TH3000
TUNKIA
1. Summary
TH3000 is a high-precision, wide-range, multi-functional AC and DC power standard, the instrument integrates AC and DC standard voltage/current/small signal voltmeter, AC/DC energy measurement, ripple measurement, harmonic measurement and other functions.
2. Features
• DC power/energy accuracy: class 0.005.
• DC voltage accuracy: class 0.002; DC current accuracy: class 0.003.
• DC voltage measurement: 10 mV ~ 1150 V.
• Direct DC current measurement: 10 μ A ~ 120 A.
• Support the transducer to expand the current range to 500 A or more.
• DC small-signal voltage measurement: 0.1 m V ~ 12 V
• Support DC ripple, AC 2nd ~ 63rd harmonic measurement.
• Comprehensive statistics analysis of the measured electricity.
• Standard energy pulse input/output function.
• USB, RS232 and LAN interfaces.
• LCD touch screen.
3. Applications
Note: When using TH30000 to calibrate electrical measuring devices or resistors, it is necessary to use the orresponding test power supply.
• Calibrating energy metering device: the instrument can be used as a high precision electric energy standard meter, used to assess AC energy meter verification device, DC energy meter verification device, standard energy meter, electric vehicle charger tester and other equipment, its DC+AC measurement mode is also especially suitable for the verification and calibration of DC energy meter verification device with ripple detection function.
• Calibrate the electrical measurement standard sources: direct calibrate AC and DC voltage sources, current sources, power sources, etc.
• Calibrating electrical measuring instruments: it can be used as a standard meter, with the test power supply to verify or calibrate voltmeters, ammeters, resistance meters, power meters, power factor meters, frequency meters, phase meters, transmitters, sensors, etc.
1. Summary
TH3000 is a high-precision, wide-range, multi-functional AC and DC power standard, the instrument integrates AC and DC standard voltage/current/small signal voltmeter, AC/DC energy measurement, ripple measurement, harmonic measurement and other functions.
2. Features
• DC power/energy accuracy: class 0.005.
• DC voltage accuracy: class 0.002; DC current accuracy: class 0.003.
• DC voltage measurement: 10 mV ~ 1150 V.
• Direct DC current measurement: 10 μ A ~ 120 A.
• Support the transducer to expand the current range to 500 A or more.
• DC small-signal voltage measurement: 0.1 m V ~ 12 V
• Support DC ripple, AC 2nd ~ 63rd harmonic measurement.
• Comprehensive statistics analysis of the measured electricity.
• Standard energy pulse input/output function.
• USB, RS232 and LAN interfaces.
• LCD touch screen.
3. Applications
Note: When using TH30000 to calibrate electrical measuring devices or resistors, it is necessary to use the orresponding test power supply.
• Calibrating energy metering device: the instrument can be used as a high precision electric energy standard meter, used to assess AC energy meter verification device, DC energy meter verification device, standard energy meter, electric vehicle charger tester and other equipment, its DC+AC measurement mode is also especially suitable for the verification and calibration of DC energy meter verification device with ripple detection function.
• Calibrate the electrical measurement standard sources: direct calibrate AC and DC voltage sources, current sources, power sources, etc.
• Calibrating electrical measuring instruments: it can be used as a standard meter, with the test power supply to verify or calibrate voltmeters, ammeters, resistance meters, power meters, power factor meters, frequency meters, phase meters, transmitters, sensors, etc.
