Availability: | |
---|---|
Quantity: | |
TI5800
TUNKIA
1. Summary
TI5800 integrates a precision voltmeter, precision ammeter, and programmable adjustable voltage source. It supplies power to a voltage or current sensor when testing it, and accurately measures the sensor's secondary output current Is or voltage Us. This instrument can be combined with a standard current source or voltage source to form a sensor testing device to test voltage or current sensors.
2. Feature
• Voltage Us Measuring Range:10 mV ~ 12 V.
• Current Is Measuring Range:1 mA ~ 1.1 A.
• Measurement Frequency: DC,AC 45 Hz~1 kHz(Optional).
• DC Measurement Uncertainty: 50 ppm.
• Built-in DC ± (5.0 V~50 V) adjustable power supply.
• Supports power consumption measurement of the tested sensor.
• Optional connection adapter to match the sensor interface of DB9.
• Optional 1~4 measurement channels.
3. Application
☆ Sensor power supply and measurement of secondary signals |
Sensor Secondary Signal Measurement Module l Built-in standard ammeter and standard voltmeter can accurately measure the secondary output Is or Us of the sensor. l It can be used with standard current sources or voltage source groups to test the proportional error of voltage or current sensors. Sensor Power Supply Module l Built-in DC ± (5 V ~ 50 V) adjustable power supply can test the power supply impact of the sensor under test. A standard ammeter is connected in series in the power supply module loop. It can measure the no-load or full-load power consumption of the sensor being tested. |
1. Summary
TI5800 integrates a precision voltmeter, precision ammeter, and programmable adjustable voltage source. It supplies power to a voltage or current sensor when testing it, and accurately measures the sensor's secondary output current Is or voltage Us. This instrument can be combined with a standard current source or voltage source to form a sensor testing device to test voltage or current sensors.
2. Feature
• Voltage Us Measuring Range:10 mV ~ 12 V.
• Current Is Measuring Range:1 mA ~ 1.1 A.
• Measurement Frequency: DC,AC 45 Hz~1 kHz(Optional).
• DC Measurement Uncertainty: 50 ppm.
• Built-in DC ± (5.0 V~50 V) adjustable power supply.
• Supports power consumption measurement of the tested sensor.
• Optional connection adapter to match the sensor interface of DB9.
• Optional 1~4 measurement channels.
3. Application
☆ Sensor power supply and measurement of secondary signals |
Sensor Secondary Signal Measurement Module l Built-in standard ammeter and standard voltmeter can accurately measure the secondary output Is or Us of the sensor. l It can be used with standard current sources or voltage source groups to test the proportional error of voltage or current sensors. Sensor Power Supply Module l Built-in DC ± (5 V ~ 50 V) adjustable power supply can test the power supply impact of the sensor under test. A standard ammeter is connected in series in the power supply module loop. It can measure the no-load or full-load power consumption of the sensor being tested. |
4. Specification
4.1 DC Voltage/Current Measurement
Current Range | 10 mA, 100 mA, 1 A,manual/automatic range switching. |
± (1 mA ~ 1.1 A) | |
Voltage Range | 100 mV, 1 V, 10 V, manual/automatic range switching. |
± (10 mV ~ 12 V) | |
Measurement Uncertainty(k=2) | 0.003%*RD + 0.002%*RG [1] |
Temperature Coefficient@ (0°C~40°C) | 2 ppm*RD/°C |
Display | 7-digit decimal |
Note | [1]: RD is the reading value, RG is the range value, the same below. |
4.2 AC Voltage/Current Measurement (Optional)
Current Range | 10 mA, 100 mA, 1 A,manual/automatic range switching. |
