TD1880 Precision Multi-function Calibrator is a multi-function calibrator with wide range and high accuracy, integrates AC/DC voltage and current standard source, can output excellent DC\AC voltage, DC\AC current, DC\AC power, phase, frequency, and analog resistance and capacitance etc. It’s applied for calibrating 6.5 digits or less digital multimeters.
DC voltage output: ± (0~1020 V)
DC current output: ± (0 ~ 20.5 A)
AC voltage output: 1mV ~ 1020 V
AC current output: 29 μA ~ 20.5 A
Sinusoidal wave frequency: 10 Hz ~ 500 kHz
Simulated DC resistance: 0 Ω ~ 1100 MΩ
2 MHz square wave frequency output
Phase output
AC/DC power output
Simulated capacitance(optional)
8 kinds of thermocouple output and measurement (option)
6 kinds of thermal resistance output (option)
Clamp meter calibration (optional)
☆ Calibrate AC\DC electrical measuring instruments |
• DCV output: ± (0 ~ 1020 V) • DCI output: ± (0 ~ 20.5 A) • ACV output: 1 mV~1020 V、10 Hz ~ 500 kHz • ACI output: 29 μA~ 20.5 A、10 Hz ~ 30 kHz • RES output: 0 Ω ~ 1100 MΩ • F output: 1 Hz ~ 2 MHz • It is suitable for calibrating digital multimeters of 6.5 digits and below, DC voltmeters and DC resistance meters of Class 0.01 and below, DC ammeters and AC voltmeters of Class 0.05 and below, AC ammeters of Class 0.1 and below, and Class 0.01 and below The frequency meter. |
☆ Calibrate AC\DC power meters |
|
☆ Calibrate capacitance meters(optional) |
|
☆ Calibrate temperature meters(optional) |
|
☆ Calibrate clamp ammeter (optional) |
• Equivalent to 1000AT current through input 20A current to the 50T Coil, which is suitable for calibrating DC clamp meters. |
TD1880 Precision Multi-function Calibrator is a multi-function calibrator with wide range and high accuracy, integrates AC/DC voltage and current standard source, can output excellent DC\AC voltage, DC\AC current, DC\AC power, phase, frequency, and analog resistance and capacitance etc. It’s applied for calibrating 6.5 digits or less digital multimeters.
DC voltage output: ± (0~1020 V)
DC current output: ± (0 ~ 20.5 A)
AC voltage output: 1mV ~ 1020 V
AC current output: 29 μA ~ 20.5 A
Sinusoidal wave frequency: 10 Hz ~ 500 kHz
Simulated DC resistance: 0 Ω ~ 1100 MΩ
2 MHz square wave frequency output
Phase output
AC/DC power output
Simulated capacitance(optional)
8 kinds of thermocouple output and measurement (option)
6 kinds of thermal resistance output (option)
Clamp meter calibration (optional)
☆ Calibrate AC\DC electrical measuring instruments |
• DCV output: ± (0 ~ 1020 V) • DCI output: ± (0 ~ 20.5 A) • ACV output: 1 mV~1020 V、10 Hz ~ 500 kHz • ACI output: 29 μA~ 20.5 A、10 Hz ~ 30 kHz • RES output: 0 Ω ~ 1100 MΩ • F output: 1 Hz ~ 2 MHz • It is suitable for calibrating digital multimeters of 6.5 digits and below, DC voltmeters and DC resistance meters of Class 0.01 and below, DC ammeters and AC voltmeters of Class 0.05 and below, AC ammeters of Class 0.1 and below, and Class 0.01 and below The frequency meter. |
☆ Calibrate AC\DC power meters |
|
☆ Calibrate capacitance meters(optional) |
|
☆ Calibrate temperature meters(optional) |
|
☆ Calibrate clamp ammeter (optional) |
• Equivalent to 1000AT current through input 20A current to the 50T Coil, which is suitable for calibrating DC clamp meters. |
☆ Front panel | |
Item | Description |
LCD touchable screen. | |
Operation panel. | |
Pulse frequency output terminal. | |
Simulated DC resistance, RTD output terminals. | |
Simulated capacitance output terminals. | |
AC/DC voltage output terminals. | |
Thermocouple output terminals. | |
AC/DC current output terminals. |
☆ Rear\Side Panel | |
Item | Description |
Cooling vents | |
Handle: used for carrying instruments | |
Foldable feet | |
USB/LAN/RS232 | |
Chassis ground binding post | |
AC 220V power interface |
☆Wide output range |
☆ High stability and accuracy |
• The DC voltage stability is up to 2 ppm/24h. |
☆ 4-terminals Simulated DC resistance |
• 2-terminals and 4-terminals resistance wiring mode. When calibrating the low value resistance, the four wire system wiring mode can effectively eliminate the measurement error caused by the resistance of the test lead and improve the measurement accuracy. • Compared with the standard physical resistance box, the analog resistance output has the following characteristics: ☑ Continuously adjustable, with better adjustment fineness and sensitivity ☑ Smaller size and lighter weight ☑ No influence of transfer switch error, residual resistance error and contact resistance error. |
☆ Security Protection Function |
|
|
☆ Easy to Operate |
|
☆ Multiple output\adjustment methods |
• “Direct output” mode: the value output can be set directly by physical key or touch screen. |
☆ Multiple output\adjustment methods |
• “Rotary Knob” mode, User can set in clockwise direction or anticlockwise direction. |
• Use the key ↑10%、↓10%、± 10、÷ 10 on the front panel, can realize the adjustment of the quantity value, and can also be used to perform the relevant test of linearity. |
☆ Special Test Software (optional) |
|
(Note: During automatic calibration, the tested equipment should have communication function and open communication protocol to realize data reading) |
☆ Front panel | |
Item | Description |
LCD touchable screen. | |
Operation panel. | |
Pulse frequency output terminal. | |
Simulated DC resistance, RTD output terminals. | |
Simulated capacitance output terminals. | |
AC/DC voltage output terminals. | |
Thermocouple output terminals. | |
AC/DC current output terminals. |
☆ Rear\Side Panel | |
Item | Description |
Cooling vents | |
Handle: used for carrying instruments | |
Foldable feet | |
USB/LAN/RS232 | |
Chassis ground binding post | |
AC 220V power interface |
☆Wide output range |
☆ High stability and accuracy |
• The DC voltage stability is up to 2 ppm/24h. |
☆ 4-terminals Simulated DC resistance |
• 2-terminals and 4-terminals resistance wiring mode. When calibrating the low value resistance, the four wire system wiring mode can effectively eliminate the measurement error caused by the resistance of the test lead and improve the measurement accuracy. • Compared with the standard physical resistance box, the analog resistance output has the following characteristics: ☑ Continuously adjustable, with better adjustment fineness and sensitivity ☑ Smaller size and lighter weight ☑ No influence of transfer switch error, residual resistance error and contact resistance error. |
☆ Security Protection Function |
|
|
☆ Easy to Operate |
|
☆ Multiple output\adjustment methods |
• “Direct output” mode: the value output can be set directly by physical key or touch screen. |
☆ Multiple output\adjustment methods |
• “Rotary Knob” mode, User can set in clockwise direction or anticlockwise direction. |
• Use the key ↑10%、↓10%、± 10、÷ 10 on the front panel, can realize the adjustment of the quantity value, and can also be used to perform the relevant test of linearity. |
☆ Special Test Software (optional) |
|
(Note: During automatic calibration, the tested equipment should have communication function and open communication protocol to realize data reading) |
Range | Resolution | Accuracy(Tcal±5°C) ±(ppm*output + μV) | Stability (Tcal±1°C) ±(ppm*output + μV) | Maximum Load [Internal Impedance] | |
90 days | 1 year | 24 hours | |||
0~330.0000 mV | 100 nV | 15 + 1.5 | 20 + 1.5 | 3 + 1 | [60 Ω] |
0~3.300000 V | 1 μV | 8 + 3 | 10 + 3 | 2 + 1.5 | 20 mA |
0~33.00000 V | 10 μV | 10 + 30 | 12 + 30 | 2 + 10 | 15 mA |
30.0000 V~330.0000 V | 100 μV | 15 + 200 | 18 + 200 | 2.5 + 100 | 10 mA |
100.000 V~1020.000 V | 1 mV | 15 + 2000 | 18 + 2000 | 3 + 300 | 10 mA |
• Manual or automatic range switch.
