TS1300 is a special instrument for measuring the AC/DC magnetic properties of electrical steel. It consists of excitation and measurement host (AC and DC), Epstein frame, Single Sheet Tester (optional), computer software, etc. It is suitable for comprehensive analysis of AC magnetic characteristics of electrical steel or core.
• Electrical parameters calibration function.
• Measurement of magnetic field by M.C and H-Coil (optional).
• Maximum magnetization field reaches 30 kA/m (700 T Epstein frame).
• The maximum frequency can be selected to 20kHz.
• 2 ~ 63 harmonics influence measurement (optional).
• Slowly reduce amplitude demagnetization for specimen.
• Software controlled automatic measurement.
• Automatic multi-point measurement.
• The software has been configured with multiple measurement schemes
• Automatic calculates magnetic parameters and curves.
• Modular design, easy to upgrade or maintenance.
• Ultra-wide range of current continuously and stably regulated.
• Switch the Permeameter by software without repeated wiring.
• IEC 60404-2 "Magnetic materials-Part 2 Methods of measurement of the magnetic properties of electrical steel strip and sheet by EP".
• IEC 60404-3 "Magnetic materials-Part 3 Methods of measurement of the magnetic properties of electrical steel strip and sheet by SST".
• IEC 60404-10 "Magnetic materials-Part 10-Methods of measurement of magnetic properties of electrical steel strip and sheet at medium frequencies".
Frequency | Parameter | Range[1] | Uncertainty (k=2) [6] | Repeatability[6] |
DC | Hm | 1 A/m...30 kA/m | 0.5% | 0.2% |
Jm | 10 mT...2.0 T | 1.0% | 0.3% | |
Bm | 10 mT...2.1 T | 1.0% | 0.3% | |
Br | —— | 1.0% | 0.3% | |
Hc | —— | 1.0% | 0.5% | |
μi | —— | 4.0% | 1.5% | |
μm | —— | 2.0% | 1.0% | |
<50Hz | Hm | 5 A/m...10 kA/m / 30 kA/m[5] | 0.5% | 0.2% |
Jm / Bm | B5[2]... B10000 / B30000[5] | 0.5% | 0.2% | |
Ps | P0.1[3]... P1.9 ( P1.7 ) | 1.0% | 0.3% | |
50Hz / 60Hz | Hm | 5 A/m...10 kA/m / 30 kA/m[5] | 0.3% | 0.2% |
Jm / Bm | B5[2]... B10000 / B30000[5] | 0.3% | 0.1% | |
Ps | P0.1[3]... P1.9 ( P1.7 ) | 0.5% | 0.2% | |
400 Hz | Hm | 5 A/m...2.5 kA/m | 0.5% | 0.2% |
Jm / Bm | B5[2]... B2500 | 0.5% | 0.2% | |
Ps | P0.1[3]... P1.7 ( P1.5 ) | 1.0% | 0.5% | |
1 kHz | Hm | 5 A/m...800 A/m | 1.0% | 0.3% |
Jm / Bm | B5[2]... B800 | 1.0% | 0.3% | |
Ps | P0.1[3]... P1.5 ( P1.0 ) | 2.0% | 0.5% | |
2.5 kHz | Hm | 5 A/m...200 A/m | 1.0% | 0.3% |
Jm / Bm | B5[2]... B200 | 1.0% | 0.3% | |
Ps | P0.1[3]... P1.0 | 2.0% | 0.5% | |
10 kHz | Hm | 5 A/m...100 A/m | 2.0% | 0.5% |
Jm / Bm | B5[2]... B100 | 2.0% | 0.5% | |
Ps | P0.1[3]... P0.5 | 4.0% | 1.0% | |
20 kHz | Hm | 5 A/m...50 A/m | 2.0% | 0.5% |
Jm / Bm | B5[2]... B50 | 2.0% | 0.5% | |
Ps | P0.1[3]... P0.2 | 4.0% | 1.0% |
• Conditions :(23 ± 5)℃, demagnetized before measurement.
• Accuracy: 1%~2% deviation will occur according to the way and position of loading the sample in the measuring frame.
• Repeatability: put the same specimens in the specified position and keep it from moving, and measure it repeatedly for 10 times. Take the maximum or minimum value of 10 measurements minus the average value divided by the average value, and the maximum value of the two represents repeatability.
• Note: [1] The table only gives reference measuring points, and the indexes such as Hm, Jm/Bm, Ps that can be reached by the specimens during actual measure are related to the performance of the material, quality of the specimens, frequency, current, voltage, etc.
• Note: [2] B5 refers to the magnetic flux density value when the magnetic field H is 5 A/m, and the typical measuring points of non-oriented steel are in parentheses.
• Note: [3] P0.1 refers to the specific total loss value when flux B is 0.1T, and the typical measure point for non-oriented steel is in parentheses.
• Note: [4] For medium and high frequency measurement, select the corresponding Epstein frame.
• Note: [5] When the selected effective excitation power is 1500VA, the magnetic field can reach 30 kA/m with the corresponding Epstein frame.
• Note: [6] The uncertainty and repeatability are at typical measuring points, and are related to the performance of the measured material.
Output Capability | Voltage | 0.1 Vrms~200 Vrms | ||
Frequency | 3 Hz... 2.5 kHz...10 kHz...20 kHz (optional) | |||
Power | 0 ~ 500VA / 1500 VA Optional (see selection guide) | |||
Current | 500VA | 500 VA / 1500 VA (optional) | ||
1500VA | 0.5 mApk ~ 15 Apk | |||
Harmonic | 2... 63 (optional) | |||
Specification | Frequency (Hz) | Uncertainty of Effective Value of Voltage and Current(k=2) | Uncertainty of Active Power(k=2) | |
3 ≤ F < 45 | 0.1%*RG [7] | 0.2%*FS [8] | ||
45 ≤ F ≤ 65 | 0.05%*RG [7] | 0.1%*FS [8] | ||
65 ≤ F ≤ 400 | 0.05%*RG [7] | 0.15%*FS [8] | ||
400 < F ≤ 1 k | 0.1%*RG [7] | 0.2%*FS [8] | ||
1 k < F ≤ 2.5 k | 0.2%*RG [7] | 0.5%*FS [8] | ||
2.5 k < F ≤ 10 k | 0.5%*RG [7] | 1.0%*FS [8] | ||
10 k < F ≤ 20 k | 1.0%*RG [7] | 2.0%*FS [8] |
• Note: [7] RG is range;
• Note: [8] FS= Voltage range x current range, when the phase Angle is within 75 degrees.
Output Capacity | Voltage | ± 0.1 V ~ ± 20 V | |
Current | ± 0.5 mA ~ ± 40 A | ||
Power | 0 ~ 800 W | ||
Adjustment Fineness of Current | 0.005%*RG | ||
Specification | Current | Uncertainty(k=2) | 0.2%*RG |
Flux | Range | 5 mWb, 100 mWb, 350 mWb | |
Zero-drift | ± (0.05%*RG)/min or ± 1 μWb/min (Whichever is greater) | ||
Uncertainty(k=2) | 0.3%*RG+ 5 μWb |
Advanced Multifunction Calibrator Advanced Electrical Reference Standards Electric Vehicle (EV)Charging Facility Tester AC/DC Standard Source /Meter Single-phase/Three-phase Standard Meter Magnetic Parameters Measurement and Calibration Magnetic Material Testing Equipment Electrical Steel Testing Equipment