TY1000 Magnetic Properties Measuring System for Hard Magnetic Materials design conforms to the standard IEC 60404-5, GB/T 3217-2013. It consists of excitation and measurement host, electromagnet, magnetic measurement sensors, computer software, etc. It is suitable for measuring the magnetic properties of various types of hard magnetic materials and drawing hysteresis loops and curves. TY1000 has the advantages of convenient operation, fast measurement, good repeatability and high reliability.
Availability: | |
---|---|
Quantity: | |
TY1000
TUNKIA
*Above picture is only for reference, subject to actual delivery
1. Summary
TY1000 design conforms to the standard IEC 60404-5, GB/T 3217-2013. It consists of excitation and measurement host, electromagnet, magnetic measurement sensors, computer software, etc. It is suitable for measuring the magnetic properties of various types of hard magnetic materials and drawing hysteresis loops and curves. TY1000 has the advantages of convenient operation, fast measurement, good repeatability and high reliability.
2. Features
• Electrical parameters calibration function.
• Slowly reduce amplitude demagnetization for sample.
• Zero drift of flux meter:<1 μWb / min.
• Low nonlinear error of the hall probe.
• Continuously adjustable bipolar magnetization current source.
• Use flux meter and coils to measure B or J.
• Use magnetometer and hall-probe or flux meter and coils or to measure H.
• Automatic compensate the residual area-turns of J-coils by software.
• Various types of coils or probes are optional.
• Coils can be customized according to specimen size.
• Hmax of the 150 mm electromagnet is up to 2.5 T.
• Hmax of the 130 mm electromagnet is up to 2.45 T.
• Typical single measurement time is 30 s.
• Computer controlled automatic measurement.
• Automatic calculates magnetic parameters and curves.
• Modular design, easy to upgrade or maintenance.
3. Application
*Above picture is only for reference, subject to actual delivery
1. Summary
TY1000 design conforms to the standard IEC 60404-5, GB/T 3217-2013. It consists of excitation and measurement host, electromagnet, magnetic measurement sensors, computer software, etc. It is suitable for measuring the magnetic properties of various types of hard magnetic materials and drawing hysteresis loops and curves. TY1000 has the advantages of convenient operation, fast measurement, good repeatability and high reliability.
2. Features
• Electrical parameters calibration function.
• Slowly reduce amplitude demagnetization for sample.
• Zero drift of flux meter:<1 μWb / min.
• Low nonlinear error of the hall probe.
• Continuously adjustable bipolar magnetization current source.
• Use flux meter and coils to measure B or J.
• Use magnetometer and hall-probe or flux meter and coils or to measure H.
• Automatic compensate the residual area-turns of J-coils by software.
• Various types of coils or probes are optional.
• Coils can be customized according to specimen size.
• Hmax of the 150 mm electromagnet is up to 2.5 T.
• Hmax of the 130 mm electromagnet is up to 2.45 T.
• Typical single measurement time is 30 s.
• Computer controlled automatic measurement.
• Automatic calculates magnetic parameters and curves.
• Modular design, easy to upgrade or maintenance.
3. Application
4. Excitation and Measurement
Magnetizing Current | Output Range | 0~± 25 A |
Max. Power | 1500 VA (peak) | |
Protection Function | Over voltage, over current, overload, overheating | |
Tesla Meter | Range | 0~4000 mT [1] |
Uncertainty(k=2) | 0.5%*RG within 2500 mT [2] | |
Characteristic | Hall-probe nonlinear correction function | |
Flux Meter | B Flux Meter Range | 0~20 mWb |
H Flux Meter Range | 0~40 mWb | |
Min. Resolution | 0.1 μWb | |
Zero-Drift | ≤ ± 1 μWb / min, auto-zero | |
Note:[1] Calibration only covers up to 2500mT. [2] RG is the range values. |
5. Reproducibility
Magnetic Parameters | Typical Values | Guaranteed Value (Al Ni Co) | Guaranteed Value (Ferrite, Fe-Cr-Co, Rare Earth Magnet ) |
Br | 0.3% | 1.0% | 2.0% |
Hcb | 0.3% | 1.0% | 2.0% |
Hcj | 0.3% | 1.0% | 2.0% |
(BH)max | 0.3% | 1.5% | 3.0% |
• Testing temperature:23℃ ± 5℃.
• Specimen size should strictly follow the requirements of IEC 60404-5.
6. General Specifications
Power Supply | AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz |
Temperature Performance | Working Temperature:0°C~40°C; Storage Temperature:-20°C~70°C |
Humidity Performance | Working Humidity:40%~80% R·H,No condensation Store Humidity:< 95% R·H,No condensation |
7. Electromagnet (optional)
Model | Type | Air gap | Imax | Hmax (10 mm Gap) |
TD8750A-13-1 | Vertical | 0~80 mm adjustable | DC,± 25 A | 2.45 T |
TD8750A-15-1 | Vertical | 0~80 mm adjustable | DC,± 25 A | 2.50 T |
Poles Type | Model | Pole Diameter | Pole Face Diameter | Hmax (10 mm Gap) |
Room temperature | TD8755A-13-13 | 130 mm | 130 mm | 1.5 T |
TD8755A-13-6 | 130 mm | 60 mm | 2.45 T | |
TD8755A-15-15 | 150 mm | 150 mm | 1.5 T | |
TD8755A-15-8 | 150 mm | 80 mm | 2.50 T | |
High temperature | TD8755D-13-8 | 130 mm | 80 mm | 1.8 T |
Note: The above accessories are optional according to the testing requirements.