• Short-circuit and overload protection.
Range | Resolution | Accuracy(Tcal±5°C) ±(ppm*output + μA) | Compliance Voltage (V) | Max Inductive Load | |
90 days | 1 year | ||||
0~330.0000 μA | 100 pA | 80 + 0.02 | 100 + 0.02 | 11 V | 400 μH |
0~3.300000 mA | 1 nA | 65 + 0.03 | 80 + 0.03 | 11 V | |
0~33.00000 mA | 10 nA | 60 + 0.25 | 80 + 0.25 | 9 V | |
0~330.0000 mA | 100 nA | 60 + 2 | 80 + 2 | 7 V | |
0~1.100000 A | 1 μA | 80 + 20 | 100 + 20 | 6 V | |
1.000000 A~3.300000 A | 1 μA | 120 + 50 | 150 + 50 | 6 V | |
3.00000 A~20.50000 A | 10 μA | 280 + 300 | 350 + 300 | 4 V |
• Manual or automatic range switch.
• Open-circuit and overload protection.
• Supports continuous and long-term output of current under full range.
Range | Resolution | Accuracy, (Tcal±5°C) ±(ppm*output + Ω) | Allowable Current | |
90 days | 1 year | |||
0 Ω~11.00000 Ω | 10 μΩ | 32 + 0.008 | 40 + 0.01 | 1 mA~150 mA |
10.00000 Ω~33.00000 Ω | 10 μΩ | 24 + 0.012 | 30 + 0.015 | 1 mA~150 mA |
30.0000 Ω~110.0000 Ω | 100 μΩ | 20 + 0.012 | 25 + 0.015 | 1 mA~80 mA |
100.0000 Ω~330.0000 Ω | 100 μΩ | 20 + 0.016 | 25 + 0.02 | 1 mA~40 mA |
0.300000 kΩ~1.100000 kΩ | 1 mΩ | 20 + 0.016 | 25 + 0.02 | 1 mA~20 mA |
1.000000 kΩ~3.300000 kΩ | 1 mΩ | 20 + 0.16 | 25 + 0.2 | 0.1 mA~6 mA |
3.00000 kΩ~11.00000 kΩ | 10 mΩ | 20 + 0.16 | 25 + 0.1 | 0.1 mA~2 mA |
10.00000 kΩ~33.00000 kΩ | 10 mΩ | 22 + 0.8 | 28 + 1 | 10 μA~0.6 mA |
30.0000 kΩ~110.0000 kΩ | 100 mΩ | 22 + 0.8 | 28 + 1 | 10 μA~0.2 mA |
100.0000 kΩ~330.0000 kΩ | 100 mΩ | 25 + 8 | 32 + 10 | 1 μA~60 μA |
0.300000 MΩ~1.100000 MΩ | 1 Ω | 25 + 8 | 32 + 10 | 1 μA~20 μA |
1.000000 MΩ~3.300000 MΩ | 1 Ω | 40 + 120 | 60 + 150 | 0.25 μA~6 μA |
3.00000 MΩ~11.00000 MΩ | 10 Ω | 104 + 200 | 130 + 250 | 0.25 μA~2 μA |
10.00000 MΩ~33.00000 MΩ | 10 Ω | 200 + 2500 | 250 + 2500 | 25 nA~600 nA |
30.0000 MΩ~110.0000 MΩ | 100 Ω | 400 + 3000 | 500 + 3000 | 25 nA~200 nA |
100.0000 MΩ~330.0000 MΩ | 100 Ω | 2400 + 100000 | 3000 + 100000 | 2.5 nA~60 nA |
300.000 MΩ~1100.000 MΩ | 1 kΩ | 11000 + 480000 | 14000 + 480000 | 1 nA~20 nA |
• Manual or automatic range switch.
• Over-current and reverse connection protection.