4. Characteristics
☆ Wide Range of Current Measurement |
• Direct current measurement: use a pair of quick connect terminals to directly measure 10 μ A ~ 120 A (DC);10 μA ~ 80 A (AC) wide range DC current. • Current conversion measurement: The instrument is equipped with a sensor-specific measurement interface, which can supply power to the sensor and measure its secondary signal at the same time, and can be connected to a precision current sensor (such as TH0540) to extend the current range to 500 A, 1 kA or even more, and directly display the size of the primary current of the measured sensor by setting the ratio of the current sensor on the instrument. |
☆ AC/DC Comprehensive Measurement |
• The instrument supports DC, AC and DC+AC measurement modes, AC fundamental frequency: 40 Hz ~ 1 kHz. • When measuring voltage or current, distortion and ripple measurements can be completed simultaneously, and the maximum measurement bandwidth of DC ripple is 5 kHz, and AC supports the measurement of the 2nd ~ 63rd harmonic. • The DC+AC measurement function is very suitable for detecting energy meter verification devices with ripple or harmonic output. |
☆ Precision Small-signal Voltage Measurement |
• Support 0.1 m V ~ 12 V (DC); 2 m V ~ 10.2 V (AC) small signal voltage measurement, equipped with ultra-low thermal electric potential terminals and test wires, and can improve measurement accuracy and withstand overload signals up to 150 Vpk at most. • Application scenario 1: With a stable high-current source and precision current sensor, it can be used to calibrate shunts or other I/V converters. • Application scenario 2: With its own precision voltage measurement function, it can be used to evaluate the indirect access DC energy meter verification device. |
☆ Power Calculation Function |
• The instrument has the function of power calculation, the user can connect the voltage divider, shunt, transformer, sensor and other range extenders, and set the scale k and offset b, measure the secondary output signal x of the extender through the instrument, and then calculate the actual value of the primary electricity to be measured according to the function f(x) = kx + b. |
☆ DC Ripple / AC Harmonic Measurement | |
S/N | Function |
With DC ripple test function, it can measure the ripple content in the range of 1 Hz ~ 5 kHz. The ripple waveform can be displayed in real time, and the rms value (RMS) and peak-to-peak value (P-P) of the ripple can be measured. | |
AC power measurement also has an oscilloscope function, which can display the waveform of the AC signal under test in real time. | |
Calculate two types of harmonic distortion: THD/T (harmonic relative to full wave) and THD/F (harmonic relative to fundamental wave). | |
The amplitude (RMS), content (%) and phase of the 2nd ~ 63rd harmonic of each phase voltage or current can be analyzed in real time. | |
Visualize the spectrum of each harmonic (100% fundamental) in the form of a histogram. |
☆ Data Statistical Analysis Function | |
S/N | Function |
Data statistical analysis: calculate the maximum (Max), minimum value (min), peak-peak (P-P), average value (AVG), standard deviation (S.dev), divergence (Div), etc. of the measured electricity. | |
Power Trend Test: Curve the power over time in real time during the test cycle. | |
Normal distribution histogram: Displays the distribution of the amount of electricity being counted during the test period. |
4. Characteristics
☆ Wide Range of Current Measurement |
• Direct current measurement: use a pair of quick connect terminals to directly measure 10 μ A ~ 120 A (DC);10 μA ~ 80 A (AC) wide range DC current. • Current conversion measurement: The instrument is equipped with a sensor-specific measurement interface, which can supply power to the sensor and measure its secondary signal at the same time, and can be connected to a precision current sensor (such as TH0540) to extend the current range to 500 A, 1 kA or even more, and directly display the size of the primary current of the measured sensor by setting the ratio of the current sensor on the instrument. |
☆ AC/DC Comprehensive Measurement |
• The instrument supports DC, AC and DC+AC measurement modes, AC fundamental frequency: 40 Hz ~ 1 kHz. • When measuring voltage or current, distortion and ripple measurements can be completed simultaneously, and the maximum measurement bandwidth of DC ripple is 5 kHz, and AC supports the measurement of the 2nd ~ 63rd harmonic. • The DC+AC measurement function is very suitable for detecting energy meter verification devices with ripple or harmonic output. |
☆ Precision Small-signal Voltage Measurement |
• Support 0.1 m V ~ 12 V (DC); 2 m V ~ 10.2 V (AC) small signal voltage measurement, equipped with ultra-low thermal electric potential terminals and test wires, and can improve measurement accuracy and withstand overload signals up to 150 Vpk at most. • Application scenario 1: With a stable high-current source and precision current sensor, it can be used to calibrate shunts or other I/V converters. • Application scenario 2: With its own precision voltage measurement function, it can be used to evaluate the indirect access DC energy meter verification device. |