1 mA ~ 1.1 A | |
Voltage Range | 100 mV, 1 V, 10 V,manual/automatic range switching. |
10 mV ~ 12 V | |
Measurement Frequency | 45 Hz ~ 1 kHz |
Measurement Uncertainty(k=2) | 0.006%*RD + 0.004%*RG |
Temperature Coefficient@ (0°C~40&°C) | 5 ppm*RD/°C |
Display | 7-digit decimal |
5. General Specification
Power Supply | AC ( 220 ± 22 ) V,(50 ± 2) Hz |
Maximum Power Consumption | 100 VA |
Warm-up Time | 2 times the last predicted warm-up time, up to 30 minutes. |
Working Environment | Temperature: 0°C ~ 40&°C Humidity: 20%R H ~ 80%R H,no condensation. Others: No electromagnetic field interference. |
Storage Environment | Temperature: -20°C ~ 70°C Humidity: <95%R H,no condensation. |
Altitude | < 3000 m |
Communication Interface | RS232 x 1 |
6. Ordering Information
7. Accessory List
S/N | Picture | Name | Specification | Quantity | Note |
1 | Test leads | 1.5m, 1.5mm2, Φ4 gun insert red, black | Red 2, Black 2 | Standard | |
2 | Test leads | 1.5m, 1.5mm2, Φ4 gun inserted green | 1 | Standard | |
3 | Power Cable | 3m, AC 250V, 16A | 1 | Standard | |
4 | Glass Fuse | F1A, 250V | 3 | Standard | |
5 | Connect Adapter | DB9 to conventional terminal block | 1 | Optional | |
6 | Sensor Cable | 0.8m, DB9 female base | 1 | Optional | |
7 | Data Communication Lines | RS232 to USB2.0 0.8m, DB9 female base | 1 | Optional |
Note: The quantity of accessories in items 1, 2, 5, and 6 above is for single-channel specifications. Multi-channel specifications should be multiplied by the corresponding multiples.
4. Specification
4.1 DC Voltage/Current Measurement
Current Range | 10 mA, 100 mA, 1 A,manual/automatic range switching. |
± (1 mA ~ 1.1 A) | |
Voltage Range | 100 mV, 1 V, 10 V, manual/automatic range switching. |
± (10 mV ~ 12 V) | |
Measurement Uncertainty(k=2) | 0.003%*RD + 0.002%*RG [1] |
Temperature Coefficient@ (0°C~40°C) | 2 ppm*RD/°C |
Display | 7-digit decimal |
Note | [1]: RD is the reading value, RG is the range value, the same below. |
4.2 AC Voltage/Current Measurement (Optional)
Current Range | 10 mA, 100 mA, 1 A,manual/automatic range switching. |
1 mA ~ 1.1 A | |
Voltage Range | 100 mV, 1 V, 10 V,manual/automatic range switching. |
10 mV ~ 12 V | |
Measurement Frequency | 45 Hz ~ 1 kHz |
Measurement Uncertainty(k=2) | 0.006%*RD + 0.004%*RG |
Temperature Coefficient@ (0°C~40&°C) | 5 ppm*RD/°C |
Display | 7-digit decimal |
5. General Specification
Power Supply | AC ( 220 ± 22 ) V,(50 ± 2) Hz |
Maximum Power Consumption | 100 VA |
Warm-up Time | 2 times the last predicted warm-up time, up to 30 minutes. |
Working Environment | Temperature: 0°C ~ 40&°C Humidity: 20%R H ~ 80%R H,no condensation. Others: No electromagnetic field interference. |
Storage Environment | Temperature: -20°C ~ 70°C Humidity: <95%R H,no condensation. |
Altitude | < 3000 m |
Communication Interface | RS232 x 1 |
6. Ordering Information
7. Accessory List
S/N | Picture | Name | Specification | Quantity | Note |
1 | Test leads | 1.5m, 1.5mm2, Φ4 gun insert red, black | Red 2, Black 2 | Standard | |
2 | Test leads | 1.5m, 1.5mm2, Φ4 gun inserted green | 1 | Standard | |
3 | Power Cable | 3m, AC 250V, 16A | 1 | Standard | |
4 | Glass Fuse | F1A, 250V | 3 | Standard | |
5 | Connect Adapter | DB9 to conventional terminal block | 1 | Optional | |
6 | Sensor Cable | 0.8m, DB9 female base | 1 | Optional | |
7 | Data Communication Lines | RS232 to USB2.0 0.8m, DB9 female base | 1 | Optional |
Note: The quantity of accessories in items 1, 2, 5, and 6 above is for single-channel specifications. Multi-channel specifications should be multiplied by the corresponding multiples.