Range | Frequency (Hz) | Accuracy(Tcal±5°C) ±(ppm*output + μV) | Resolution | Max Burden [Source internal resistance] | |
90 days | 1 year | ||||
1.00000 mV~ 33.00000 mV
| 10≤F≤45 | 600 + 6 | 800 + 6 | 10 nV | [10 Ω] |
45<F≤10k | 100 + 6 | 120 + 6 | |||
10k<F≤20k | 160 + 6 | 200 + 6 | |||
20k<F≤50k | 800 + 6 | 1000 + 6 | |||
50k<F≤100k | 2800 + 12 | 3500 + 12 | |||
100k<F≤500k | 6000 + 50 | 8000 + 50 | |||
30.0000 mV~ 330.0000 mV
| 10≤F≤45 | 250 + 8 | 300 + 8 | 100 nV | [60 Ω] |
45<F≤10k | 112 + 8 | 140 + 8 | |||
10k<F≤20k | 130 + 8 | 160 + 8 | |||
20k<F≤50k | 280 + 8 | 350 + 8 | |||
50k<F≤100k | 600 + 20 | 750 + 20 | |||
100k<F≤500k | 1600 + 70 | 2000 + 70 | |||
0.300000 V~ 3.300000 V
| 10≤F≤45 | 250 + 50 | 300 + 50 | 1 μV | 20 mA |
45<F≤10k | 80 + 50 | 100 + 50 | |||
10k<F≤20k | 150 + 50 | 180 + 50 | |||
20k<F≤50k | 240 + 50 | 300 + 50 | |||
50k<F≤100k | 550 + 100 | 700 + 100 | |||
100k<F≤500k | 2000 + 600 | 2400 + 600 | |||
3.00000 V~ 33.00000 V
| 10≤F≤45 | 160 + 650 | 200 + 650 | 10 μV | 15 mA |
45<F≤10k | 80 + 500 | 100 + 500 | |||
10k<F≤20k | 160 + 500 | 200 + 500 | |||
20k<F≤50k | 280 + 500 | 350 + 500 | |||
50k<F≤100k | 350 + 1500 | 550 + 1500 | |||
30.0000 V~ 330.0000 V
| 45≤F≤1k | 80 + 2000 | 100 + 2000 | 100 μV | 30 mA [1] |
1k<F≤10k | 80 + 6000 | 100 + 6000 | |||
10k<F≤20k | 160 + 6000 | 200 + 6000 | |||
20k<F≤50k | 240 + 6000 | 300 + 6000 | |||
50k<F≤100k | 1200 + 50000 | 1500 + 50000 | |||
300.000 V~ 1020.000 V | 45≤F≤1k | 100 + 10000 | 120 + 10000 | 1 mV | 8 mA [2] |
1k<F≤5k | 120 + 10000 | 150 + 10000 | |||
5k<F≤10k | 160 + 10000 | 200 + 10000 | |||
Note: [1] When the output frequency ≤ 3kHz, the maximum load is 30 mA; When the output frequency >3kHz, the maximum load is 5 mA. [2] When the output frequency ≤ 3kHz, the maximum load is 8 mA; When the output frequency >3kHz, the maximum load is 3 mA. |
• Short-circuit and overload protection.
Range | Frequency (Hz) | Accuracy(Tcal±5°C) ±(%* output + μA) | Resolution | Compliance Voltage (rms) | Max Inductive Load | |
90 days | 1 year | |||||
29.0000 μA~ 330.0000 μA | 10≤F≤20 | 0.08 + 0.1 | 0.1 + 0.1 | 0.1 nA | 7 V | 200 |
20<F≤45 | 0.04 + 0.1 | 0.05 + 0.1 | ||||
45<F≤1k | 0.024 + 0.1 | 0.03 + 0.1 | ||||
1k<F≤5k | 0.08 + 0.1 | 0.1 + 0.1 | ||||
5k<F≤10k | 0.16 + 0.2 | 0.2 + 0.2 | ||||
10k<F≤30k | 0.64 + 0.4 | 0.8 + 0.4 | ||||
0.300000 mA~ 3.300000 mA | 10≤F≤20 | 0.04 + 1.5 | 0.05 + 1.5 | 1 nA | 7 V | 200 |
20<F≤45 | 0.028 + 0.1 | 0.035 + 0.1 | ||||
45<F≤1k | 0.024 + 0.1 | 0.03 + 0.1 | ||||
1k<F≤5k | 0.024 + 0.2 | 0.03 + 0.2 | ||||
5k<F≤10k | 0.024 + 0.5 | 0.03 + 0.5 | ||||
10k<F≤30k | 0.16 + 0.6 | 0.2 + 0.6 | ||||
3.00000 mA~ 33.00000 mA | 10≤F≤20 | 0.04 + 2 | 0.05 + 2 | 10 nA | 7 V | 50 |
20<F≤45 | 0.02 + 2 | 0.025 + 2 | ||||
45<F≤1k | 0.016 + 2 | 0.02 + 2 | ||||
1k<F≤5k | 0.016 + 3 | 0.02 + 3 | ||||
5k<F≤10k | 0.04 + 5 | 0.05 + 5 | ||||
10k<F≤30k | 0.16 + 6 | 0.2 + 6 | ||||
30.0000 mA~ 330.0000 mA | 10≤F≤20 | 0.04 + 20 | 0.05 + 20 | 100 nA | 5 V | 50 |
20<F≤45 | 0.02 + 20 | 0.025 + 20 | ||||
45<F≤1k | 0.012 + 30 | 0.015 + 30 | ||||
1k<F≤5k | 0.016 + 30 | 0.02 + 30 | ||||
5k<F≤10k | 0.016 + 100 | 0.02 + 100 | ||||
10k<F≤30k | 0.08 + 500 | 0.1 + 500 | ||||
0.100000 A~ 1.100000 A | 10≤F≤20 | 0.04 + 100 | 0.05 + 100 | 1 μA | 5 V | 2.5 |
20<F≤45 | 0.024 + 50 | 0.03 + 50 | ||||
45<F≤1k | 0.016 + 50 | 0.02 + 50 | ||||
1k<F≤5k | 0.016 + 100 | 0.02 + 100 | ||||
5k<F≤10k | 0.04 + 500 | 0.05 + 500 | ||||
1.000000 A~ 3.300000 A | 10≤F≤20 | 0.04 + 100 | 0.05 + 100 | 1 μA | 4 V | 2.5 |
20<F≤45 | 0.024 + 100 | 0.03 + 100 | ||||
45<F≤1k | 0.016 + 100 | 0.02 + 100 | ||||
1k<F≤5k | 0.032 + 100 | 0.04 + 100 | ||||
5k<F≤10k | 0.04 + 900 | 0.05 + 900 | ||||
3.00000 A~ 20.50000 A | 45≤F≤100 | 0.024 + 1000 | 0.03 + 1000 | 10 μA | 3 V | 1 |
100<F≤1k | 0.032 + 1000 | 0.04 + 1000 | ||||
1k<F≤5k | 0.048 + 2000 | 0.06 + 2000 |
• Manual or automatic range switch.
• Open-circuit and overload protection.
• Supports continuous and long-term output of current under full range.