☆ Power Calculation Function |
• The instrument has the function of power calculation, the user can connect the voltage divider, shunt, transformer, sensor and other range extenders, and set the scale k and offset b, measure the secondary output signal x of the extender through the instrument, and then calculate the actual value of the primary electricity to be measured according to the function f(x) = kx + b. |
☆ DC Ripple / AC Harmonic Measurement | |
S/N | Function |
With DC ripple test function, it can measure the ripple content in the range of 1 Hz ~ 5 kHz. The ripple waveform can be displayed in real time, and the rms value (RMS) and peak-to-peak value (P-P) of the ripple can be measured. | |
AC power measurement also has an oscilloscope function, which can display the waveform of the AC signal under test in real time. | |
Calculate two types of harmonic distortion: THD/T (harmonic relative to full wave) and THD/F (harmonic relative to fundamental wave). | |
The amplitude (RMS), content (%) and phase of the 2nd ~ 63rd harmonic of each phase voltage or current can be analyzed in real time. | |
Visualize the spectrum of each harmonic (100% fundamental) in the form of a histogram. |
☆ Data Statistical Analysis Function | |
S/N | Function |
Data statistical analysis: calculate the maximum (Max), minimum value (min), peak-peak (P-P), average value (AVG), standard deviation (S.dev), divergence (Div), etc. of the measured electricity. | |
Power Trend Test: Curve the power over time in real time during the test cycle. | |
Normal distribution histogram: Displays the distribution of the amount of electricity being counted during the test period. |
5. Specifications
5.1 AC and DC voltage measurement
Voltage range | Resolution | Measurement uncertainty (k=2), (ppm*RD + ppm*RG or V) [1]. | |||
DC | 40 Hz≤F≤200 Hz | 200 Hz<f≤1 khz | Temperature coefficient (°C) @(15 ~ 30)°C | ||
100mV | 10 nV | 20 + 2 m | 80 + 40 | 100 + 50 | 0.2 |
300 mV | 10 nV | 12 + 15 | 80 + 40 | 100 + 50 | 0.2 |
1 V | 0.1 μV | 12 + 8 | 80 + 40 | 100 + 50 | 0.1 |
3 V | 0.1 μV | 12 + 8 | 80 + 40 | 100 + 50 | 0.1 |
10 V | 1 μV | 12 + 8 | 80 + 40 | 100 + 50 | 0.1 |
30 V | 1 μV | 12 + 8 | 80 + 40 | 100 + 50 | 0.1 |
100 V | 10 μV | 12 + 8 | 80 + 40 | 100 + 50 | 0.1 |
300 V | 10 μV | 12 + 8 | 80 + 40 | 100 + 50 | 0.1 |
1000 V | 0.1 mV | 12 + 8 | 80 + 40 | 100 + 50 | 0.1 |
Note: [1] RD is the reading value and RG is the range value |
• Direct flow measurement range: ± (10 mV~1150 V), manual/automatic shifting.
• Intersection measurement range: 10 mV~800 V (RMS), manual/automatic shifting.
• Input bandwidth: 50 kHz
• Minimum resolution: 10 nV, 8-digit decimal number display.
• Input resistance: 1 MΩ
• Input protection: 1200 Vpk, continuous.
5.2 AC and DC current measurement
Current range | Resolution | Measurement uncertainty (k=2), (ppm*RD +ppm*RG or nA) | |||
DC | 40Hz≤F≤200Hz | 200Hz<f≤1 khz | Temperature coefficient (/°C) @ (15 ~ 30)°C | ||
100 μA | 10 pA | 30 + 3 nA | 150 + 50 | 200 + 50 | 0.1 |
300 μA | 10 pA | 30 + 10 nA | 150 + 50 | 200 + 50 | 0.1 |
1mA | 0.1 nA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
3 mA | 0.1 nA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
10 mA | 1 nA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
30 mA | 1 nA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
100 mA | 10 nA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
300 mA | 10 nA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
1 A | 0.1 μA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
3 A | 0.1 μA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
10 A | 1 μA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
30 A | 1 μA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
100 A | 10 μA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
200 A [2] | 10 μA | 30 + 10 | 150 + 50 | 200 + 50 | 0.1 |
500 A [2] | 10 μA | 30 + 10 | 150 + 50 | 200 + 50 | 0.1 |
Note [2]: 200 A and 500 A range currents are optional, and optional current sensors such as TH0540 must be used for range extension. |
• DC measuring range: ±(10 μA~120 A), manual/automatic shifting.
The mating sensor extends the DC current limit to 240 A or 600 A.
• AC measuring range: 10 μA ~ 80A (RMS), manual/automatic shifting.
Mating sensors extends the AC current up to 160 A or 400 A (RMS).
• Input bandwidth: 50 kHz
• Minimum resolution: 0.1nA, 8-digit decimal number display.
5.3 AC/DC small-signal voltage measurement
Voltage range | Resolution | Absolute measurement uncertainty (k=2), ± (ppm*RD + ppm*RG or μV). | |||
DC | 40Hz≤F≤200Hz | 200Hz<f≤1 khz | Temperature coefficient (/°C) @ (15 ~ 30)°C | ||
1 mV | 0.1n V | 50 + 0.3 μV | - | - | 0.5 |
5 mH | 0.1n V | 50 + 0.5 μV | - | - | 0.5 |
20 mV | 1 nV | 50 + 0.6 μV | 100 + 3.0 μV | 120 + 4.0 μV | 0.5 |
100 mV | 10 nV | 40 + 15 | 80 + 40 | 100 + 50 | 0.2 |
500 mV | 10 nV | 40 + 15 | 80 + 40 | 100 + 50 | 0.2 |
2 V | 0.1 mV | 30 + 10 | 80 + 40 | 100 + 50 | 0.1 |
10 V | 1 mV | 30 + 10 | 80 + 40 | 100 + 50 | 0.1 |
• DC measuring range: ± (0.1 mV~12 V), manual/automatic shifting.
