 0.1~99.9 time output;Amplitude: 0~40%;Phase: 0~359.99°;Accuracy:± 1% |
Flicker | Flicker momentum range:0~40 %;Frequency range:0.1~200 Hz; Modulation mode:sine or square wave;Accuracy:± 1% |
Voltage Swell or Dip & Short Interruption | Output range:0~120%*Range;time of swell/dip:0~120s; Duration time:1ms~300s |
7. General Specifications
Power Supply | AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz |
Temperature Performance | Working temperature:0°C~45°C; Storage temperature:-20°C~70°C |
Humidity Performance | Working humidity:< 80% @ 30°C,< 70% @ 40°C,< 40% @ 50°C Storage humidity:(20%~80%) R·H,non-condensing |
Interface | USB、LAN、RS232 |
8. Ordering Information
Advanced Multifunction Calibrator Advanced Electrical Reference Standards Electric Vehicle (EV)Charging Facility Tester AC/DC Standard Source /Meter Single-phase/Three-phase Standard Meter Magnetic Parameters Measurement and Calibration Magnetic Material Testing Equipment Electrical Steel Testing Equipment