Range [1] | Resolution | Accuracy |
10.00000 Hz ≤ F ≤ 99.99999 Hz | 10 μHz | ± 0.005% |
100.0000 Hz ≤ F ≤ 999.9999 Hz | 0.1 mHz | ± 0.005% |
1.000000 kHz ≤ F ≤ 9.999999 kHz | 1 mHz | ± 0.005% |
10.00000 kHz ≤ F ≤ 99.99999 kHz | 10 mHz | ± 0.005% |
100.0000 kHz ≤ F ≤ 500.0000 kHz | 0.1 Hz | ± 0.005% |
Note [1]: Output mode: AC voltage or AC current |
Period | Current Range Voltage Range | Accuracy ± (%* power output)[1][2] | ||
3 mA ~ 300 mA | 300 mA ~ 3 A | 3 A ~ 20.5 A | ||
90 days | 30 mV ~ 1020 V | 0.016 | 0.018 | 0.039 |
1 year | 30 mV ~ 1020 V | 0.018 | 0.021 | 0.046 |
Period | Current Range Voltage Range | Accuracy ± (%* power output)[1][2] | ||
3 mA ~ 300 mA | 300 mA ~ 3 A | 3 A ~ 20.5 A | ||
90 days | 30 mV ~ 330 mV | 0.119 | 0.051 | 0.069 |
330 mV ~ 1020 V | 0.115 | 0.041 | 0.064 | |
1 year | 30 mV ~ 330 mV | 0.122 | 0.055 | 0.076 |
330 mV ~ 1020 V | 0.118 | 0.046 | 0.069 | |
Frequency (Hz) | Voltage Range (U) | Current Range (I) | Phase Range [2] (φ) | Power Factor Range [3] (λ) |
DC | 0~±1020 V | 0~±20.5 V | — | — |
10~45 | 30 mV~33 V | 3 mA~3.3 A | 0.000°~359.999° | -1~0~1 |
45~1k | 30 mV~1020 V | 3 mA~20.5 A | 0.000°~359.999° | -1~0~1 |
1k~5k | 3 V~1020 V | 30 mA~3.3 A | 0.000°~359.999° | -1~0~1 |
5k~10k | 3 V~1020 V | 30 mA~3.3 A | 0.000°~359.999° | -1~0~1 |
10k~30k | 3 V~330 V | 30 mA~330 mA | 0.000°~359.999° | -1~0~1 |
Note [1]: Phase resolution: 0.001° Note [2]: Power factor resolution: 0.000 01 |
Phase | Accuracy(k=2) (Tcal±5°C) | ||||||
10~65Hz | 65~500Hz | 500~1kHz | 1k~5kHz | 5k~10kHz | 10k~30kHz | ||
φ | 0.1° | 0.2° | 0.5° | 2° | 5° | 8° | |
phase (φ) | factor (λ) | Power accuracy caused by phase [4] | |||||
10~65Hz | 65~500Hz | 500~1kHz | 1k~5kHz | 5k~10kHz | 10k~30kHz | ||
0° | 1.00000 | 0.000% | 0.001% | 0.004% | 0.061% | 0.381% | 0.973% |
10° | 0.98481 | 0.031% | 0.062% | 0.158% | 0.676% | 1.917% | 3.427% |
20° | 0.93969 | 0.064% | 0.128% | 0.321% | 1.331% | 3.553% | 6.039% |
30° | 0.86603 | 0.101% | 0.202% | 0.508% | 2.076% | 5.412% | 9.008% |
40° | 0.76604 | 0.147% | 0.294% | 0.736% | 2.989% | 7.694% | 12.651% |
50° | 0.64279 | 0.208% | 0.417% | 1.044% | 4.220% | 10.767% | 17.559% |
60° | 0.50000 | 0.302% | 0.605% | 1.515% | 6.106% | 15.476% | 25.079% |
70° | 0.34202 | 0.480% | 0.960% | 2.041% | 9.649% | 24.326% | 39.211% |
80° | 0.17365 | 0.990% | 1.980% | 4.953% | 19.853% | 49.809% | 79.902% |
90° | 0.00000 | — | — | — | — | — | — |
Note [4]: Calculation formula: |
Range [1] | Resolution | Accuracy(Tcal ± 5°C) ± (ppm*RD + μHz) | Shake |
1.000000 Hz ≤ F ≤ 9.999999 Hz | 1 μHz | 20 + 20 | <2 ns |
10.00000 Hz ≤ F ≤ 99.99999 Hz | 10 μHz | ||
100.0000 Hz ≤ F ≤ 999.9999 Hz | 0.1 mHz | ||
1.000000 kHz ≤ F ≤ 9.999999 kHz | 1 mHz | ||
10.00000 kHz ≤ F ≤ 99.99999 kHz | 10 mHz | ||
100.0000 kHz ≤ F ≤ 999.9999 kHz | 0.1 Hz | ||
1.000000 MHz ≤ F ≤ 2.000000 MHz | 1 Hz | ||
Note [1]: The output type is TTL level. |
Range | Accuracy ±(%*output + Baseline Value) | Resolution | Allowable Current | |
90 days | 1 year | |||
1.100 0 nF~3.299 9 nF | 0.4 + 0.04 nF | 0.5 + 0.04 nF | 0.1 pF | 10 Hz~3 kHz |
3.300 0 nF~10.999 9 nF | 0.2 + 0.04 nF | 0.25 + 0.04 nF | 0.1 pF | 10 Hz~1 kHz |
11.000 0 nF~32.999 9 nF | 0.2 + 0.4 nF | 0.25 + 0.4 nF | 0.1 pF | 10 Hz~1 kHz |
33.000 nF~109.999 nF | 0.2 + 0.4 nF | 0.25 + 0.4 nF | 1 pF | 10 Hz~1 kHz |
110.000 nF~329.999 nF | 0.2 + 0.3 nF | 0.25 + 0.3 nF | 1 pF | 10 Hz~1 kHz |
0.330 00 μF~1.099 99 μF | 0.2 + 1 nF | 0.25 + 1 nF | 10 pF | 10 Hz~600 Hz |
1.100 00 μF~3.299 99 μF | 0.2 + 3 nF | 0.25 + 3 nF | 10 pF | 10 Hz~300 Hz |
3.300 0 μF~10.999 9 μF | 0.2 + 10 nF | 0.25 + 10 nF | 100 pF | 10 Hz~150 Hz |
11.000 μF~32.999 9 μF | 0.32 + 30 nF | 0.40 + 30 nF | 100 pF | 10 Hz~120 Hz |
33.000 μF~109.999 μF | 0.36 + 100 nF | 0.45 + 100 nF | 1 nF | 10 Hz~80 Hz |
110.000 μF~329.999 μF | 0.36 + 300 nF | 0.45 + 300 nF | 1 nF | 0 Hz~50 Hz |
0.330 00 mF~1.099 99 mF | 0.36 + 1 μF | 0.45 + 1 μF | 10 nF | 0 Hz~20 Hz |
1.100 00 mF~3.299 99 mF | 0.36 + 3 μF | 0.45 + 3 μF | 10 nF | 0 Hz~6 Hz |