• AC measurement range: 2 mV~10.2 V (RMS), manual/automatic shifting.
• Input bandwidth: 50 kHz
• Minimum resolution: 0.1 nV, 8-digit decimal number display.
• Input resistance: > 1 GΩ.
• Input protection: 150 Vpk, continuous.
5.4 Ripple measurement function
• Ripple measurement band: 1 Hz~5 kHz
• Ripple measurement uncertainty (k=2): 0.1%*RG@ (50 Hz~1 kHz), rms
• Ripple energy accuracy: ± 0.05%*RD@AC 300 Hz
5.5 Power/energy measurement
DC power / electrical energy | Measuring range | Combination of DC voltage and DC current |
Measurement uncertainty | DC voltage uncertainty + DC current uncertainty | |
AC power / electrical energy | Measuring range | The combination of AC voltage and AC current |
Measurement uncertainty | AC voltage uncertainty + AC current uncertainty, | PF|=1 | |
Electrical energy pulses | Standard energy pulse input | It can directly receive the photoelectric pulse of the energy of the meter to be checked |
Standard energy pulse output | The signal amplitude is 5 V at TTL level FH=120 kHz ×power value÷voltage range value÷current range value FL=120 kHz × power value÷ voltage range value÷ current range value÷ 10000 | |
Electrical energy error display | Automatic display with a resolution of 0.00001%, 8-bit display |
6. Ordering Information
7. General Specifications
Power supply | AC (220 ± 22 ) V, ( 50 ± 2 ) Hz, maximum power consumption: 100 VA |
Warm-up time | 30 minutes |
Temperature performance | Working temperature: 15°C ~ 35°C; Storage temperature: -10°C ~ 50°C |
Humidity performance | Operating humidity: < 80% @ 30°C, < 70% @ 40°C, < 40% @ 50°C Storage humidity: (20% ~ 80%) R·H, no condensation |
altitude | < 3000 m |
Quality | Approx. 11 kg |
Communication interface | RS232 |
Size | 400 mm(W) × 302 mm(D) × 206 mm(H) |
5. Specifications
5.1 AC and DC voltage measurement
Voltage range | Resolution | Measurement uncertainty (k=2), (ppm*RD + ppm*RG or V) [1]. | |||
DC | 40 Hz≤F≤200 Hz | 200 Hz<f≤1 khz | Temperature coefficient (°C) @(15 ~ 30)°C | ||
100mV | 10 nV | 20 + 2 m | 80 + 40 | 100 + 50 | 0.2 |
300 mV | 10 nV | 12 + 15 | 80 + 40 | 100 + 50 | 0.2 |
1 V | 0.1 μV | 12 + 8 | 80 + 40 | 100 + 50 | 0.1 |
3 V | 0.1 μV | 12 + 8 | 80 + 40 | 100 + 50 | 0.1 |
10 V | 1 μV | 12 + 8 | 80 + 40 | 100 + 50 | 0.1 |
30 V | 1 μV | 12 + 8 | 80 + 40 | 100 + 50 | 0.1 |
100 V | 10 μV | 12 + 8 | 80 + 40 | 100 + 50 | 0.1 |
300 V | 10 μV | 12 + 8 | 80 + 40 | 100 + 50 | 0.1 |
1000 V | 0.1 mV | 12 + 8 | 80 + 40 | 100 + 50 | 0.1 |
Note: [1] RD is the reading value and RG is the range value |
• Direct flow measurement range: ± (10 mV~1150 V), manual/automatic shifting.
• Intersection measurement range: 10 mV~800 V (RMS), manual/automatic shifting.
• Input bandwidth: 50 kHz
• Minimum resolution: 10 nV, 8-digit decimal number display.
• Input resistance: 1 MΩ
• Input protection: 1200 Vpk, continuous.