8. Configuration List
No. | Item Name | Number | Configuration | Note |
1 | TY1000 | 1 | Standard | |
2 | Hall probe and bracket | 1 | Standard | |
3 | J coils or J-H coils | 1 | Standard | According to specimen size |
4 | Testing software | 1 | Standard | Installation CD-ROM |
5 | Test wire and junction box | 1 | Standard | |
6 | TY1500 electromagnet | 1 | Optional | |
7 | TY1550 electromagnet poles | 1 | Optional | |
8 | TY1900 Specimen Heating Device | 1 | Optional | Used for high temperature performance testing |
9 | High temperature J-H coils | 1 | Optional | |
10 | Reference specimen | 1 | Optional | Including certificate |
11 | Segment poles | 1 | Optional | According to specimen size |
12 | Workbench | 1 | Optional | Third party Products |
13 | Computer | 1 | Optional | Third party Products |
14 | Printer | 1 | Optional | Third party Products |
• Note: The above is only for reference. The specific configuration list depends on the technical protocol.
4. Excitation and Measurement
Magnetizing Current | Output Range | 0~± 25 A |
Max. Power | 1500 VA (peak) | |
Protection Function | Over voltage, over current, overload, overheating | |
Tesla Meter | Range | 0~4000 mT [1] |
Uncertainty(k=2) | 0.5%*RG within 2500 mT [2] | |
Characteristic | Hall-probe nonlinear correction function | |
Flux Meter | B Flux Meter Range | 0~20 mWb |
H Flux Meter Range | 0~40 mWb | |
Min. Resolution | 0.1 μWb | |
Zero-Drift | ≤ ± 1 μWb / min, auto-zero | |
Note:[1] Calibration only covers up to 2500mT. [2] RG is the range values. |
5. Reproducibility
Magnetic Parameters | Typical Values | Guaranteed Value (Al Ni Co) | Guaranteed Value (Ferrite, Fe-Cr-Co, Rare Earth Magnet ) |
Br | 0.3% | 1.0% | 2.0% |
Hcb | 0.3% | 1.0% | 2.0% |
Hcj | 0.3% | 1.0% | 2.0% |
(BH)max | 0.3% | 1.5% | 3.0% |
• Testing temperature:23℃ ± 5℃.
• Specimen size should strictly follow the requirements of IEC 60404-5.
6. General Specifications
Power Supply | AC ( 220 ± 22 ) V,( 50 ± 2 ) Hz |
Temperature Performance | Working Temperature:0°C~40°C; Storage Temperature:-20°C~70°C |
Humidity Performance | Working Humidity:40%~80% R·H,No condensation Store Humidity:< 95% R·H,No condensation |
7. Electromagnet (optional)
Model | Type | Air gap | Imax | Hmax (10 mm Gap) |
TD8750A-13-1 | Vertical | 0~80 mm adjustable | DC,± 25 A | 2.45 T |
TD8750A-15-1 | Vertical | 0~80 mm adjustable | DC,± 25 A | 2.50 T |
Poles Type | Model | Pole Diameter | Pole Face Diameter | Hmax (10 mm Gap) |
Room temperature | TD8755A-13-13 | 130 mm | 130 mm | 1.5 T |
TD8755A-13-6 | 130 mm | 60 mm | 2.45 T | |
TD8755A-15-15 | 150 mm | 150 mm | 1.5 T | |
TD8755A-15-8 | 150 mm | 80 mm | 2.50 T | |
High temperature | TD8755D-13-8 | 130 mm | 80 mm | 1.8 T |
Note: The above accessories are optional according to the testing requirements.
8. Configuration List
No. | Item Name | Number | Configuration | Note |
1 | TY1000 | 1 | Standard | |
2 | Hall probe and bracket | 1 | Standard | |
3 | J coils or J-H coils | 1 | Standard | According to specimen size |
4 | Testing software | 1 | Standard | Installation CD-ROM |
5 | Test wire and junction box | 1 | Standard | |
6 | TY1500 electromagnet | 1 | Optional | |
7 | TY1550 electromagnet poles | 1 | Optional | |
8 | TY1900 Specimen Heating Device | 1 | Optional | Used for high temperature performance testing |
9 | High temperature J-H coils | 1 | Optional | |
10 | Reference specimen | 1 | Optional | Including certificate |
11 | Segment poles | 1 | Optional | According to specimen size |
12 | Workbench | 1 | Optional | Third party Products |
13 | Computer | 1 | Optional | Third party Products |
14 | Printer | 1 | Optional | Third party Products |
• Note: The above is only for reference. The specific configuration list depends on the technical protocol.