3.300 0 mF~10.999 9 mF | 0.36 + 10 μF | 0.45 + 10 μF | 100 nF | 0 Hz~2 Hz |
11.000 0 mF~30.000 0 mF | 0.6 + 30 μF | 0.75 + 30 μF | 100 nF | 0 Hz~0.6 Hz |
Note [1]: The output is continuously adjustable. |
Type | Output Range[ [1] [2] °C | Accuracy[3], (Tcal±5°C) °C | ||
min | max | 90days | 1year | |
B | 410 | 600 | 0.30 | 0.35 |
600 | 900 | 0.30 | 0.31 | |
900 | 1800 | 0.26 | 0.30 | |
E | -200 | 0 | 0.10 | 0.11 |
0 | 600 | 0.07 | 0.08 | |
600 | 1000 | 0.09 | 0.10 | |
J | -200 | -100 | 0.12 | 0.13 |
-100 | 750 | 0.09 | 0.09 | |
750 | 1200 | 0.10 | 0.10 | |
K | -200 | -100 | 0.15 | 0.20 |
-100 | 1000 | 0.10 | 0.10 | |
1000 | 1370 | 0.12 | 0.12 | |
N | -200 | -100 | 0.20 | 0.25 |
-100 | 400 | 0.12 | 0.12 | |
400 | 1300 | 0.10 | 0.12 | |
R | -50 | 50 | 0.30 | 0.40 |
50 | 300 | 0.28 | 0.32 | |
300 | 1000 | 0.22 | 0.23 | |
1000 | 1750 | 0.25 | 0.25 | |
S | -50 | 50 | 0.30 | 0.40 |
50 | 300 | 0.31 | 0.34 | |
300 | 1000 | 0.24 | 0.24 | |
1000 | 1750 | 0.22 | 0.23 | |
T | -200 | 100 | 0.17 | 0.25 |
-100 | 0 | 0.08 | 0.11 | |
0 | 400 | 0.08 | 0.08 | |
Note [1]: Resolution: 0.01 ° C Note [2]: Internal resistance of output source: 10 Ω Note [3]: Excluding thermocouple error Note [4]: Use external compensation, S, R, B, K, N, E, J, T conform to ITS-90 international temperature standard. |
Type
| Output Range [1] °C | Accuracy, (Tcal±5°C) °C | ||
Min | Max | 90 days | 1 year | |
Pt385, 100 Ω | -200 | 0 | 0.05 | 0.05 |
0 | 300 | 0.08 | 0.08 | |
300 | 850 | 0.12 | 0.12 | |
Pt385, 200 Ω | -200 | 250 | 0.04 | 0.04 |
250 | 630 | 0.10 | 0.15 | |
Pt385, 500 Ω | -200 | -30 | 0.36 | 0.40 |
-30 | 630 | 0.10 | 0.11 | |
Pt385, 1000 Ω | -200 | 0 | 0.027 | 0.03 |
0 | 300 | 0.054 | 0.06 | |
300 | 600 | 0.063 | 0.07 | |
Cu50 | -50 | 150 | 0.09 | 0.09 |
Cu100 | -50 | 150 | 0.045 | 0.045 |
Note[1]: Resolution: 0.001°C |
Power supply | AC (220±22) V, (50±2) Hz |
Maximum power consumption | 600 VA |
Warm up time | Twice the time since last warmed up, to a maximum of 30 minutes. |
Temperature performance | Operating temperature: 0°C ~ 50°C Calibration temperature: 15°C ~35°C Storage temperature: -20°C ~ 50°C |
Humidity performance | Operating humidity: <80% @ 30°C, <70% @ 40°C, <40% @ 50°C Storage humidity: <95%, No condensation |
Communication interface | RS232×1、USB×1、LAN×1 |
Dimensions | 440 mm(W)× 462 mm (D) × 206 mm (H) , Handles and feets excluded. |
Weight | About 24 kg |
Function | TD1880A | TD1880B |
AC/DC Voltage standard source (U) | ★ | ★ |
AC/DC current standard source (I) | ★ | ★ |
Simulated DC resistance (R) | ★ | ★ |
AC/DC power output (P) | ★ | ★ |
Square wave frequency output (F) | ★ | ★ |
Simulated capacitance (C) | — | ★ |
Simulated TC output &TC measurement(TC) | — | ★ |
Simulated RTD output (RTD) | — | ★ |
AC/DC auxiliary voltage output (AUX) | — | ★ |
Range | Resolution | Accuracy(Tcal±5°C) ±(ppm*output + μV) | Stability (Tcal±1°C) ±(ppm*output + μV) | Maximum Load [Internal Impedance] | |
90 days | 1 year | 24 hours | |||
0~330.0000 mV | 100 nV | 15 + 1.5 | 20 + 1.5 | 3 + 1 | [60 Ω] |
0~3.300000 V | 1 μV | 8 + 3 | 10 + 3 | 2 + 1.5 | 20 mA |
0~33.00000 V | 10 μV | 10 + 30 | 12 + 30 | 2 + 10 | 15 mA |
30.0000 V~330.0000 V | 100 μV | 15 + 200 | 18 + 200 | 2.5 + 100 | 10 mA |
100.000 V~1020.000 V | 1 mV | 15 + 2000 | 18 + 2000 | 3 + 300 | 10 mA |
• Manual or automatic range switch.
• Short-circuit and overload protection.
Range | Resolution | Accuracy(Tcal±5°C) ±(ppm*output + μA) | Compliance Voltage (V) | Max Inductive Load | |
90 days | 1 year | ||||
0~330.0000 μA | 100 pA | 80 + 0.02 | 100 + 0.02 | 11 V | 400 μH |
0~3.300000 mA | 1 nA | 65 + 0.03 | 80 + 0.03 | 11 V | |
0~33.00000 mA | 10 nA | 60 + 0.25 | 80 + 0.25 | 9 V | |
0~330.0000 mA | 100 nA | 60 + 2 | 80 + 2 | 7 V | |
0~1.100000 A | 1 μA | 80 + 20 | 100 + 20 | 6 V | |
1.000000 A~3.300000 A | 1 μA | 120 + 50 | 150 + 50 | 6 V | |
3.00000 A~20.50000 A | 10 μA | 280 + 300 | 350 + 300 | 4 V |
• Manual or automatic range switch.
• Open-circuit and overload protection.
• Supports continuous and long-term output of current under full range.