5.2 AC and DC current measurement
Current range | Resolution | Measurement uncertainty (k=2), (ppm*RD +ppm*RG or nA) | |||
DC | 40Hz≤F≤200Hz | 200Hz<f≤1 khz | Temperature coefficient (/°C) @ (15 ~ 30)°C | ||
100 μA | 10 pA | 30 + 3 nA | 150 + 50 | 200 + 50 | 0.1 |
300 μA | 10 pA | 30 + 10 nA | 150 + 50 | 200 + 50 | 0.1 |
1mA | 0.1 nA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
3 mA | 0.1 nA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
10 mA | 1 nA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
30 mA | 1 nA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
100 mA | 10 nA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
300 mA | 10 nA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
1 A | 0.1 μA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
3 A | 0.1 μA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
10 A | 1 μA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
30 A | 1 μA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
100 A | 10 μA | 20 + 10 | 100 + 50 | 150 + 50 | 0.1 |
200 A [2] | 10 μA | 30 + 10 | 150 + 50 | 200 + 50 | 0.1 |
500 A [2] | 10 μA | 30 + 10 | 150 + 50 | 200 + 50 | 0.1 |
Note [2]: 200 A and 500 A range currents are optional, and optional current sensors such as TH0540 must be used for range extension. |
• DC measuring range: ±(10 μA~120 A), manual/automatic shifting.
The mating sensor extends the DC current limit to 240 A or 600 A.
• AC measuring range: 10 μA ~ 80A (RMS), manual/automatic shifting.
Mating sensors extends the AC current up to 160 A or 400 A (RMS).
• Input bandwidth: 50 kHz
• Minimum resolution: 0.1nA, 8-digit decimal number display.
5.3 AC/DC small-signal voltage measurement
Voltage range | Resolution | Absolute measurement uncertainty (k=2), ± (ppm*RD + ppm*RG or μV). | |||
DC | 40Hz≤F≤200Hz | 200Hz<f≤1 khz | Temperature coefficient (/°C) @ (15 ~ 30)°C | ||
1 mV | 0.1n V | 50 + 0.3 μV | - | - | 0.5 |
5 mH | 0.1n V | 50 + 0.5 μV | - | - | 0.5 |
20 mV | 1 nV | 50 + 0.6 μV | 100 + 3.0 μV | 120 + 4.0 μV | 0.5 |
100 mV | 10 nV | 40 + 15 | 80 + 40 | 100 + 50 | 0.2 |
500 mV | 10 nV | 40 + 15 | 80 + 40 | 100 + 50 | 0.2 |
2 V | 0.1 mV | 30 + 10 | 80 + 40 | 100 + 50 | 0.1 |
10 V | 1 mV | 30 + 10 | 80 + 40 | 100 + 50 | 0.1 |
• DC measuring range: ± (0.1 mV~12 V), manual/automatic shifting.
• AC measurement range: 2 mV~10.2 V (RMS), manual/automatic shifting.
• Input bandwidth: 50 kHz
• Minimum resolution: 0.1 nV, 8-digit decimal number display.
• Input resistance: > 1 GΩ.
• Input protection: 150 Vpk, continuous.
5.4 Ripple measurement function
• Ripple measurement band: 1 Hz~5 kHz
• Ripple measurement uncertainty (k=2): 0.1%*RG@ (50 Hz~1 kHz), rms
• Ripple energy accuracy: ± 0.05%*RD@AC 300 Hz
5.5 Power/energy measurement
DC power / electrical energy | Measuring range | Combination of DC voltage and DC current |
Measurement uncertainty | DC voltage uncertainty + DC current uncertainty | |
AC power / electrical energy | Measuring range | The combination of AC voltage and AC current |
Measurement uncertainty | AC voltage uncertainty + AC current uncertainty, | PF|=1 | |
Electrical energy pulses | Standard energy pulse input | It can directly receive the photoelectric pulse of the energy of the meter to be checked |
Standard energy pulse output | The signal amplitude is 5 V at TTL level FH=120 kHz ×power value÷voltage range value÷current range value FL=120 kHz × power value÷ voltage range value÷ current range value÷ 10000 | |
Electrical energy error display | Automatic display with a resolution of 0.00001%, 8-bit display |
6. Ordering Information
7. General Specifications
Power supply | AC (220 ± 22 ) V, ( 50 ± 2 ) Hz, maximum power consumption: 100 VA |
Warm-up time | 30 minutes |
Temperature performance | Working temperature: 15°C ~ 35°C; Storage temperature: -10°C ~ 50°C |
Humidity performance | Operating humidity: < 80% @ 30°C, < 70% @ 40°C, < 40% @ 50°C Storage humidity: (20% ~ 80%) R·H, no condensation |
altitude | < 3000 m |
Quality | Approx. 11 kg |
Communication interface | RS232 |
Size | 400 mm(W) × 302 mm(D) × 206 mm(H) |