Home

Products

Advanced Multiproduct/Multifunction Calibrator

EV Charger/EV Charging Station Tester and Loads

Advanced Electrical Reference Standards

Electrical Resistance Standard

AC DC Current Source and Meter

DC Power/Energy Standards

Wideband Power Source/Meter

Single-phase/Three-phase Electrical Meter

Single-phase/Three-phase Energy Meter standard Calibration Device

Safety Instrument Detection

High Voltage Standard Instrument

Industrial Instrument Testing

Calibrator for ESU Analyzer and Defibrillator Analyzer

Magnetic Parameters Measurement and Calibration

Electrical Steel Testing Equipment

Magnetic Material Testing Equipment

Videos

News

Solution

Download

About Us

Contact Us

You are here: Home » News » Product News » Energy Meter Type Evaluation Solution 3-TD3720 Load Current Rapid Change Test Device

Energy Meter Type Evaluation Solution 3-TD3720 Load Current Rapid Change Test Device

Publish Time: 2024-08-08     Origin: Site

* Reference only. Configuration and details may vary depending on the application scenario.。

1. TD3720 Load Current Rapid Change Test Device

TD3720 is a set of equipment used to test the impact of rapid changes in load current on the error of three-phase electric energy meters. It is mainly composed of a three-phase electric energy meter calibration device (whose current channel can be quickly shut down and connected), system and software, etc. The device accuracy is class 0.05.


2. Features

The three-phase standard source has a maximum output of 456 V / 120 A, which can ensure the stability of the value output under full load conditions.

The user can set the duration of ton and toff, the total duration of a single test, etc. The switching of off and on is completed within a nominal frequency cycle, a single test can be repeated for more than 4 hours, and multiple tests can be performed continuously.

Supports AC voltage sag and short-term interruption tests for voltage circuit power supply.

Zero-crossing switching: The current can be set to complete the switch switching at the zero-crossing point, which can ensure the accuracy of the load rapid change signal.

By reading the cumulative electric energy value of the standard device and comparing it with the cumulative electric energy value of the DUT, the error deviation of the DUT under the condition of rapid change of load current is measured.

Mobile measurement station: The value output can be observed or controlled through its LCD touch screen, which is convenient for manual testing.

Automatic test software: It can complete the automatic test records of the inherent error of the energy meter, the initial indication of the energy meter, the end indication of the energy meter, the energy meter metering, the energy meter metering of standard device, the tested pulse, the metering error and other items. The whole process is automatic and the test report is exported.

Breakpoint resume function: After the device is powered off and restarted, the software supports detection from the current error point.

Crimping wiring: Equipped with a crimping wiring device, it can automatically crimp the energy meter that meets the standard type requirements of the State Grid and the Southern Grid, which is convenient for users to quickly check the meter.

Multi-meter position calibration test bench: 3/6 positions are optional, and the test bench is equipped with interfaces such as power output, energy pulse, and communication.

Single-three-phase compatibility (optional): Optional compatible single-phase meter testing


3. Specification

3.1 Three-phase Voltage Output

Range

Resolution

Short-term Stability

( % / min )

Measurement Uncertainty (k=2)

( ppm*RD+ppm*RG)

Distortion ( % )

60 V

0.1 mV

0.01

200 + 50

< 0.3

120 V

1 mV

0.01

200 + 50

< 0.3

240 V

1 mV

0.01

200 + 50

< 0.3

380 V

1 mV

0.01

200 + 50

< 0.3

Note: ① RD is the reading, RG is the range, the same below

Output range: 6 V ~ 456 V, 6-digit display, load capacity: 15 VA / (phasemeter position)

Symmetry: voltage ≤ 0.2%, phase ≤ 0.5°

Protection function: short circuit protection, overload protection

3.2 Three-phase Current Output

Range

Resolution

Short-term Stability

( % / min )

Measurement Uncertainty (k=2)

( ppm*RD+ppm*RG)

Distortion ( % )

100 mA

1 μA

0.01

200 + 50

< 0.5

200 mA

1 μA

0.01

200 + 50

< 0.5

500 mA

1 μA

0.01

200 + 50

< 0.5

1 A

10 μA

0.01

200 + 50

< 0.5

2 A

10 μA

0.01

200 + 50

< 0.5

5 A

10 μA

0.01

200 + 50

< 0.5

10 A

100 μA

0.01

200 + 50

< 0.5

20 A

100 μA

0.01

200 + 50

< 0.5

50 A

100 μA

0.01

200 + 50

< 0.5

100 A

1 mA

0.01

200 + 50

< 0.5

Output range: 10 mA ~ 120 A, 6-digit display, load capacity: 30 VA / (phasemeter position)

Symmetry: current ≤ 0.5%, phase ≤ 0.5°

Protection function: open circuit protection, overload protection.


3.3 Frequency/Phase

Measurement

Class 0.05

Frequency

Measuring range

45 Hz~65 Hz

Minimum resolution

0.000 01 Hz

Measurement uncertainty (k=2)

0.005%*RD

Phase

(I ≥ 50mA)

Measuring range

0~360°

Minimum resolution

0.000 1°

Measurement uncertainty (k=2)

0.012°


3.4 Power/Energy

Voltage Range

Current Range

Power Factor

Measurement Uncertainty (k=2)

Class 0.05

30 V ≤ U ≤ 380 V

50 mA ≤ I ≤120 A

0.5L~1~0.5C

0.05%*RD

10 mA ≤ I < 50 mA

1

0.05%*RD

0.5L~1~0.5C

0.08%*RD

Output power stability: 0.02% / 2min;

Power/ energy measurement range: combination of AC voltage range and AC current range;

Power factor measurement range: -1.000 00…0.000 00…1.000 00;

Standard electric energy pulse output: high frequency full range value corresponds to 60 kHz, low frequency full scale range corresponds to 6 Hz;

Standard electric energy pulse input: frequency ≤ 200 kHz, voltage: 0…3.3 V…24 V;

Energy error display: automatic, with a resolution of 0.0001%.


4. General Specification

Power Supply

Single-phase AC ( 220 ± 22 ) V, ( 50 ± 2 ) Hz

Power consumption

/

Warm-up Time

30mins

Temperature Performance

Working temperatures: 5°C~45°C;

Storage temperature: -10°C~55°C

Humidity Performance

Working humidity: < 80% @ 30°C, < 70% @ 40°C, < 40% @ 50°C

Storage humidity: (20%~80%) RH, no condensation

Altitude

< 3000 m

Communication Interface

RS232, RS485, LAN

Test Bench Dimensions

(L*W*H /total height, mm)


Positions

3-position

6-position

Crimp wiring

1350*750*750/1395

1850*750*750/1395

Computer Table Dimensions (optional)

((L*W*H, mm)

1000*800*750


5. Ordering Information