Range | Resolution | Accuracy, (Tcal±5°C) ±(ppm*output + Ω) | Allowable Current | |
90 days | 1 year | |||
0 Ω~11.00000 Ω | 10 μΩ | 32 + 0.008 | 40 + 0.01 | 1 mA~150 mA |
10.00000 Ω~33.00000 Ω | 10 μΩ | 24 + 0.012 | 30 + 0.015 | 1 mA~150 mA |
30.0000 Ω~110.0000 Ω | 100 μΩ | 20 + 0.012 | 25 + 0.015 | 1 mA~80 mA |
100.0000 Ω~330.0000 Ω | 100 μΩ | 20 + 0.016 | 25 + 0.02 | 1 mA~40 mA |
0.300000 kΩ~1.100000 kΩ | 1 mΩ | 20 + 0.016 | 25 + 0.02 | 1 mA~20 mA |
1.000000 kΩ~3.300000 kΩ | 1 mΩ | 20 + 0.16 | 25 + 0.2 | 0.1 mA~6 mA |
3.00000 kΩ~11.00000 kΩ | 10 mΩ | 20 + 0.16 | 25 + 0.1 | 0.1 mA~2 mA |
10.00000 kΩ~33.00000 kΩ | 10 mΩ | 22 + 0.8 | 28 + 1 | 10 μA~0.6 mA |
30.0000 kΩ~110.0000 kΩ | 100 mΩ | 22 + 0.8 | 28 + 1 | 10 μA~0.2 mA |
100.0000 kΩ~330.0000 kΩ | 100 mΩ | 25 + 8 | 32 + 10 | 1 μA~60 μA |
0.300000 MΩ~1.100000 MΩ | 1 Ω | 25 + 8 | 32 + 10 | 1 μA~20 μA |
1.000000 MΩ~3.300000 MΩ | 1 Ω | 40 + 120 | 60 + 150 | 0.25 μA~6 μA |
3.00000 MΩ~11.00000 MΩ | 10 Ω | 104 + 200 | 130 + 250 | 0.25 μA~2 μA |
10.00000 MΩ~33.00000 MΩ | 10 Ω | 200 + 2500 | 250 + 2500 | 25 nA~600 nA |
30.0000 MΩ~110.0000 MΩ | 100 Ω | 400 + 3000 | 500 + 3000 | 25 nA~200 nA |
100.0000 MΩ~330.0000 MΩ | 100 Ω | 2400 + 100000 | 3000 + 100000 | 2.5 nA~60 nA |
300.000 MΩ~1100.000 MΩ | 1 kΩ | 11000 + 480000 | 14000 + 480000 | 1 nA~20 nA |
• Manual or automatic range switch.
• Over-current and reverse connection protection.
Range | Frequency (Hz) | Accuracy(Tcal±5°C) ±(ppm*output + μV) | Resolution | Max Burden [Source internal resistance] | |
90 days | 1 year | ||||
1.00000 mV~ 33.00000 mV
| 10≤F≤45 | 600 + 6 | 800 + 6 | 10 nV | [10 Ω] |
45<F≤10k | 100 + 6 | 120 + 6 | |||
10k<F≤20k | 160 + 6 | 200 + 6 | |||
20k<F≤50k | 800 + 6 | 1000 + 6 | |||
50k<F≤100k | 2800 + 12 | 3500 + 12 | |||
100k<F≤500k | 6000 + 50 | 8000 + 50 | |||
30.0000 mV~ 330.0000 mV
| 10≤F≤45 | 250 + 8 | 300 + 8 | 100 nV | [60 Ω] |
45<F≤10k | 112 + 8 | 140 + 8 | |||
10k<F≤20k | 130 + 8 | 160 + 8 | |||
20k<F≤50k | 280 + 8 | 350 + 8 | |||
50k<F≤100k | 600 + 20 | 750 + 20 | |||
100k<F≤500k | 1600 + 70 | 2000 + 70 | |||
0.300000 V~ 3.300000 V
| 10≤F≤45 | 250 + 50 | 300 + 50 | 1 μV | 20 mA |
45<F≤10k | 80 + 50 | 100 + 50 | |||
10k<F≤20k | 150 + 50 | 180 + 50 | |||
20k<F≤50k | 240 + 50 | 300 + 50 | |||
50k<F≤100k | 550 + 100 | 700 + 100 | |||
100k<F≤500k | 2000 + 600 | 2400 + 600 | |||
3.00000 V~ 33.00000 V
| 10≤F≤45 | 160 + 650 | 200 + 650 | 10 μV | 15 mA |
45<F≤10k | 80 + 500 | 100 + 500 | |||
10k<F≤20k | 160 + 500 | 200 + 500 | |||
20k<F≤50k | 280 + 500 | 350 + 500 | |||
50k<F≤100k | 350 + 1500 | 550 + 1500 | |||
30.0000 V~ 330.0000 V
| 45≤F≤1k | 80 + 2000 | 100 + 2000 | 100 μV | 30 mA [1] |
1k<F≤10k | 80 + 6000 | 100 + 6000 | |||
10k<F≤20k | 160 + 6000 | 200 + 6000 | |||
20k<F≤50k | 240 + 6000 | 300 + 6000 | |||
50k<F≤100k | 1200 + 50000 | 1500 + 50000 | |||
300.000 V~ 1020.000 V | 45≤F≤1k | 100 + 10000 | 120 + 10000 | 1 mV | 8 mA [2] |
1k<F≤5k | 120 + 10000 | 150 + 10000 | |||
5k<F≤10k | 160 + 10000 | 200 + 10000 | |||
Note: [1] When the output frequency ≤ 3kHz, the maximum load is 30 mA; When the output frequency >3kHz, the maximum load is 5 mA. [2] When the output frequency ≤ 3kHz, the maximum load is 8 mA; When the output frequency >3kHz, the maximum load is 3 mA. |
• Short-circuit and overload protection.
Range | Frequency (Hz) | Accuracy(Tcal±5°C) ±(%* output + μA) | Resolution | Compliance Voltage (rms) | Max Inductive Load | |
90 days | 1 year | |||||
29.0000 μA~ 330.0000 μA | 10≤F≤20 | 0.08 + 0.1 | 0.1 + 0.1 | 0.1 nA | 7 V | 200 |
20<F≤45 | 0.04 + 0.1 | 0.05 + 0.1 | ||||
45<F≤1k | 0.024 + 0.1 | 0.03 + 0.1 | ||||
1k<F≤5k | 0.08 + 0.1 | 0.1 + 0.1 | ||||
5k<F≤10k | 0.16 + 0.2 | 0.2 + 0.2 | ||||
10k<F≤30k | 0.64 + 0.4 | 0.8 + 0.4 | ||||
0.300000 mA~ 3.300000 mA | 10≤F≤20 | 0.04 + 1.5 | 0.05 + 1.5 | 1 nA | 7 V | 200 |
20<F≤45 | 0.028 + 0.1 | 0.035 + 0.1 | ||||
45<F≤1k | 0.024 + 0.1 | 0.03 + 0.1 | ||||
1k<F≤5k | 0.024 + 0.2 | 0.03 + 0.2 | ||||
5k<F≤10k | 0.024 + 0.5 | 0.03 + 0.5 | ||||
10k<F≤30k | 0.16 + 0.6 | 0.2 + 0.6 | ||||
3.00000 mA~ 33.00000 mA | 10≤F≤20 | 0.04 + 2 | 0.05 + 2 | 10 nA | 7 V | 50 |
20<F≤45 | 0.02 + 2 | 0.025 + 2 | ||||
45<F≤1k | 0.016 + 2 | 0.02 + 2 | ||||
1k<F≤5k | 0.016 + 3 | 0.02 + 3 | ||||
5k<F≤10k | 0.04 + 5 | 0.05 + 5 | ||||
10k<F≤30k | 0.16 + 6 | 0.2 + 6 | ||||
30.0000 mA~ 330.0000 mA | 10≤F≤20 | 0.04 + 20 | 0.05 + 20 | 100 nA | 5 V | 50 |
20<F≤45 | 0.02 + 20 | 0.025 + 20 | ||||
45<F≤1k | 0.012 + 30 | 0.015 + 30 | ||||
1k<F≤5k | 0.016 + 30 | 0.02 + 30 | ||||
5k<F≤10k | 0.016 + 100 | 0.02 + 100 | ||||
10k<F≤30k | 0.08 + 500 | 0.1 + 500 | ||||
0.100000 A~ 1.100000 A | 10≤F≤20 | 0.04 + 100 | 0.05 + 100 | 1 μA | 5 V | 2.5 |
20<F≤45 | 0.024 + 50 | 0.03 + 50 | ||||
45<F≤1k | 0.016 + 50 | 0.02 + 50 | ||||
1k<F≤5k | 0.016 + 100 | 0.02 + 100 | ||||
5k<F≤10k | 0.04 + 500 | 0.05 + 500 | ||||
1.000000 A~ 3.300000 A | 10≤F≤20 | 0.04 + 100 | 0.05 + 100 | 1 μA | 4 V | 2.5 |
20<F≤45 | 0.024 + 100 | 0.03 + 100 | ||||
45<F≤1k | 0.016 + 100 | 0.02 + 100 | ||||
1k<F≤5k | 0.032 + 100 | 0.04 + 100 | ||||
5k<F≤10k | 0.04 + 900 | 0.05 + 900 | ||||
3.00000 A~ 20.50000 A | 45≤F≤100 | 0.024 + 1000 | 0.03 + 1000 | 10 μA | 3 V | 1 |
100<F≤1k | 0.032 + 1000 | 0.04 + 1000 | ||||
1k<F≤5k | 0.048 + 2000 | 0.06 + 2000 |
• Manual or automatic range switch.
• Open-circuit and overload protection.
• Supports continuous and long-term output of current under full range.
Range [1] | Resolution | Accuracy |
10.00000 Hz ≤ F ≤ 99.99999 Hz | 10 μHz | ± 0.005% |
100.0000 Hz ≤ F ≤ 999.9999 Hz | 0.1 mHz | ± 0.005% |
1.000000 kHz ≤ F ≤ 9.999999 kHz | 1 mHz | ± 0.005% |
10.00000 kHz ≤ F ≤ 99.99999 kHz | 10 mHz | ± 0.005% |
100.0000 kHz ≤ F ≤ 500.0000 kHz | 0.1 Hz | ± 0.005% |
Note [1]: Output mode: AC voltage or AC current |
Period | Current Range Voltage Range | Accuracy ± (%* power output)[1][2] | ||
3 mA ~ 300 mA | 300 mA ~ 3 A | 3 A ~ 20.5 A | ||
90 days | 30 mV ~ 1020 V | 0.016 | 0.018 | 0.039 |
1 year | 30 mV ~ 1020 V | 0.018 | 0.021 | 0.046 |
Period | Current Range Voltage Range | Accuracy ± (%* power output)[1][2] | ||
3 mA ~ 300 mA | 300 mA ~ 3 A | 3 A ~ 20.5 A | ||
90 days | 30 mV ~ 330 mV | 0.119 | 0.051 | 0.069 |
330 mV ~ 1020 V | 0.115 | 0.041 | 0.064 | |
1 year | 30 mV ~ 330 mV | 0.122 | 0.055 | 0.076 |
330 mV ~ 1020 V | 0.118 | 0.046 | 0.069 | |
Frequency (Hz) | Voltage Range (U) | Current Range (I) | Phase Range [2] (φ) | Power Factor Range [3] (λ) |
DC | 0~±1020 V | 0~±20.5 V | — | — |
10~45 | 30 mV~33 V | 3 mA~3.3 A | 0.000°~359.999° | -1~0~1 |
45~1k | 30 mV~1020 V | 3 mA~20.5 A | 0.000°~359.999° | -1~0~1 |
1k~5k | 3 V~1020 V | 30 mA~3.3 A | 0.000°~359.999° | -1~0~1 |
5k~10k | 3 V~1020 V | 30 mA~3.3 A | 0.000°~359.999° | -1~0~1 |
10k~30k | 3 V~330 V | 30 mA~330 mA | 0.000°~359.999° | -1~0~1 |
Note [1]: Phase resolution: 0.001° Note [2]: Power factor resolution: 0.000 01 |
Phase | Accuracy(k=2) (Tcal±5°C) | ||||||
10~65Hz | 65~500Hz | 500~1kHz | 1k~5kHz | 5k~10kHz | 10k~30kHz | ||
φ | 0.1° | 0.2° | 0.5° | 2° | 5° | 8° | |
phase (φ) | factor (λ) | Power accuracy caused by phase [4] | |||||
10~65Hz | 65~500Hz | 500~1kHz | 1k~5kHz | 5k~10kHz | 10k~30kHz | ||
0° | 1.00000 | 0.000% | 0.001% | 0.004% | 0.061% | 0.381% | 0.973% |
10° | 0.98481 | 0.031% | 0.062% | 0.158% | 0.676% | 1.917% | 3.427% |
20° | 0.93969 | 0.064% | 0.128% | 0.321% | 1.331% | 3.553% | 6.039% |
30° | 0.86603 | 0.101% | 0.202% | 0.508% | 2.076% | 5.412% | 9.008% |
40° | 0.76604 | 0.147% | 0.294% | 0.736% | 2.989% | 7.694% | 12.651% |
50° | 0.64279 | 0.208% | 0.417% | 1.044% | 4.220% | 10.767% | 17.559% |
60° | 0.50000 | 0.302% | 0.605% | 1.515% | 6.106% | 15.476% | 25.079% |
70° | 0.34202 | 0.480% | 0.960% | 2.041% | 9.649% | 24.326% | 39.211% |
80° | 0.17365 | 0.990% | 1.980% | 4.953% | 19.853% | 49.809% | 79.902% |
90° | 0.00000 | — | — | — | — | — | — |
Note [4]: Calculation formula: |
Range [1] | Resolution | Accuracy(Tcal ± 5°C) ± (ppm*RD + μHz) | Shake |
1.000000 Hz ≤ F ≤ 9.999999 Hz | 1 μHz | 20 + 20 | <2 ns |
10.00000 Hz ≤ F ≤ 99.99999 Hz | 10 μHz | ||
100.0000 Hz ≤ F ≤ 999.9999 Hz | 0.1 mHz | ||
1.000000 kHz ≤ F ≤ 9.999999 kHz | 1 mHz | ||
10.00000 kHz ≤ F ≤ 99.99999 kHz | 10 mHz | ||
100.0000 kHz ≤ F ≤ 999.9999 kHz | 0.1 Hz | ||
1.000000 MHz ≤ F ≤ 2.000000 MHz | 1 Hz | ||
Note [1]: The output type is TTL level. |
Range | Accuracy ±(%*output + Baseline Value) | Resolution | Allowable Current | |
90 days | 1 year | |||
1.100 0 nF~3.299 9 nF | 0.4 + 0.04 nF | 0.5 + 0.04 nF | 0.1 pF | 10 Hz~3 kHz |
3.300 0 nF~10.999 9 nF | 0.2 + 0.04 nF | 0.25 + 0.04 nF | 0.1 pF | 10 Hz~1 kHz |
11.000 0 nF~32.999 9 nF | 0.2 + 0.4 nF | 0.25 + 0.4 nF | 0.1 pF | 10 Hz~1 kHz |
33.000 nF~109.999 nF | 0.2 + 0.4 nF | 0.25 + 0.4 nF | 1 pF | 10 Hz~1 kHz |
110.000 nF~329.999 nF | 0.2 + 0.3 nF | 0.25 + 0.3 nF | 1 pF | 10 Hz~1 kHz |
0.330 00 μF~1.099 99 μF | 0.2 + 1 nF | 0.25 + 1 nF | 10 pF | 10 Hz~600 Hz |
1.100 00 μF~3.299 99 μF | 0.2 + 3 nF | 0.25 + 3 nF | 10 pF | 10 Hz~300 Hz |
3.300 0 μF~10.999 9 μF | 0.2 + 10 nF | 0.25 + 10 nF | 100 pF | 10 Hz~150 Hz |
11.000 μF~32.999 9 μF | 0.32 + 30 nF | 0.40 + 30 nF | 100 pF | 10 Hz~120 Hz |
33.000 μF~109.999 μF | 0.36 + 100 nF | 0.45 + 100 nF | 1 nF | 10 Hz~80 Hz |
110.000 μF~329.999 μF | 0.36 + 300 nF | 0.45 + 300 nF | 1 nF | 0 Hz~50 Hz |
0.330 00 mF~1.099 99 mF | 0.36 + 1 μF | 0.45 + 1 μF | 10 nF | 0 Hz~20 Hz |
1.100 00 mF~3.299 99 mF | 0.36 + 3 μF | 0.45 + 3 μF | 10 nF | 0 Hz~6 Hz |
3.300 0 mF~10.999 9 mF | 0.36 + 10 μF | 0.45 + 10 μF | 100 nF | 0 Hz~2 Hz |
11.000 0 mF~30.000 0 mF | 0.6 + 30 μF | 0.75 + 30 μF | 100 nF | 0 Hz~0.6 Hz |
Note [1]: The output is continuously adjustable. |
Type | Output Range[ [1] [2] °C | Accuracy[3], (Tcal±5°C) °C | ||
min | max | 90days | 1year | |
B | 410 | 600 | 0.30 | 0.35 |
600 | 900 | 0.30 | 0.31 | |
900 | 1800 | 0.26 | 0.30 | |
E | -200 | 0 | 0.10 | 0.11 |
0 | 600 | 0.07 | 0.08 | |
600 | 1000 | 0.09 | 0.10 | |
J | -200 | -100 | 0.12 | 0.13 |
-100 | 750 | 0.09 | 0.09 | |
750 | 1200 | 0.10 | 0.10 | |
K | -200 | -100 | 0.15 | 0.20 |
-100 | 1000 | 0.10 | 0.10 | |
1000 | 1370 | 0.12 | 0.12 | |
N | -200 | -100 | 0.20 | 0.25 |
-100 | 400 | 0.12 | 0.12 | |
400 | 1300 | 0.10 | 0.12 | |
R | -50 | 50 | 0.30 | 0.40 |
50 | 300 | 0.28 | 0.32 | |
300 | 1000 | 0.22 | 0.23 | |
1000 | 1750 | 0.25 | 0.25 | |
S | -50 | 50 | 0.30 | 0.40 |
50 | 300 | 0.31 | 0.34 | |
300 | 1000 | 0.24 | 0.24 | |
1000 | 1750 | 0.22 | 0.23 | |
T | -200 | 100 | 0.17 | 0.25 |
-100 | 0 | 0.08 | 0.11 | |
0 | 400 | 0.08 | 0.08 | |
Note [1]: Resolution: 0.01 ° C Note [2]: Internal resistance of output source: 10 Ω Note [3]: Excluding thermocouple error Note [4]: Use external compensation, S, R, B, K, N, E, J, T conform to ITS-90 international temperature standard. |
Type
| Output Range [1] °C | Accuracy, (Tcal±5°C) °C | ||
Min | Max | 90 days | 1 year | |
Pt385, 100 Ω | -200 | 0 | 0.05 | 0.05 |
0 | 300 | 0.08 | 0.08 | |
300 | 850 | 0.12 | 0.12 | |
Pt385, 200 Ω | -200 | 250 | 0.04 | 0.04 |
250 | 630 | 0.10 | 0.15 | |
Pt385, 500 Ω | -200 | -30 | 0.36 | 0.40 |
-30 | 630 | 0.10 | 0.11 | |
Pt385, 1000 Ω | -200 | 0 | 0.027 | 0.03 |
0 | 300 | 0.054 | 0.06 | |
300 | 600 | 0.063 | 0.07 | |
Cu50 | -50 | 150 | 0.09 | 0.09 |
Cu100 | -50 | 150 | 0.045 | 0.045 |
Note[1]: Resolution: 0.001°C |
Power supply | AC (220±22) V, (50±2) Hz |
Maximum power consumption | 600 VA |
Warm up time | Twice the time since last warmed up, to a maximum of 30 minutes. |
Temperature performance | Operating temperature: 0°C ~ 50°C Calibration temperature: 15°C ~35°C Storage temperature: -20°C ~ 50°C |
Humidity performance | Operating humidity: <80% @ 30°C, <70% @ 40°C, <40% @ 50°C Storage humidity: <95%, No condensation |
Communication interface | RS232×1、USB×1、LAN×1 |
Dimensions | 440 mm(W)× 462 mm (D) × 206 mm (H) , Handles and feets excluded. |
Weight | About 24 kg |
Function | TD1880A | TD1880B |
AC/DC Voltage standard source (U) | ★ | ★ |
AC/DC current standard source (I) | ★ | ★ |
Simulated DC resistance (R) | ★ | ★ |
AC/DC power output (P) | ★ | ★ |
Square wave frequency output (F) | ★ | ★ |
Simulated capacitance (C) | — | ★ |
Simulated TC output &TC measurement(TC) | — | ★ |
Simulated RTD output (RTD) | — | ★ |
AC/DC auxiliary voltage output (AUX) | — | ★ |